The influence of the dose assessment method on the LET dependence of the relative luminescence efficiency of LiF:Mg,Ti and LiF:Mg,Cu,P

被引:19
|
作者
Parisi, Alessio [1 ,2 ]
Van Hoey, Olivier [1 ]
Megret, Patrice [2 ]
Vanhavere, Filip [1 ]
机构
[1] CEN, Belgian Nucl Res Ctr SCK, Boeretang 200, Mol, Belgium
[2] Univ Mons, Fac Engn, Blvd Dolez 31, Mons, Belgium
关键词
LiF:Mg; Ti; Cu; P; TLD; Efficiency; LET; TEMPERATURE RATIO HTR; IONIZATION DENSITY; GLOW-CURVE; THERMOLUMINESCENT EFFICIENCY; COSMIC-RADIATION; LIF DETECTORS; ENERGY IONS; LIF-MG; TI; TL;
D O I
10.1016/j.radmeas.2017.02.001
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Thermoluminescent detectors (TLDs) are solid state detectors commonly used for dose monitoring and dose mapping in many earth and space applications. Among all the crystals available, the ones based on lithium fluoride (LiF) are the most widely spread and studied around the world. In the last years the use of these TLDs in complex heavy charged particles (HCPs) fields, as in spacecraft or in radiotherapy beams, has grown rapidly leading to the necessity of an accurate characterization of the detectors' response in these radiation environments. In this work, (LiF)-Li-6:Mg,Ti, (LiF)-Li-7:Mg,Ti, (LiF)-Li-6:Mg,Cu,P and (LiF)-Li-7:Mg,Cu,P detectors were exposed to different radiation qualities in order to investigate the dependence of their relative luminescence efficiency on the linear energy transfer (LET) of the incident particles. The measured efficiencies were in good agreement with literature. In relation to this issue, different dose assessment methods, i.e. peak height, signal integration over different regions of interest (ROIs) and glow curve deconvolution, were compared and their influence on the measured relative luminescence efficiency values was assessed. For both LiF:Mg,Ti and LiF:Mg,Cu,P detectors, no relevant differences were found in the relative luminescence efficiency evaluated by means of main peak height, main peak ROI and main peak deconvolution. On the other hand, if the integration of the signal is extended to the high temperature zone (248-310 degrees C) of LiF:Mg,Ti or the low temperature zone (125-185 degrees C) of LiF:Mg,Cu,P, significant deviations were found. (C) 2017 Elsevier Ltd. All rights reserved.
引用
收藏
页码:34 / 40
页数:7
相关论文
共 50 条
  • [31] Sensitivity changes of LiF:Mg,Ti and LiF:Mg,Cu,P TL detectors after proton exposures
    Sadel, M.
    Bilski, P.
    Swakon, J.
    Rydygier, M.
    RADIATION MEASUREMENTS, 2015, 74 : 26 - 30
  • [32] Comparison between the low temperature thermoluminescence spectra in annealed LiF:Mg,Cu, LiF:Mg,Cu,P and LiF:Mg,Cu,Si
    Yang, B.
    Wang, L.
    Townsend, P. D.
    Gao, H.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2008, 266 (11): : 2581 - 2586
  • [33] Optical and luminescence properties of single crystals of LiF:Cu and LiF:Mg, Cu
    Shalaev A.A.
    Bobina N.S.
    Paklin A.S.
    Shendrik R.Y.
    Nepomnyashchikh A.I.
    Bulletin of the Russian Academy of Sciences: Physics, 2015, 79 (2) : 263 - 266
  • [34] Sensitivity loss and recovery for individual TL peaks in LiF:Mg,Ti and LiF:Mg,Cu,P after high-dose irradiation
    Bilski, P.
    Cybulski, T.
    Puchalska, M.
    Ptaszkiewicz, M.
    RADIATION MEASUREMENTS, 2008, 43 (2-6) : 357 - 360
  • [35] Proton irradiation of 7LiF:Mg,Ti thermoluminescent detectors:: influence of dopant concentration on dose response and let dependence of TL efficiency
    Loncol, T
    Denis, JM
    Vynckier, S
    Scalliet, P
    Bilski, P
    Budzanowski, M
    Waligorski, MPR
    Olko, P
    RADIATION PROTECTION DOSIMETRY, 1999, 84 (1-4) : 179 - 184
  • [36] Comparative studies of glow peaks and kinetic trapping parameters of LiF(Mg,Cu), LiF(Mg,Cu,P), LiF(Mg,Cu,Si) and LiF(Mg,Cu,Na,Si) single crystals
    Laopaiboon, R.
    Bootjomchai, C.
    Pencharee, S.
    Laopaiboon, J.
    RADIATION EFFECTS AND DEFECTS IN SOLIDS, 2014, 169 (02): : 129 - 136
  • [37] The average distance between Mg-based trapping structures in LiF:Mg,Ti and LiF:Mg,Cu,P and the relevance to microdosimetry
    Horowitz, YS
    RADIATION PROTECTION DOSIMETRY, 1999, 82 (01) : 51 - 54
  • [38] Glow curve analysis of long-term stability of LiF:Mg,Cu,P as compared to LiF:Mg,Ti
    Duggan, L
    Kron, T
    RADIATION PROTECTION DOSIMETRY, 1999, 85 (1-4) : 213 - 216
  • [39] Behaviour of LiF:Mg,Cu,P and LiF:Mg,Ti thermoluminescent detectors for electron doses up to 1 MGy
    Bilski, Pawel
    Obryk, Barbara
    Stuglik, Zofia
    RADIATION MEASUREMENTS, 2010, 45 (3-6) : 576 - 578
  • [40] UV-INDUCED BLEACHING OF DEEP TRAPS IN HARSHAW TLD LiF:Mg,Cu,P and LiF:Mg,Ti
    Benevides, L.
    Voss, S.
    Nita, I.
    Rotunda, J.
    Velbeck, K.
    Luo, L. Z.
    Moscovitch, M.
    RADIATION PROTECTION DOSIMETRY, 2011, 144 (1-4) : 199 - 201