Evaluation of a lithium formate EPR dosimetry system for dose measurements around 192Ir brachytherapy sources

被引:29
|
作者
Antonovic, Laura [1 ]
Gustafsson, Hakan [1 ]
Carlsson, Gudrun Alm [1 ]
Tedgren, Asa Carlsson [1 ]
机构
[1] Linkoping Univ, Dept Med & Hlth Sci IMH, Fac Hlth Sci, SE-58185 Linkoping, Sweden
关键词
brachytherapy; dosimetry; organic compounds; paramagnetic resonance; phantoms; thermoluminescent dosimeters; MONTE-CARLO DOSIMETRY; DISTRIBUTIONS; ENERGY; TLD;
D O I
10.1118/1.3110068
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A dosimetry system using lithium formate monohydrate (HCO2Li center dot H2O) as detector material and electron paramagnetic resonance (EPR) spectroscopy for readout has been used to measure absorbed dose distributions around clinical Ir-192 sources. Cylindrical tablets with diameter of 4.5 mm, height of 4.8 mm, and density of 1.26 g/cm(3) were manufactured. Homogeneity test and calibration of the dosimeters were performed in a 6 MV photon beam. Ir-192 irradiations were performed in a PMMA phantom using two different source models, the GammaMed Plus HDR and the microSelectron PDR-v1 model. Measured absorbed doses to water in the PMMA phantom were converted to the corresponding absorbed doses to water in water phantoms of dimensions used by the treatment planning systems (TPSs) using correction factors explicitly derived for this experiment. Experimentally determined absorbed doses agreed with the absorbed doses to water calculated by the TPS to within +/- 2.9%. Relative standard uncertainties in the experimentally determined absorbed doses were estimated to be within the range of 1.7%-1.3% depending on the radial distance from the source, the type of source (HDR or PDR), and the particular absorbed doses used. This work shows that a lithium formate dosimetry system is well suited for measurements of absorbed dose to water around clinical HDR and PDR Ir-192 sources. Being less energy dependent than the commonly used thermoluminescent lithium fluoride (LiF) dosimeters, lithium formate monohydrate dosimeters are well suited to measure absorbed doses in situations where the energy dependence cannot easily be accounted for such as in multiple-source irradiations to verify treatment plans. Their wide dynamic range and linear dose response over the dose interval of 0.2-1000 Gy make them suitable for measurements on sources of the strengths used in clinical applications. The dosimeter size needs, however, to be reduced for application to single-source dosimetry.
引用
收藏
页码:2236 / 2247
页数:12
相关论文
共 50 条
  • [1] Dosimetry of 192Ir sources used for endovascular brachytherapy
    Reynaert, N
    Van Eijkeren, M
    Taeymans, Y
    Thierens, H
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (02): : 499 - 516
  • [2] A standard graphite calorimeter for dosimetry in brachytherapy with high dose rate 192Ir sources
    Guerra, A. S.
    Loreti, S.
    Pimpinella, M.
    Quini, M.
    D'Arienzo, M.
    Astefanoaei, I.
    Caporali, C.
    Bolzan, C.
    Pagliari, M.
    [J]. METROLOGIA, 2012, 49 (05) : S179 - S183
  • [3] Radiochromic Film Based Reference Dosimetry System for 192Ir Brachytherapy Sources
    Aldelaijan, S.
    Mohammed, H.
    Tomic, N.
    Liang, L.
    Deblois, F.
    Sarfehnia, A.
    Abdel-Rahman, W.
    Seuntjens, J.
    Devic, S.
    [J]. MEDICAL PHYSICS, 2011, 38 (06) : 3574 - +
  • [4] INVESTIGATING LITHIUM FORMATE EPR DOSIMETRY FOR ABSOLUTE DOSE MEASUREMENTS AROUND 1921R SOURCES
    Antonovic, L.
    Gustafsson, H.
    Lund, E.
    Carlsson, G. Alm
    Tedgren, Carlsson
    [J]. RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S400 - S400
  • [5] A standard dosimetry procedure for 192Ir sources used for endovascular brachytherapy
    Piermattei, A
    Fidanzio, A
    Azario, L
    Russo, A
    Perrone, F
    Capote, R
    Toni, MP
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (23): : 4205 - 4221
  • [6] Determination of Dose at the Reference Point for 192Ir HDR Brachytherapy Sources Using Fricke Dosimetry
    Austerlitz, Carlos
    Benhabib, Sidi
    Zhu, Dengsong
    Mota, Helvecio
    Campos, Diana
    Allison, Ron
    Sibina, Claudio
    [J]. AMERICAN JOURNAL OF CLINICAL ONCOLOGY-CANCER CLINICAL TRIALS, 2008, 31 (06): : 617 - 617
  • [7] Dosimetry comparison of 192Ir Sources
    Papagiannis, P
    Angelopoulos, A
    Pantelis, E
    Sakelliou, L
    Baltas, D
    Karaiskos, P
    Sandilos, P
    Vlachos, L
    [J]. MEDICAL PHYSICS, 2002, 29 (10) : 2239 - 2246
  • [8] A microDiamondfor determination of absorbed dose around high-dose-rate 192Ir brachytherapy sources
    Kaveckyte, V.
    Malusek, A.
    Benmaklouf, H.
    Carlsson, G. Alm
    Tedgren, A. Carlsson
    [J]. RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S407 - S407
  • [9] Dose rate calculations around 192Ir brachytherapy sources using a Sievert integration model
    Karaiskos, P
    Angelopoulos, A
    Baras, P
    Rozaki-Mavrouli, H
    Sandilos, P
    Vlachos, L
    Sakelliou, L
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (02): : 383 - 398
  • [10] Patient effective dose from endovascular brachytherapy with 192Ir sources
    Perna, L
    Bianchi, C
    Nicolini, G
    Novario, R
    Tanzi, F
    Conte, L
    [J]. RADIATION PROTECTION DOSIMETRY, 2002, 102 (02) : 121 - 128