In-situ sensitivity of a totally-implantable microphone

被引:2
|
作者
D'hondt, Christiane [1 ,2 ]
Wouters, Jan [1 ]
Verhaert, Nicolas [1 ,3 ]
机构
[1] Univ Leuven, Dept Neurosci, Res Grp ExpORL, KU Leuven, Onderwijs Navorsing 2,Herestr 49,Bus 721, B-3000 Leuven, Belgium
[2] Cochlear Technol Ctr Belgium, Mechelen, Belgium
[3] Univ Hosp Leuven, Dept Otolaryngol Head & Neck Surg, Leuven, Belgium
关键词
middle ear implant; fully-implantable middle ear implant; subcutaneous microphone; microphone sensitivity; HEARING-AIDS; THICKNESS; SYSTEM;
D O I
10.1016/j.heares.2020.108018
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
One of the key components of fully-implantable hearing devices is the implantable microphone. A crucial parameter when characterizing implantable microphones is the acoustic sensitivity, as it is one of the input parameters for the fitting algorithm and influences the achievable gain. The aim of our study was to investigate the sensitivity of an implanted subcutaneous microphone over time to answer two research questions: (1) How does the sensitivity change once the microphone is implanted under the skin (pre-op versus in-situ)? and (2) How does the sensitivity change from short-term to mid-term? We have measured the in-situ microphone sensitivity in three subjects implanted with a fully-implantable active middle ear implant from 2 weeks up to 20 weeks after implantation with a research software. The microphone sensitivity changed after implantation with clinically relevant changes around the resonance frequency of the microphone. Based on our results, it is essential to measure the in-situ microphone sensitivity at the time of initial fitting of the implant. Once implanted no clinically relevant changes in the microphone sensitivity were observed over time, with a clear decrease in variability over time. (C) 2020 Elsevier B.V. All rights reserved.
引用
收藏
页数:10
相关论文
共 50 条
  • [21] The totally implantable hearing aid
    Zenner, HP
    FORTSCHRITT UND FORTBILDUNG IN DER MEDIZIN, BAND 25 (2001/2002), 2001, 25 : 251 - +
  • [22] Implantable microphone with acoustic tube for fully implantable hearing devices
    Jung, Eui Sung
    Seong, Ki Woong
    Lim, Hyung Gyu
    Lee, Jyung Hyun
    Cho, Jin Ho
    IEICE ELECTRONICS EXPRESS, 2011, 8 (04): : 215 - 219
  • [23] An implantable microphone for electronic hearing implants
    Leysieffer, H
    Muller, G
    Zenner, HP
    HNO, 1997, 45 (10) : 816 - 827
  • [24] The Frequency Response of a Floating Piezoelectric Microphone for the Implantable Middle Ear Microphone
    Gao, Na
    Chen, Yong Zheng
    Chi, Fang Lu
    Zhang, Tian Yu
    Xu, Hai Dong
    Kang, Hou Yong
    Pan, Tie Zheng
    LARYNGOSCOPE, 2013, 123 (06): : 1506 - 1513
  • [25] In-situ impedance and absorption coefficient measurements using a double-layer microphone array
    Hald, Jorgen
    Song, Wookeun
    Haddad, Karim
    Jeong, Cheol-Ho
    Richard, Antoine
    APPLIED ACOUSTICS, 2019, 143 : 74 - 83
  • [26] MICROPHONE SENSITIVITY RATINGS
    LORONA, A
    AUDIO, 1976, 60 (12): : 32 - &
  • [27] An Implantable Umbo Microphone For Fully-Implantable Assistive Hearing Devices
    Cary, Benjamin G.
    Zhang, John Z.
    McHugh, Christopher, I
    Kymissis, Ioannis
    Olson, Elizabeth S.
    Nakajima, Hideko Heidi
    Lang, Jeffrey H.
    2021 IEEE SENSORS, 2021,
  • [28] Calculation of the sensitivity of the in-situ seismograph using the collocation test
    Xu WeiWei
    Yang DaKe
    Yuan SongYong
    Ma JieMei
    Xie JianBo
    Li XiaoJun
    CHINESE JOURNAL OF GEOPHYSICS-CHINESE EDITION, 2021, 64 (12): : 4572 - 4581
  • [29] In-situ closure studies on scattering and backscattering coefficients and their sensitivity
    Wex, H.
    Koziar, C.
    Neusuess, C.
    Wiedensohler, A.
    Journal of Aerosol Science, 1999, 30 (Suppl. 1)
  • [30] Development of a totally implantable hearing aid
    Vujanic, A
    Pavelka, R
    Adamovic, N
    Kment, C
    Mitic, S
    Brenner, W
    Popovic, G
    2002 23RD INTERNATIONAL CONFERENCE ON MICROELECTRONICS, VOLS 1 AND 2, PROCEEDINGS, 2002, : 235 - 238