COMPARISON OF RADIATION-DOSIMETRY IN WATER AND IN SOLID PHANTOM MATERIALS FOR I-125 AND PD-103 BRACHYTHERAPY SOURCES - EGS4 MONTE-CARLO STUDY

被引:55
|
作者
LUXTON, G
机构
[1] University of Southern California School of Medicine, Los Angeles, California
关键词
BRACHYTHERAPY; RADIOISOTOPES; MONTE-CARLO; RADIATION DOSIMETRY; I-125; PD-103;
D O I
10.1118/1.597317
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A model based on the EGS4 Monte Carlo system is developed for calculating dose fate to water from an embedded low-energy brachytherapy source, given measurement data of dose rate to water within a water-substitute solid phantom for a source of given strength. The EGS4-based model is used to calculate point source dose fate distributions per unit source strength for water and for several species of solid phantoms. As a model for the dose rate distribution to an infinitesimal water inhomogeneity embedded in the solid phantom, the Monte Carlo is used to calculate dose rate to a thin spherical shell of water contained within the solid phantom at various distances centered on the source. Comparing this distribution to that calculated for a homogeneous water phantom, one can derive correction factors to determine dose rate to homogeneous water medium from measurements with water dose-calibrated detectors within the solid phantom. These factors are calculated for polymethylmethacrylate (PMMA or acrylic), solid water (WT1), and RW-1, a material optimized for low-energy dosimetry, with photon spectra from Pd-103 and from two commercial models of I-125 seed used as input. For model 6711 I-125 seeds at 1 cm in PMMA and WT1, the calculated ratios of dose rate to water to dose rate to water in the solid phantom are 0.893 and 1.038, respectively. Applying these factors to published, apparently divergent measurements of the specific dose rate constant based on measurements in PMMA and WT1 phantoms, the specific dose rate constant inferred for the model 6711 seed are found to be in excellent agreement: 0.878 cGy h-1U-1 (1.115 cGy- cm2h-1mCi-1) from the mean of the WT1 measurements and 0.879 cGy h-1U-1 from the PMMA measurements. Among the phantoms studied for Pd-103 and 1-125, measurements in RW-1 would most closely approximate those for water.
引用
收藏
页码:631 / 641
页数:11
相关论文
共 15 条
  • [2] RADIATION-DOSIMETRY FOR I-125 USING MONTE-CARLO SIMULATION METHOD
    CHIUTSAO, ST
    OBRIEN, K
    SANNA, R
    [J]. MEDICAL PHYSICS, 1983, 10 (04) : 532 - 532
  • [3] A COMPARISON OF SOLID PHANTOMS WITH WATER FOR THE DOSIMETRY OF I-125 BRACHYTHERAPY SOURCES
    MEIGOONI, AS
    MELI, JA
    NATH, R
    [J]. MEDICAL PHYSICS, 1987, 14 (03) : 476 - 476
  • [4] Radial dose functions for 103Pd, 125I, 169Yb and 192Ir brachytherapy sources:: an EGS4 Monte Carlo study
    Mainegra, E
    Capote, R
    López, E
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (03): : 703 - 717
  • [5] Dose rate constants for 125I, 103Pd, 192Ir and 169Yb brachytherapy sources:: an EGS4 Monte Carlo study
    Mainegra, E
    Capote, R
    Lopez, E
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (06): : 1557 - 1566
  • [6] Ocular brachytherapy dosimetry for 103Pd and 125I in the presence of gold nanoparticles: a Monte Carlo study
    Asadi, Somayeh
    Vaez-zadeh, Mehdi
    Vahidian, Mohammad
    Marghchouei, Mahdieh
    Masoudi, S. Farhad
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (03): : 90 - 99
  • [7] Ocular Brachytherapy Dosimetry for 103Pd and 125I in The Presence of Gold Nanoparticles: Monte Carlo Study
    Asadi, S.
    Vaez-zadeh, M.
    Vahidian, M.
    Marghchouei, M.
    Masoudi, S. Farhad
    [J]. RADIOTHERAPY AND ONCOLOGY, 2016, 118 : S105 - S105
  • [8] Application of Hybrid Source in Slotted Plaque Brachytherapy for the Treatment of Uveal Melanoma and Dosimetric Comparison of Plans with I-125 and Pd-103 Sources
    Awan, S.
    Meigooni, A.
    Howard, S.
    Tejwani, A.
    Badkul, R.
    Saleh, H.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E579 - E579
  • [9] Quantitative comparison between initial DNA-damage RBE of I-125 and Pd-103 brachytherapy sources through a hybrid simulation approach
    Hamid Reza Baghani
    Shiva Nasrollahi
    [J]. The European Physical Journal Plus, 136
  • [10] Quantitative comparison between initial DNA-damage RBE of I-125 and Pd-103 brachytherapy sources through a hybrid simulation approach
    Baghani, Hamid Reza
    Nasrollahi, Shiva
    [J]. EUROPEAN PHYSICAL JOURNAL PLUS, 2021, 136 (02):