Quiescent-Inflow Single-Shot Magnetic Resonance Angiography Using a Highly Undersampled Radial k-Space Trajectory

被引:20
|
作者
Edelman, R. R. [1 ,2 ]
Giri, S. [3 ]
Dunkle, E. [1 ]
Galizia, M. [2 ]
Amin, P. [2 ]
Koktzoglou, I. [1 ,4 ]
机构
[1] NorthShore Univ HealthSyst, Evanston, IL 60201 USA
[2] Northwestern Univ, Feinberg Sch Med, Chicago, IL 60611 USA
[3] Siemens Healthcare, Chicago, IL USA
[4] Univ Chicago, Pritzker Sch Med, Chicago, IL 60637 USA
基金
美国国家卫生研究院;
关键词
nonenhanced MR angiography; radial k-space trajectory; highly accelerated MRI; quiescent-inflow single-shot; peripheral arterial disease; MR-ANGIOGRAPHY; PHASED-ARRAY; CONTRAST; DISEASE; FAT;
D O I
10.1002/mrm.24596
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeWe hypothesized that high undersampling factors could be used in conjunction with radial quiescent-inflow single-shot magnetic resonance angiography (MRA) to accelerate the data acquisition and enable multislice acquisitions. MethodsSeven subjects were imaged on a 1.5 T MRI system. For multislice quiescent-inflow single-shot MRA, the venous saturation radiofrequency pulse, in-plane saturation radiofrequency pulse, and quiescent interval were applied only once before the first slice. ResultsThe mean (standard deviation) measurements for the intra-arterial signal-to-noise ratio were as follows: Cartesian 1 slice29.3 (5.5); radial 1 slice, 92 views22.3 (3.6); radial 1 slice, 46 views18.5 (2.0); radial 2 slices, 46 views18.3 (3.2); and radial 3 slices, 32 views21.7 (3.9), normalized for pixel size to 15.8. Horizontal striping was present with multislice radial quiescent-inflow single-shot MRA (especially with the three-slice acquisition) due to variable T-1 relaxation between the concurrently acquired slices, but the image quality remained diagnostic. Vascular pathology in patients with peripheral arterial disease was well shown by all techniques. ConclusionVery high undersampling factors in excess of 18 have been demonstrated for nonenhanced MRA using a radial quiescent-inflow single-shot technique, enabling the acquisition of two to three slices per cardiac cycle. Scan time for a complete peripheral MRA could be shortened to 2 min or less. Magn Reson Med 70:1662-1668, 2013. (c) 2012 Wiley Periodicals, Inc.
引用
收藏
页码:1662 / 1668
页数:7
相关论文
共 50 条
  • [41] Real-time reduction of motion artefacts using k-space weighting in magnetic resonance imaging
    Sinkus, R
    Bornert, P
    [J]. PHILIPS JOURNAL OF RESEARCH, 1998, 51 (02) : 339 - 352
  • [42] Centrally fat-saturated three-dimensional magnetic resonance angiography of the abdomen using selective central fat-saturation of K-space
    Amano, Y
    Takahama, K
    Matsuda, T
    Amano, M
    Kumazaki, T
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2003, 18 (05) : 567 - 574
  • [43] Magnetic resonance phase contrast velocity mapping for flow quantification in irregular heart rhythms using radial k-space ultrashort echo time imaging
    Hell, Michaela M.
    Francis, Jane M.
    d'Arcy, Joanna
    Robson, Matthew D.
    Neubauer, Stefan
    Achenbach, Stephan
    Myerson, Saul G.
    [J]. INTERNATIONAL JOURNAL OF CARDIOLOGY, 2020, 317 : 211 - 215
  • [44] MAGNETIC RESONANCE IMAGE RECONSTRUCTION FROM PARTIAL K-SPACE USING ANALOGICAL BASIS DECONSTRUCTION THEORY
    Liu, Fengqin
    Miao, Peng
    [J]. ADVANCES IN BIOMEDICAL PHOTONICS AND IMAGING, 2008, : 344 - +
  • [45] Single-shot vector magnetic measurements using synchronous coherent population trapping resonance in a feedback compensation system
    Pati, Gour
    Tripathi, Renu
    Pulido, Mauricio
    [J]. JOURNAL OF APPLIED PHYSICS, 2023, 134 (18)
  • [46] Parallel acquisition of q-space using second order magnetic fields for single-shot diffusion measurements
    Kittler, W. C.
    Galvosas, P.
    Hunter, M. W.
    [J]. JOURNAL OF MAGNETIC RESONANCE, 2014, 244 : 46 - 52
  • [47] Contrast-free magnetic resonance imaging protocol for TAVI guidance: 3D, whole heart' and quiescent-interval single-shot angiography in comparison with contrast-enhanced CT
    Plappert, David
    Pamminger, Mathias
    Klug, Gert
    Kranewitter, Christof
    Reindl, Martin
    Reinstadler, Sebastian J.
    Henninger, Benjamin
    Tiller, Christina
    Holzknecht, Magdalena
    Kremser, Christian
    Jaschke, Werner
    Metzler, Bernhard
    Mayr, Agnes
    [J]. WIENER KLINISCHE WOCHENSCHRIFT, 2019, 131 : 342 - 342
  • [48] Correcting magnetic resonance k-space data for in-plane motion using an optical position tracking system
    Marxen, M.
    Marmurek, J.
    Baker, N.
    Graham, S. J.
    [J]. MEDICAL PHYSICS, 2009, 36 (12) : 5580 - 5585
  • [49] Model-based myocardial T1 mapping with sparsity constraints using single-shot inversion-recovery radial FLASH cardiovascular magnetic resonance
    Xiaoqing Wang
    Florian Kohler
    Christina Unterberg-Buchwald
    Joachim Lotz
    Jens Frahm
    Martin Uecker
    [J]. Journal of Cardiovascular Magnetic Resonance, 21
  • [50] Model-based myocardial T1 mapping with sparsity constraints using single-shot inversion-recovery radial FLASH cardiovascular magnetic resonance
    Wang, Xiaoqing
    Kohler, Florian
    Unterberg-Buchwald, Christina
    Lotz, Joachim
    Frahm, Jens
    Uecker, Martin
    [J]. JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2019, 21 (01)