Aortic valve replacement with small-sized disc prostheses (Medtronic Hall)

被引:0
|
作者
Brazao, AJ [1 ]
Prieto, D [1 ]
de Oliveira, JF [1 ]
Eugenio, L [1 ]
Antunes, MJ [1 ]
机构
[1] Univ Hosp, Dept Cardiothorac Surg, P-3049 Coimbra, Portugal
来源
JOURNAL OF HEART VALVE DISEASE | 1999年 / 8卷 / 06期
关键词
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background and aim of the study: Several studies have demonstrated a worse performance of small prostheses in the narrow aortic root. However, modern low-profile mechanical prostheses have improved hemodynamic performance, are easy to implant, and have been used increasingly in elderly patients, where narrow roots are most frequently observed. We describe our experience in patients with aortic annuli >21 mm with the use of Medtronic Hall prostheses. Methods: Between April 1988 and December 1997, a total of 332 patients (218 females, 114 males; mean age 59.3 +/- 9.8 years (range: 29-75 years) received Medtronic Hall size 20 (n = 140), 21 (n = 96) or 22 (n = 96) prostheses. The mean body surface area was 1.59 +/- 0.12 m(2) (range: 1.27-2.01 m(2)); 140 patients were in NYHA functional classes III/IV. There were no significant differences in clinical characteristics of patients in the three prosthesis-size groups. Results: The mean preoperative systolic left ventricular/aorta gradient was 64.5 +/- 24.8 mmHg (no significant inter-group difference). After cardiopulmonary bypass, peak gradients through the aortic prostheses were 13.9 +/- 8.0 mmHg in size 20 valves, 14.0 +/- 8.1 mmHg in size 21, and 10.1 +/- 8.9 mmHg in size 22. Four patients died in hospital (mortality rate 1.2%); there were no significant inter-group differences in hospital morbidity. Follow up was complete for 97% of the patients (mean 4.3 years; range: 1-11 years). The late mortality rate was 9.3% (n = 31; 2.56% per pt-yr); of these patients, 20 (14.3%) had size 20 prostheses, seven (7.3%) size 21, and four (4.2%) size 22 (p = 0.01). Fourteen patients (4.2%) died from cardiac causes, and six (1.8%) from prosthesis-related causes. Preoperative older age and aortic regurgitation were the only independent predictors of late mortality. Eight patients had systemic thromboembolic events (0.66% per pt-yr), two had prosthetic thrombosis (0.17% per pt-yr) and five had hemorrhagic episodes (0.41% per pt-yr). Seven patients had prosthetic valve endocarditis (0.58% per pt-yr). Among survivors, 97% are currently in NYHA class I/II. Conclusions: The small (size 20, 21 and 22) Medtronic Hall prostheses have good hemodynamic performance and are an excellent option as valve substitutes in patients with narrow aortic roots.
引用
收藏
页码:680 / 686
页数:7
相关论文
共 50 条
  • [21] THE MEDTRONIC-HALL VALVE - A DESIGN IN 1977 TO IMPROVE THE RESULTS OF VALVE-REPLACEMENT
    HALL, KV
    EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1992, 6 : S64 - S67
  • [22] A retrospective comparative study of aortic valve replacement with St. Jude Medical and Medtronic-Hall prostheses:: A 20-year follow-up study
    Anttila, V
    Heikkinen, J
    Biancari, F
    Oikari, K
    Pokela, R
    Lepojärvi, M
    Salmela, E
    Juvonen, T
    SCANDINAVIAN CARDIOVASCULAR JOURNAL, 2002, 36 (01) : 53 - 59
  • [23] Serial Changes of Hemodynamic Performance With Medtronic Hall Valve in Aortic Position
    Cho, Yang Hyun
    Jeong, Dong Seop
    Park, Pyo Won
    Park, Kay-Hyun
    Sung, Kiick
    Kim, Wook Sung
    Lee, Young Tak
    ANNALS OF THORACIC SURGERY, 2011, 91 (02): : 424 - 431
  • [24] The 20 mm Medtronic Hall prosthesis in the small aortic root
    Aris, A
    Ramirez, I
    Camara, ML
    Carreras, F
    Borras, X
    PonsLlado, G
    JOURNAL OF HEART VALVE DISEASE, 1996, 5 (04): : 459 - 462
  • [25] Aortic valve replacement with small sized pericardial bioprostheses - hemodynamic and clinical features
    Kurt, M.
    Litmathe, J.
    Boeken, U.
    Feindt, P.
    Gams, E.
    ZEITSCHRIFT FUR HERZ THORAX UND GEFASSCHIRURGIE, 2005, 19 (04): : 156 - 161
  • [26] Aortic valve replacement with Medtronic Freestyle Bioprosthesis: Operative technique and results
    Doty, JR
    Flores, JH
    Millar, RC
    Doty, DB
    JOURNAL OF CARDIAC SURGERY, 1998, 13 (03) : 208 - 217
  • [27] Concept of fracturing a small-sized bioprothesis for further valve in valve implantation
    Tanase, Daniel
    Georgiev, Stanimir
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2022, 99 (01) : E47 - E47
  • [28] Biological versus mechanical prostheses for aortic valve replacement
    Rodriguez-Caulo, Emiliano A.
    Blanco-Herrera, Oscar R.
    Berastegui, Elisabet
    Arias-Dachary, Javier
    Souaf-Khalafi, Souhayla
    Parody-Cuerda, Gertrudis
    Laguna, Gregorio
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2023, 165 (02): : 609 - +
  • [29] Recurrent ischemic strokes in a patient with Medtronic-Hall prosthetic aortic valve and valve strands
    Hutchinson, K
    Hafeez, F
    Woods, TD
    Chopra, PS
    Warner, TF
    Levine, RL
    Kamp, TJ
    JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 1998, 11 (07) : 755 - 757
  • [30] Serial Doppler echocardiographic evaluation of small-sized Sorin Bicarbon prostheses
    De Carlo, M
    Milano, AD
    Nardi, C
    Mecozzi, G
    Bortolotti, U
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2003, 126 (02): : 337 - 343