Improving broadband ultrasound attenuation assessment in cancellous bone by mitigating the influence of cortical bone: Phantom and in-vitro study

被引:5
|
作者
Tasinkevych, Yuriy [1 ]
Falinska, Katarzyna [1 ]
Lewin, Peter A. [2 ]
Litniewski, Jerzy [1 ]
机构
[1] Polish Acad Sci, Inst Fundamental Technol Res, Dept Ultrasound, Warsaw, Poland
[2] Drexel Univ, Philadelphia, PA 19104 USA
关键词
Broadband ultrasound attenuation; Correction of influence of cortical bone; Trabecular bone; X-RAY ABSORPTIOMETRY; TRABECULAR BONE; ELDERLY-WOMEN; HIP FRACTURE; SLOW WAVES; BACKSCATTER; TRANSMISSION; PARAMETERS; DENSITY; OSTEOPOROSIS;
D O I
10.1016/j.ultras.2018.06.018
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The purpose of this work was to present a new approach that allows the influence of cortical bone on non-invasive measurement of broadband ultrasound attenuation (BUA) to be corrected. The method, implemented here at 1 MHz makes use of backscattered signal and once refined and clinically confirmed, it would offer an alternative to ionizing radiation based methods, such as DEXA (Dual-energy X-ray absorptiometry), quantitative computed tomography (QCT), radiographic absorptiometry (RA) or single X-ray absorptiometry (SXA), which are clinically approved for assessment of progress of osteoporosis. In addition, as the method employs reflected waves, it might substantially enhance the applicability of BUA - from being suitable to peripheral bones only it would extend this applicability to include such embedded bones as hip and femoral neck. The proposed approach allows the cortical layer parameters used for correction and the corrected value and parameter of the cancellous bone (BUA) to be determined simultaneously from the single (pulse-echo) bone backscattered wave; to the best of the authors' knowledge such approach was not previously reported. The validity of the method was tested using acoustic data obtained from a custom-designed bone-mimicking phantom and a calf femur. The relative error of the attenuation coefficient assessment was determined to be 3.9% and 4.7% for the bone phantom and calf bone specimens, respectively. When the cortical shell influence was not taken into account the corresponding errors were considerably higher 8.3% (artificial bone) and 9.2% (calf femur). As indicated above, once clinically proven, the use of this BUA measurement technique in reflection mode would augment diagnostic power of the attending physician by permitting to include bones, which are not accessible for transmission mode evaluation, e.g. hip, spine, humerus and femoral neck.
引用
收藏
页码:382 / 390
页数:9
相关论文
共 46 条
  • [41] Influence of Exposure Parameters and Implant Position in Peri-Implant Bone Assessment in CBCT Images: An In Vitro Study
    Sawicki, Pawel
    Regulski, Piotr
    Winiarski, Artur
    Zawadzki, Pawel J.
    JOURNAL OF CLINICAL MEDICINE, 2022, 11 (13)
  • [42] Broadband ultrasound attenuation (BUA) of the heel bone and its correlates in men and women in the EPIC-Norfolk cohort: A cross-sectional population-based study
    Welch A.
    Camus J.
    Dalzell N.
    Oakes S.
    Reeve J.
    Khaw K.T.
    Osteoporosis International, 2004, 15 (3) : 217 - 225
  • [43] Use of embedded strain gages for the in-vitro study of proximal tibial cancellous bone deformation during knee flexion-extension movement: development, reproducibility and preliminary results of feasibility after frontal low femoral osteotomy
    Sobczak, Stephane
    Salvia, Patrick
    Dugailly, Pierre-Michel
    Lefevre, Philippe
    Feipel, Veronique
    Jan, Serge Van Sint
    Rooze, Marcel
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2011, 6
  • [44] Use of embedded strain gages for the in-vitro study of proximal tibial cancellous bone deformation during knee flexion-extension movement: development, reproducibility and preliminary results of feasibility after frontal low femoral osteotomy
    Stéphane Sobczak
    Patrick Salvia
    Pierre-Michel Dugailly
    Philippe Lefèvre
    Véronique Feipel
    Serge Van Sint Jan
    Marcel Rooze
    Journal of Orthopaedic Surgery and Research, 6
  • [45] Influence of Implant Geometry on Primary Insertion Stability and Simulated Pen-implant Bone Loss: An In Vitro Study Using Resonance Frequency Analysis and Damping Capacity Assessment
    Lachmann, Stefan
    Laval, Johannes Yves
    Axmann, Detlef
    Weber, Heiner
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2011, 26 (02) : 347 - 355
  • [46] Influence of age and gender in the assessment of inter-radicular and cortical bone thickness of the anterior maxilla and mandible for the placement of orthodontic mini-implants by cone-beam computed tomography: A retrospective study
    Bramhe, Pragati
    Bhoosreddy, Ajay
    Bhadage, Chetan
    Rathod, Prutha
    Chaudhari, Prajakta
    Utekar, Dipika
    JOURNAL OF THE INTERNATIONAL CLINICAL DENTAL RESEARCH ORGANIZATION, 2020, 12 (01) : 49 - 54