Verification of the Elekta Monaco TPS Monte Carlo in modelling radiation transmission through metals in a water equivalent phantom

被引:5
|
作者
Byrnes, Kurt [1 ]
Ford, Andriana [1 ]
Bennie, Nick [1 ]
机构
[1] Northern NSW Local Hlth Dist, Lismore Base Hosp, NCCI, Lismore, NSW 2480, Australia
关键词
Monaco; Monte Carlo; MCP96; Metals; Hip prosthetics; HIP PROSTHESES; THERAPY;
D O I
10.1007/s13246-019-00749-2
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Many studies have performed dosimetric studies using various metal implants however these are difficult to translate to other implants of a different geometry or material (Rijken and Colyer, J Appl Clin Med Phys 18:5:301-306, 2017; Ade and du Plessis, J Appl Clin Med Phys 18:5:162-173, 2017; Prabhakar et al. Rep Pract Oncol Radiother 18:209-213, 2013; Ng et al. Rep Pract Oncol Radiother 20:273-277, 2015; Reft et al. Med Phys 30:1162-1182, 2003; Sasaki et al., Nihon Hoshasen Gijutsu Gakkai Zasshi 72(9):735-745, 2016). In this study, the ability of the Monaco Monte Carlo algorithm (Elekta AB, Stockholm, Sweden) to model radiation transport through different types of metals was evaluated. Investigation of the capabilities and limitations of the algorithm is required for the potential use of Monaco for planning radiotherapy treatments when avoidance of metal implants is clinically undesirable. A MapCHECK 2 diode array (Sun Nuclear Corp, Melbourne, USA) and a PTW 30013 Farmer chamber was used to measure the dose at depth, downstream of 1cmx5cmx5cm metal blocks of three known compositions; stainless steel, aluminium and MCP96. The setup was imaged using a CT scanner and imported into the Monaco TPS where the beam arrangement was replicated. The density of the metals was overridden using the known electron density of each (IMPAC Medical Systems Inc, Monaco dose calculation technical reference. IMPAC Medical Systems, Sunnydale, CA, 2013). The differences between the dose measured using the ion chamber and calculated using Monaco downstream of the 1cm metal blocks were respectively: - 1.2%, - 2.2% and 9.5% when irradiated using a 6 MV beam, and -0.9%, - 1.3% and 14%, when irradiated using a 15 MV beam. This was then repeated using 2cm and 3cm of each metal type giving similar results for aluminium and stainless steel and increased discrepancy for MCP96. Discrepancies between treatment planning software and measurements at depth have been shown to give uncertainties between 5 and 23% in previous studies (Rijken and Colyer, J Appl Clin Med Phys 18:5:301-306, 2017; Ade and du Plessis, J Appl Clin Med Phys 18:5:162-173, 2017; Prabhakar et al. Rep Pract Oncol Radiother 18:209-213, 2013; Ng et al. Rep Pract Oncol Radiother 20:273-277, 2015; Reft et al. Med Phys 30:1162-1182, 2003; Sasaki et al., Nihon Hoshasen Gijutsu Gakkai Zasshi 72(9):735-745, 2016). This study uses basic shapes providing results that remove the uncertainties in geometry and can therefore be applied to any shape. This will help determine whether errors in dose calculations are due to the TPS particle transport algorithms or due to other effects, such as inaccurate contouring or incorrect densities. Thus giving the planner an additional degree of freedom in their planning and decision making process.
引用
收藏
页码:639 / 645
页数:7
相关论文
共 17 条
  • [1] Verification of the Elekta Monaco TPS Monte Carlo in modelling radiation transmission through metals in a water equivalent phantom
    Kurt Byrnes
    Andriana Ford
    Nick Bennie
    [J]. Australasian Physical & Engineering Sciences in Medicine, 2019, 42 : 639 - 645
  • [2] Modelling the MLC Agility transmission radiation through the closed leaf gap in the Monaco TPS
    Vicedo Gonzalez, M. A.
    Garcia Hernandez, M. T.
    Planes Meseguer, D.
    Garcia Molla, R.
    Borderia Navarro, B.
    Calatayud Cuesta, P.
    Alcaraz Lozano, M. L.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2022, 170 : S1296 - S1297
  • [3] Verification of the physics model of the TPS for asymmetric fields through Monte Carlo simulations
    Brovchuk, S.
    Cora, S.
    Francescon, P.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S997 - S998
  • [4] FURTHER DOSIMETRIC SPECIFICATION AND MONTE CARLO MODELLING OF A NEW ELEKTA RADIATION HEAD WITH INTEGRATED 160-LEAF MLC
    Thompson, C.
    Christofides, D.
    Evans, C.
    Weston, S.
    Cosgrove, V.
    Thwaites, D. I.
    Thompson, G.
    Brown, K.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S555 - S555
  • [5] Description and dosimetric verification of the PEREGRINE Monte Carlo dose calculation system for photon beams incident on a water phantom
    Siantar, CLH
    Walling, RS
    Daly, TP
    Faddegon, B
    Albright, N
    Bergstrom, P
    Chuang, C
    Garrett, D
    House, RK
    Knapp, D
    Wieczorek, DJ
    Verhey, LJ
    [J]. MEDICAL PHYSICS, 2001, 28 (07) : 1322 - 1337
  • [6] Verification of water-equivalent model for calculation of multiple scattering effects in simplified Monte Carlo dose calculation
    Kohno, R
    Takada, Y
    Sakae, T
    Terunuma, T
    Matsumoto, K
    Nohtomi, A
    Matsuda, H
    [J]. JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 2003, 42 (6A): : 3728 - 3729
  • [7] New insights into photodynamic therapy treatment through the use of 3D Monte Carlo radiation transfer modelling.
    Campbell, C. Louise
    Wood, Kenneth
    Brown, C. Tom A.
    Moseley, Harry
    [J]. PHOTONIC THERAPEUTICS AND DIAGNOSTICS XII, 2016, 9689
  • [8] Effect of boiler feed water composition on inorganic scaling in once-through steam generators estimated using a Monte Carlo modelling approach
    Klyukin, Yury I.
    Mayer, Bernhard
    Tutolo, Benjamin
    [J]. APPLIED GEOCHEMISTRY, 2023, 148
  • [9] Bottled water safety evaluation: A comprehensive health risk assessment of oral exposure to heavy metals through deterministic and probabilistic approaches by Monte Carlo simulation
    Askari, Masoomeh
    Soleimani, Hamed
    Nalosi, Kamal Babakrpur
    Saeedi, Reza
    Abolli, Samaneh
    Ghani, Maryam
    Abtahi, Mehrnoosh
    Alimohammadi, Mahmood
    [J]. FOOD AND CHEMICAL TOXICOLOGY, 2024, 185
  • [10] COMPARISON OF RADIATION-DOSIMETRY IN WATER AND IN SOLID PHANTOM MATERIALS FOR I-125 AND PD-103 BRACHYTHERAPY SOURCES - EGS4 MONTE-CARLO STUDY
    LUXTON, G
    [J]. MEDICAL PHYSICS, 1994, 21 (05) : 631 - 641