Improving broadband ultrasound attenuation assessment in cancellous bone by mitigating the influence of cortical bone: Phantom and in-vitro study

被引:5
|
作者
Tasinkevych, Yuriy [1 ]
Falinska, Katarzyna [1 ]
Lewin, Peter A. [2 ]
Litniewski, Jerzy [1 ]
机构
[1] Polish Acad Sci, Inst Fundamental Technol Res, Dept Ultrasound, Warsaw, Poland
[2] Drexel Univ, Philadelphia, PA 19104 USA
关键词
Broadband ultrasound attenuation; Correction of influence of cortical bone; Trabecular bone; X-RAY ABSORPTIOMETRY; TRABECULAR BONE; ELDERLY-WOMEN; HIP FRACTURE; SLOW WAVES; BACKSCATTER; TRANSMISSION; PARAMETERS; DENSITY; OSTEOPOROSIS;
D O I
10.1016/j.ultras.2018.06.018
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The purpose of this work was to present a new approach that allows the influence of cortical bone on non-invasive measurement of broadband ultrasound attenuation (BUA) to be corrected. The method, implemented here at 1 MHz makes use of backscattered signal and once refined and clinically confirmed, it would offer an alternative to ionizing radiation based methods, such as DEXA (Dual-energy X-ray absorptiometry), quantitative computed tomography (QCT), radiographic absorptiometry (RA) or single X-ray absorptiometry (SXA), which are clinically approved for assessment of progress of osteoporosis. In addition, as the method employs reflected waves, it might substantially enhance the applicability of BUA - from being suitable to peripheral bones only it would extend this applicability to include such embedded bones as hip and femoral neck. The proposed approach allows the cortical layer parameters used for correction and the corrected value and parameter of the cancellous bone (BUA) to be determined simultaneously from the single (pulse-echo) bone backscattered wave; to the best of the authors' knowledge such approach was not previously reported. The validity of the method was tested using acoustic data obtained from a custom-designed bone-mimicking phantom and a calf femur. The relative error of the attenuation coefficient assessment was determined to be 3.9% and 4.7% for the bone phantom and calf bone specimens, respectively. When the cortical shell influence was not taken into account the corresponding errors were considerably higher 8.3% (artificial bone) and 9.2% (calf femur). As indicated above, once clinically proven, the use of this BUA measurement technique in reflection mode would augment diagnostic power of the attending physician by permitting to include bones, which are not accessible for transmission mode evaluation, e.g. hip, spine, humerus and femoral neck.
引用
收藏
页码:382 / 390
页数:9
相关论文
共 46 条
  • [31] Effect of Gaussian smoothing filter for CT-based attenuation correction mapon quantitative assessment in bone SPECT/CT: a phantom study
    Tomita, Yoya
    Ichikawa, Yasutaka
    Hashizume, Kengo
    Kamigiri, Akira
    Yamada, Tsuyoshi
    Maki, Hiroaki
    Sakuma, Hajime
    JOURNAL OF NUCLEAR MEDICINE, 2018, 59
  • [32] Effect of Gaussian Smoothing Filter Size for CT-Based Attenuation Correction on Quantitative Assessment of Bone SPECT/CT: A Phantom Study
    Yoya Tomita
    Yasutaka Ichikawa
    Kengo Hashizume
    Hajime Sakuma
    Journal of Digital Imaging, 2023, 36 (5) : 2313 - 2321
  • [33] Effect of Gaussian Smoothing Filter Size for CT-Based Attenuation Correction on Quantitative Assessment of Bone SPECT/CT: A Phantom Study
    Tomita, Yoya
    Ichikawa, Yasutaka
    Hashizume, Kengo
    Sakuma, Hajime
    JOURNAL OF DIGITAL IMAGING, 2023, 36 (05) : 2313 - 2321
  • [34] ASSESSMENT OF BONE MINERAL DENSITY: A COMPARATIVE PILOT STUDY WITH PHANTOM LESS QUANTITATIVE COMPUTED TOMOGRAPHY (QCT) AND QUANTITATIVE ULTRASOUND (QUS)
    David, L. R.
    OSTEOPOROSIS INTERNATIONAL, 2014, 25 : S41 - S41
  • [35] Influence of cutting force on temperature, microcracks and chip morphology during rotary ultrasonic bone drilling: An in-vitro study
    Raj Agarwal
    Ravinder Pal Singh
    Vishal Gupta
    Jaskaran Singh
    Journal of the Brazilian Society of Mechanical Sciences and Engineering, 2022, 44
  • [36] Influence of cutting force on temperature, microcracks and chip morphology during rotary ultrasonic bone drilling: An in-vitro study
    Agarwal, Raj
    Singh, Ravinder Pal
    Gupta, Vishal
    Singh, Jaskaran
    JOURNAL OF THE BRAZILIAN SOCIETY OF MECHANICAL SCIENCES AND ENGINEERING, 2022, 44 (07)
  • [37] ARTHROSCOPIC ULTRASOUND TECHNIQUE FOR SIMULTANEOUS QUANTITATIVE ASSESSMENT OF ARTICULAR CARTILAGE AND SUBCHONDRAL BONE: AN IN VITRO AND IN VIVO FEASIBILITY STUDY
    Liukkonen, J.
    Hirvasniemi, J.
    Joukainen, A.
    Penttila, P.
    Viren, T.
    Saarakkala, S.
    Kroger, H.
    Jurvelin, J. S.
    Toyras, J.
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2013, 39 (08): : 1460 - 1468
  • [38] Low grade inflammation is associated with lower velocity of sound and broadband ultrasound attenuation in older men, but not with bone loss or fracture risk in a longitudinal aging study
    den Uyl, Debby
    van Schoor, Natasja M.
    Bravenboer, Nathalie
    Lips, Paul
    Lems, Willem F.
    BONE, 2015, 81 : 270 - 276
  • [39] Influence of chitosan and hydroxyapatite incorporation on properties of electrospun PVA/HA nanofibrous mats for bone tissue regeneration: Nanofibers optimization and in-vitro assessment
    Salim, Samar A.
    Loutfy, Samah A.
    El-Fakharany, Esmail M.
    Taha, Tarek H.
    Hussien, Yasmen
    Kamoun, Elbadawy A.
    JOURNAL OF DRUG DELIVERY SCIENCE AND TECHNOLOGY, 2021, 62
  • [40] Assessment of the effect of strontium, lead, and aluminum in bone on dual-energy x-ray absorptiometry and quantitative ultrasound measurements: A phantom study
    Jang, Deok Hyun
    Da Silva, Eric
    Tavakkoli, Jahan
    Slatkovska, Lubomira
    Cheung, Angela M.
    Pejovic-Milic, Ana
    MEDICAL PHYSICS, 2018, 45 (01) : 81 - 91