Developing a MLC modifier program to improve fiducial detection for MV/kV imaging during hypofractionated prostate volumetric modulated arc therapy

被引:12
|
作者
Happersett, Laura [1 ]
Wang, Ping [1 ]
Zhang, Pengpeng [1 ]
Mechalakos, James [1 ]
Li, Guang [1 ]
Eley, Eleanor [1 ]
Zelefsky, Michael [1 ]
Mageras, Gig [1 ]
Damato, Antonio L. [1 ]
Hunt, Margie [1 ]
机构
[1] Mem Sloan Kettering Canc Ctr, 1275 York Ave, New York, NY 10021 USA
来源
关键词
fiducial tracking; image-guided radiotherapy; motion management; MV; kV imaging; prostate;
D O I
10.1002/acm2.12614
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To develop an Eclipse plug-in (MLC_MODIFIER) that automatically modifies control points to expose fiducials obscured by MLC during VMAT, thereby facilitating tracking using periodic MV/kV imaging. Method Three-dimensional fiducial tracking was performed during VMAT by pairing short-arc (3 degrees) MV digital tomosynthesis (DTS) images to triggered kV images. To evaluate MLC_MODIFIER efficacy, two cohorts of patients were considered. For first 12 patients, plans were manually edited to expose one fiducial marker. Next for 15 patients, plans were modified using MLC_MODIFIER script. MLC_MODIFIER evaluated MLC apertures at appropriate angles for marker visibility. Angles subtended by control points were compressed and low-dose "imaging" control points were inserted and exposed one marker with 1 cm margin. Patient's images were retrospectively reviewed to determine rate of MV registration failures. Failure categories were poor DTS image quality, MLC blockage of fiducials, or unknown reasons. Dosimetric differences in rectum, bladder, and urethra D1 cc, PTV maximum dose, and PTV dose homogeneity (PTV HI) were evaluated. Statistical significance was evaluated using Fisher's exact and Student's t test. Result Overall MV registration failures, failures due to poor image quality, MLC blockage, and unknown reasons were 33% versus 8.9% (P < 0.0001), 8% versus 6.4% (P < 0.05), 13.6% versus 0.1% (P < 0.0001), and 7.6% versus 2.4% (P < 0.0001) for manually edited and MLC_MODIFIER plans, respectively. PTV maximum and HI increased on average from unmodified plans by 2.1% and 0.3% (P < 0.004) and 22.0% and 3.3% (P < 0.004) for manually edited and MLC_MODIFIED plans, respectively. Changes in bladder, rectum, and urethra D1CC were similar for each method and less than 0.7%. Conclusion Increasing fiducial visibility via an automated process comprised of angular compression of control points and insertion of additional "imaging" control points is feasible. Degradation of plan quality is minimal. Fiducial detection and registration success rates are significantly improved compared to manually edited apertures.
引用
收藏
页码:120 / 124
页数:5
相关论文
共 50 条
  • [21] Optimizing Fiducial Visibility On Periodically Acquired Megavoltage and Kilovoltage Image Pairs During Volumetric Modulated Arc Therapy
    Zhang, P.
    Wang, P.
    Laura, H.
    Ravindranath, B.
    Mageras, G.
    Hunt, M.
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3195 - 3195
  • [22] Comparison of Moderate Hypofractionated Volumetric-Modulated Arc Therapy Plans With and Without Flattening Filter for Localized Prostate Cancer
    Takakusagi, Yosuke
    Usui, Keisuke
    Mizoguchi, Nobutaka
    Nagatsuka, Jun
    Hikage, Takeshi
    Kodama, Yasuhiko
    Ezura, Takatomo
    Kusunoki, Terufumi
    Oizumi, Yukio
    [J]. CUREUS JOURNAL OF MEDICAL SCIENCE, 2021, 13 (09)
  • [23] Ultra hypofractionated extended nodal irradiation using volumetric modulated arc therapy for oligorecurrent pelvic nodal prostate cancer
    Slevin, Finbar
    Thompson, Christopher M.
    Speight, Richard
    Murray, Louise J.
    Lilley, John
    Henry, Ann M.
    [J]. MEDICAL DOSIMETRY, 2021, 46 (04) : 411 - 418
  • [24] Role of fiducial markers in the assessment of prostate bed motion in post-prostatectomy patients treated with volumetric modulated arc therapy
    Vlachaki, Maria T.
    Ahmad, Salahuddin
    Kennedy, Erika
    Aref, Amr M.
    Chuba, Paul J.
    [J]. JOURNAL OF RADIOTHERAPY IN PRACTICE, 2020, 19 (03) : 299 - 304
  • [25] A Novel Approach to Tracking Intrafraction Prostate Motion during Volumetric Modulated Arc Therapy with Tomosynthesis
    Zhou, J.
    Turian, J.
    Lee, E.
    Templeton, A.
    Liao, Y.
    Chu, J.
    [J]. MEDICAL PHYSICS, 2010, 37 (06)
  • [26] Dynamic Collimator Angle Adjustments During Volumetric Modulated Arc Therapy to Account for Prostate Rotations
    de Boer, Johan
    Wolf, Anne Lisa
    Szeto, Yenny Z.
    van Herk, Marcel
    Sonke, Jan-Jakob
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 91 (05): : 1009 - 1016
  • [27] Determination of Action Thresholds for Electromagnetic Tracking System-guided Hypofractionated Prostate Radiotherapy using Volumetric Modulated Arc Therapy
    Zhang, P.
    Mah, D.
    Happersett, L.
    Cox, B.
    Hunt, M.
    Mageras, G.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (03): : S675 - S676
  • [28] Determination of action thresholds for electromagnetic tracking system-guided hypofractionated prostate radiotherapy using volumetric modulated arc therapy
    Zhang, Pengpeng
    Mah, Dennis
    Happersett, Laura
    Cox, Brett
    Hunt, Margie
    Mageras, Gig
    [J]. MEDICAL PHYSICS, 2011, 38 (07) : 4001 - 4008
  • [29] Tracking of organ motion during arc therapy with MV treatment beam imaging and as-needed kV imaging
    Liu, W.
    Mao, W.
    Wiersma, R.
    Luxton, G.
    Riaz, N.
    Schlaefer, A.
    Xing, L.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (01): : S609 - S609
  • [30] Automatic Prostate Tracking and Motion Assessment in Volumetric Modulated Arc Therapy With an Electronic Portal Imaging Device
    Diego Azcona, Juan
    Li, Ruijiang
    Mok, Edward
    Hancock, Steven
    Xing, Lei
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 86 (04): : 762 - 768