Micromotion of implant-abutment interfaces (IAI) after loading: correlation of finite element analysis with in vitro performances

被引:15
|
作者
Li, Zhi [1 ,2 ]
Gao, Shanshan [1 ]
Chen, Hongyu [1 ]
Ma, Ruiyang [1 ]
Wu, Tingting [1 ]
Yu, Haiyang [1 ]
机构
[1] Sichuan Univ, State Key Lab Oral Dis, Natl Clin Res Ctr Oral Dis, West China Hosp Stomatol, Chengdu, Sichuan, Peoples R China
[2] Aerosp Ctr Hosp, Beijing, Peoples R China
基金
中国国家自然科学基金;
关键词
Dental implants; Finite element analysis; Microleakage; Micromotion; Locked condition; TITANIUM ABUTMENTS; BACTERIAL LEAKAGE; DENTAL IMPLANT; REMOVAL FORCE; CONNECTIONS; ZIRCONIA; COLONIZATION; MICROLEAKAGE; PARAMETERS; COMPONENTS;
D O I
10.1007/s11517-018-1937-6
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Micromotion between IAI affects long-term survival rate of dental implants. The use of practical implants for mechanical test is costly. Finite element analysis (FEA) could test the micron level deformation changes, but whether it reflects the in vitro mechanical performances remains unknown. This study aims to investigate the correlation between IAI micromotion of FEA and in vitro performances. The two-step-two-component FEA method was used to test the relative deformation between IAI for three implant designs (M1, M2, and M3) during torque loading and cyclic oblique loading. The micromotion was divided into directions that perpendicular to (x-axis) and parallel to (y-axis) IAI. In vitro experiments on the micromotion relevant performance of IAI microleakage (tested by toluidine blue releasing with a spectrometer) and IAI locked condition (tested by abutment removal force tests after detaching the central screws) were also conducted for the identical implant systems (G1, G2, and G3). One-way ANOVA and Pearson's correlation tests were performed for data analysis. FEA illustrated that the three implant systems performed different micromotion patterns. Significant differences were found in the IAI microleakage and removal force among the groups. Positive correlations were found between FEA and in vitro outcomes. Therefore, the two-step-two-component FEA method is an appropriate method to evaluate the IAI micromotion after loading.
引用
收藏
页码:1133 / 1144
页数:12
相关论文
共 50 条
  • [41] Effect of Different Implant-Abutment Mismatches on Stress Distribution: A 3-Dimensional Finite Element Study
    Esfahanian, Vahid
    Khosravi, Armin
    Ketabi, Mohammad
    Deporter, Douglas
    JOURNAL OF ORAL IMPLANTOLOGY, 2022, 48 (05) : 370 - 374
  • [42] Evaluation of the mechanical characteristics of the implant-abutment complex of a reduced-diameter morse-taper implant -: A nonlinear finite element stress analysis
    Akça, K
    Çehreli, MC
    Iplikçioglu, H
    CLINICAL ORAL IMPLANTS RESEARCH, 2003, 14 (04) : 444 - 454
  • [43] Comparative Evaluation of Three Abutment-Implant Interfaces on Stress Distribution in and Around Different Implant Systems: A Finite Element Analysis
    Kharsan, Vishwas
    Bandgar, Vaibhav
    Mirza, Aftab
    Jagtiani, Kiran
    Dhariwal, Neha
    Kore, Rahul
    CONTEMPORARY CLINICAL DENTISTRY, 2019, 10 (04) : 590 - 594
  • [44] A conical implant-abutment interface at the level of the marginal bone improves the distribution of stresses in the supporting bone - An axisymmetric finite element analysis
    Hansson, S
    CLINICAL ORAL IMPLANTS RESEARCH, 2003, 14 (03) : 286 - 293
  • [45] Effects of Marginal Bone Loss Progression on Stress Distribution in Different Implant-Abutment Connections and Abutment Materials: A 3D Finite Element Analysis Study
    Lin, Ching-Ping
    Shyu, Yi-Ting
    Wu, Yu-Ling
    Tsai, Ming-Hsu
    Chen, Hung-Shyong
    Wu, Aaron Yu-Jen
    MATERIALS, 2022, 15 (17)
  • [46] Pull-off resistance of a screwless implant-abutment connection and surface evaluation after cyclic loading
    Alevizakos, Vasilios
    Mosch, Richard
    Mitov, Gergo
    Othman, Ahmed
    von See, Constantin
    JOURNAL OF ADVANCED PROSTHODONTICS, 2021, 13 (03): : 152 - 159
  • [47] Influence of the interface design on the yield force of the implant-abutment complex before and after cyclic mechanical loading
    Dittmer, Marc Philipp
    Dittmer, Stephanie
    Borchers, Lothar
    Kohorst, Philipp
    Stiesch, Meike
    JOURNAL OF PROSTHODONTIC RESEARCH, 2012, 56 (01) : 19 - 24
  • [48] Effects of Implant-Abutment Connection Type and Inter-Implant Distance on Inter-Implant Bone Stress and Microgap: Three-Dimensional Finite Element Analysis
    Matsuoka, Takashi
    Nakano, Tamaki
    Yamaguchi, Satoshi
    Ono, Shinji
    Watanabe, Shota
    Sato, Takumi
    Yatani, Hirofumi
    MATERIALS, 2021, 14 (09)
  • [49] The influence of platform switching on the biomechanical aspects of the implant-abutment system. A three dimensional finite element study
    Canullo, Luigi
    Pace, Francesco
    Coelho, Paulo
    Sciubba, Enrico
    Vozza, Iole
    MEDICINA ORAL PATOLOGIA ORAL Y CIRUGIA BUCAL, 2011, 16 (06): : E852 - E856
  • [50] In Vitro Evaluation of Bacterial Leakage Along the Implant-Abutment Interface of an External-Hex Implant After Saliva Incubation
    do Nascimento, Cassio
    Miani, Paola Kirsten
    Watanabe, Evandro
    Pedrazzi, Vinicius
    de Albuquerque, Rubens Ferreira, Jr.
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2011, 26 (04) : 782 - 787