SURGICAL COMPLICATIONS WITH THE COCHLEAR MULTIPLE-CHANNEL INTRACOCHLEAR IMPLANT - EXPERIENCE AT HANNOVER AND MELBOURNE

被引:120
|
作者
WEBB, RL
LASZIG, R
LEHNHARDT, E
PYMAN, BC
CLARK, GM
FRANZ, BKHG
机构
[1] UNIV MELBOURNE,ROYAL VICTORIAN EYE & EAR HOSP,DEPT OTOLARYNGOL,MELBOURNE,VIC 3002,AUSTRALIA
[2] MED ACAD HANNOVER,DEPT OTOLARYNGOL,HANNOVER,GERMANY
来源
关键词
COCHLEAR IMPLANT; COMPLICATIONS; DEAFNESS; SURGERY;
D O I
10.1177/000348949110000208
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
The surgical complications for the first 153 multiple-channel cochlear implant operations carried out at the Medizinische Hochschule in Hannover and the first 100 operations at the University of Melbourne Clinic, The Royal Victorian Eye and Ear Hospital, are presented. In the Hannover experience the major complications were wound breakdown, wound infection, electrode tie erosion through the external auditory canal, electrode slippage, a persistent increase in tinnitus, and facial nerve stimulation. The incidence of wound breakdown requiring removal of the package was 0.6% in Hannover and 1.0% in Melbourne. The complications for the operation at both clinics were at acceptable levels. It was considered that wound breakdown requiring implant removal could be kept to a minimum by making a generous incision and suturing the flap without tension.
引用
收藏
页码:131 / 136
页数:6
相关论文
共 46 条
  • [1] SURGERY FOR AN IMPROVED MULTIPLE-CHANNEL COCHLEAR IMPLANT
    CLARK, GM
    PYMAN, BC
    BAILEY, QE
    WEBB, RL
    SHEPHERD, RK
    [J]. ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1984, 93 (03): : 204 - 207
  • [2] AN ACOUSTIC MODEL OF A MULTIPLE-CHANNEL COCHLEAR IMPLANT
    BLAMEY, PJ
    DOWELL, RC
    TONG, YC
    CLARK, GM
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1984, 76 (01): : 97 - 103
  • [3] PRELIMINARY-REPORT ON A MULTIPLE-CHANNEL COCHLEAR IMPLANT OPERATION
    TONG, YC
    BLACK, RC
    CLARK, GM
    FORSTER, IC
    MILLAR, JB
    OLOUGHLIN, BJ
    PATRICK, JF
    [J]. JOURNAL OF LARYNGOLOGY AND OTOLOGY, 1979, 93 (07): : 679 - 695
  • [4] PSYCHOPHYSICAL STUDIES FOR 2 MULTIPLE-CHANNEL COCHLEAR IMPLANT PATIENTS
    TONG, YC
    CLARK, GM
    BLAMEY, PJ
    BUSBY, PA
    DOWELL, RC
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1982, 71 (01): : 153 - 160
  • [5] A MULTIPLE-CHANNEL COCHLEAR IMPLANT - AN EVALUATION USING NONSENSE SYLLABLES
    CLARK, GM
    TONG, YC
    DOWELL, RC
    MARTIN, LF
    SELIGMAN, PM
    BUSBY, PA
    PATRICK, JF
    [J]. ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1981, 90 (03): : 227 - 230
  • [6] A MULTIPLE-CHANNEL COCHLEAR IMPLANT - EVALUATION USING SPEECH TRACKING
    MARTIN, LFA
    TONG, YC
    CLARK, GM
    [J]. ARCHIVES OF OTOLARYNGOLOGY-HEAD & NECK SURGERY, 1981, 107 (03) : 157 - 159
  • [7] A MULTIPLE-CHANNEL COCHLEAR IMPLANT - A SUMMARY OF RESULTS FOR 2 PATIENTS
    CLARK, GM
    TONG, YC
    [J]. ARCHIVES OF OTOLARYNGOLOGY-HEAD & NECK SURGERY, 1982, 108 (04) : 214 - 217
  • [8] MULTIPLE-CHANNEL STIMULATION OF THE COCHLEAR NUCLEUS
    EVANS, DE
    NIPARKO, JK
    MILLER, JM
    JYUNG, RW
    ANDERSON, DJ
    [J]. OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 1989, 101 (06) : 651 - 657
  • [9] A 12-CONSONANT CONFUSION STUDY ON A MULTIPLE-CHANNEL COCHLEAR IMPLANT PATIENT
    DOWELL, RC
    MARTIN, LFA
    TONG, YC
    CLARK, GM
    SELIGMAN, PM
    PATRICK, JF
    [J]. JOURNAL OF SPEECH AND HEARING RESEARCH, 1982, 25 (04): : 509 - 516
  • [10] A MULTIPLE-CHANNEL COCHLEAR IMPLANT AND WEARABLE SPEECH-PROCESSOR - AN AUDIOLOGICAL EVALUATION
    TONG, YC
    CLARK, GM
    DOWELL, RC
    MARTIN, LFA
    SELIGMAN, PM
    PATRICK, JF
    [J]. ACTA OTO-LARYNGOLOGICA, 1981, 92 (3-4) : 193 - 198