Intracochlear Inflammatory Response to Cochlear Implant Electrodes in Humans

被引:2
|
作者
Seyyedi, Mohammad [1 ,2 ]
Nadol, Joseph B., Jr. [1 ,2 ]
机构
[1] Massachusetts Eye & Ear Infirm, Dept Otolaryngol, Boston, MA 02114 USA
[2] Harvard Univ, Sch Med, Dept Otol & Laryngol, Boston, MA 02115 USA
关键词
Cochlear implant; Endocochlear; Foreign body; Giant cell; Labyrinthitis; Tissue response; FOREIGN-BODY REACTION; HUMAN TEMPORAL BONES; CHRONIC ELECTRICAL-STIMULATION; AUDITORY-NERVE; RESIDUAL HEARING; TISSUE FORMATION; HISTOPATHOLOGY; BIOMATERIALS; COMPLICATIONS; GRANULOMA;
D O I
暂无
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Hypothesis: This study evaluates the types and degrees of tissue response adjacent to the electrode of multichannel cochlear implants. Background: Cochlear implant electrodes have been classified as biocompatible prostheses. Nevertheless, in some reports, electrode extrusion, chronic inflammation, and even soft failure of the implant system have been attributed to a tissue response to the electrode. Methods: All celloidin-embedded temporal bones with multichannel cochlear implants from the temporal bone collection of the Massachusetts Eye and Ear Infirmary were included in the study. A total of 28 temporal bones from 21 subjects were identified and processed for histology. The severity of cellular response including eosinophil and lymphocytic infiltration, giant cell reaction, new bone formation, and fibrosis were scored on a scale from 0 to 3 at three 1-mm segments along the electrode: first 1 mm at the cochleostomy, last 1 mm from the tip of the electrode, and midway between these proximal and distal segments. The values were compared using the Wilcoxon test. Results: A granulomatous reaction to the electrode was observed in 27 (96.4%) temporal bones. Eosinophil infiltration was observed in 7 (25%) temporal bones, suggesting an allergic reaction. The Inflammatory response to the electrode was significantly greater at the basal turn of cochlea close to the cochleostomy (p < 0.05) than distal to it. Conclusion: An inflammatory response is common after cochlear implantation, and it is more robust at the cochleostomy than distal to it, suggesting the role of trauma of insertion as a contributing factor.
引用
收藏
页码:1545 / 1551
页数:7
相关论文
共 50 条
  • [41] Automatic Localization of Cochlear Implant Electrodes in CT
    Zhao, Yiyuan
    Dawant, Benoit M.
    Labadie, Robert F.
    Noble, Jack H.
    MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION - MICCAI 2014, PT I, 2014, 8673 : 331 - +
  • [42] Potential insertion complications with cochlear implant electrodes
    Ishiyama, Akira
    Risi, Frank
    Boyd, Paul
    COCHLEAR IMPLANTS INTERNATIONAL, 2020, 21 (04) : 206 - 219
  • [43] Measurements of generated potentials by the electrodes of a cochlear implant
    Pesatori, A.
    Norgia, M.
    Sibella, F.
    Burdo, S.
    Tognola, G.
    Svelto, C.
    MEMEA 2007: SECOND - IEEE INTERNATIONAL WORKSHOP ON MEDICAL MEASUREMENT AND APPLICATIONS, 2007, : 10 - +
  • [44] Cochlear implant electrodes as electrochemical sensors in vivo
    Weltin, A.
    Kieninger, J.
    Urban, G. A.
    Arndt, S.
    Rosskothen-Kuhl, N.
    2021 10TH INTERNATIONAL IEEE/EMBS CONFERENCE ON NEURAL ENGINEERING (NER), 2021, : 617 - 620
  • [45] SURGICAL COMPLICATIONS WITH THE COCHLEAR MULTIPLE-CHANNEL INTRACOCHLEAR IMPLANT - EXPERIENCE AT HANNOVER AND MELBOURNE
    WEBB, RL
    LASZIG, R
    LEHNHARDT, E
    PYMAN, BC
    CLARK, GM
    FRANZ, BKHG
    ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1991, 100 (02): : 131 - 136
  • [46] Histologic evaluation of the tissue seal and biologic response around cochlear implant electrodes in the human
    Nadol, JB
    Eddington, DK
    OTOLOGY & NEUROTOLOGY, 2004, 25 (03) : 257 - 262
  • [47] Intracochlear New Fibro-Ossification and Neuronal Degeneration Following Cochlear Implant Electrode Translocation: Long-Term Histopathological Findings in Humans
    Knoll, Renata M.
    Trakimas, Danielle R.
    Wu, Matthew J.
    Lubner, Rory J.
    Nadol, Joseph B.
    Ishiyama, Akira
    Santos, Felipe
    Jung, David H.
    Remenschneider, Aaron K.
    Kozin, Elliott D.
    OTOLOGY & NEUROTOLOGY, 2022, 43 (02) : E153 - E164
  • [48] The Use of Micro-CT to Evaluate Cochlear Implant Electrode Position and Intracochlear Damage
    Fama, Anthony
    Carlson, Matthew
    Driscoll, Colin
    Lane, John
    LARYNGOSCOPE, 2010, 120 : S205 - S205
  • [49] Intracochlear Recording of Electrocochleography During and After Cochlear Implant Insertion Dependent on the Location in the Cochlea
    Haumann, Sabine
    Timm, Max E.
    Buechner, Andreas
    Lenarz, Thomas
    Salcher, Rolf B.
    TRENDS IN HEARING, 2024, 28
  • [50] Cochlear implant material effects on inflammatory cell function and foreign body response
    Jensen, Megan J.
    Claussen, Alexander D.
    Higgins, Timon
    Vielman-Quevedo, Rene
    Mostaert, Brian
    Xu, Linjing
    Kirk, Jonathon
    Hansen, Marlan R.
    HEARING RESEARCH, 2022, 426