Twin machines validation for VMAT treatments using electronic portal-imaging device: a multicenter study

被引:2
|
作者
Fenoglietto, P. [1 ]
Khodri, M. [2 ]
Nguyen, D. [3 ]
Josserand-Pietri, F. [4 ]
Ailleres, N. [1 ]
机构
[1] Inst Reg Canc Montpellier ICM, Dept Oncol Radiotherapie, F-34298 Montpellier 5, France
[2] ICL, Dept Oncol Radiotherapie, St Etienne, France
[3] Grp ORLAM, Serv Radiotherapie, Macon, France
[4] Grp ORLAM, Serv Radiotherapie, Villeurbanne, France
关键词
VOLUMETRIC-MODULATED ARC; QUALITY-ASSURANCE; CLINICAL IMPLEMENTATION; PROSTATE-CANCER; DYNAMIC IMRT; RADIOTHERAPY; DELIVERY; THERAPY; RAPIDARC; VERIFICATION;
D O I
10.1186/s13014-015-0577-3
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To verify the accuracy of volumetric arc therapy (VMAT) using the RapidArc (TM) device when switching patients from one single linear accelerator (linac) to a paired energy and mechanics "twin" linac without reoptimization of the original treatment plan. Patients and Methods: Four centers using 8 linacs were involved in this study. Seventy-four patients previously treated with the 6MV photon RapidArc (TM) technique were selected for analysis, using 242 measurements. In each institution, all patients were planned on linac A, and their plans were verified both on linac A and on the twin linac B. Verifications were done using the amorphous silicium electronic portal imager (EPID) of the linacs and were analyzed with the EpiQa software (Epidos, Bratislavia, Slovakia). The gamma index formalism was used for validation with a double threshold of 3 % and 3 mm with a measurement resolution of 0.39 mm/pixel, and a smoothed resolution of approximately 2.5 mm. Results: The number of points passing the gamma criteria between the measured and computed doses was 94.79 +/- 2.57 % for linac A and 94.61 +/- 2.46 % for linac B. Concerning the smoothed measurement analysis, 98.67 +/- 1.26 % and 98.59 +/- 1.20 % points passing the threshold were obtained for linacs A and B, respectively. The difference between the 2 dose matrices acquired on the EPID was very small, with 99.92 +/- 0.06 % of the points passing the criteria. Conclusion: For linacs sharing the same mechanical and energy parameters, this study tends to indicate that patients may be safely switched from treatment with one linac to treatment with its twin linac using the same VMAT plan.
引用
收藏
页数:9
相关论文
共 50 条
  • [31] Dosimetric investigation and portal dose image prediction using an amorphous silicon electronic portal imaging device
    McCurdy, BMC
    Luchka, K
    Pistorius, S
    MEDICAL PHYSICS, 2001, 28 (06) : 911 - 924
  • [32] Portal dose measurements with a video-based electronic portal imaging device using a deconvolution algorithm
    Pasma, KL
    Kroonwijk, M
    Visser, AG
    Heijmen, BJM
    PROCEEDINGS OF THE XIITH INTERNATIONAL CONFERENCE ON THE USE OF COMPUTERS IN RADIATION THERAPY, 1997, : 282 - 284
  • [33] Study of X-ray field junction dose using an a-Si electronic portal imaging device
    Madebo, Mebratu
    Perkins, A.
    Fox, C.
    Johnston, P.
    Kron, T.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2010, 33 (01) : 45 - 50
  • [34] Study of X-ray field junction dose using an a-Si electronic portal imaging device
    Mebratu Madebo
    A. Perkins
    C. Fox
    P. Johnston
    T. Kron
    Australasian Physical & Engineering Sciences in Medicine, 2010, 33 : 45 - 50
  • [35] Comparison of a MLC calibration method using a diode array and electronic portal imaging device
    Simon, T.
    Fox, C.
    Li, J.
    Palta, I.
    Simon, W.
    Liu, C.
    MEDICAL PHYSICS, 2007, 34 (06) : 2337 - 2337
  • [36] Development and clinical implementation of in-vivo dosimetry using an electronic portal imaging device
    Kroonwijk, M
    Pasma, KL
    Quint, S
    Visser, AG
    Heijmen, BJM
    PROCEEDINGS OF THE XIITH INTERNATIONAL CONFERENCE ON THE USE OF COMPUTERS IN RADIATION THERAPY, 1997, : 276 - 278
  • [37] Commissioning of An In-Vivo Quality Assurance Method Using the Electronic Portal Imaging Device
    Morin, O.
    Held, M.
    Pouliot, J.
    MEDICAL PHYSICS, 2014, 41 (06) : 326 - 327
  • [38] A Multileaf Collimator Calibration and Quality Assurance Technique Using An Electronic Portal Imaging Device
    Lebron, S.
    Yan, G.
    Li, J.
    Lu, B.
    Liu, C.
    MEDICAL PHYSICS, 2016, 43 (06) : 3685 - 3685
  • [39] Radiological thickness measurement using a liquid ionization chamber electronic portal imaging device
    Evans, PM
    Donovan, EM
    Partridge, M
    Bidmead, AM
    Garton, A
    Mubata, C
    PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (06): : N89 - N97
  • [40] Transit Dosimetry for Verification of Dose Delivery Using Electronic Portal Imaging Device (EPID)
    Baek, T.
    Chung, E.
    Lee, S.
    Yoon, M.
    MEDICAL PHYSICS, 2014, 41 (06) : 301 - 301