Development of a golden beam data set for the commissioning of a proton double-scattering system in a pencil-beam dose calculation algorithm

被引:3
|
作者
Slopsema, R. L. [1 ]
Lin, L. [2 ]
Flampouri, S. [1 ]
Yeung, D. [1 ]
Li, Z. [1 ]
McDonough, J. E. [2 ]
Palta, J. [3 ]
机构
[1] Univ Florida, Proton Therapy Inst, Jacksonville, FL 32205 USA
[2] Univ Penn, Dept Radiat Oncol, Philadelphia, PA 19104 USA
[3] Virginia Commonwealth Univ, VCU Massey Canc Ctr, Richmond, VA 23298 USA
关键词
proton therapy; commissioning; passive scattering; golden beam data; ACCELERATORS; SERIES;
D O I
10.1118/1.4893281
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this investigation is to determine if a single set of beam data, described by a minimal set of equations and fitting variables, can be used to commission different installations of a proton double-scattering system in a commercial pencil-beam dose calculation algorithm. Methods: The beam model parameters required to commission the pencil-beam dose calculation algorithm (virtual and effective SAD, effective source size, and pristine-peak energy spread) are determined for a commercial double-scattering system. These parameters are measured in a first room and parameterized as function of proton energy and nozzle settings by fitting four analytical equations to the measured data. The combination of these equations and fitting values constitutes the golden beam data (GBD). To determine the variation in dose delivery between installations, the same dosimetric properties are measured in two additional rooms at the same facility, as well as in a single room at another facility. The difference between the room-specific measurements and the GBD is evaluated against tolerances that guarantee the 3D dose distribution in each of the rooms matches the GBD-based dose distribution within clinically reasonable limits. The pencil-beam treatment-planning algorithm is commissioned with the GBD. The three-dimensional dose distribution in water is evaluated in the four treatment rooms and compared to the treatment-planning calculated dose distribution. Results: The virtual and effective SAD measurements fall between 226 and 257 cm. The effective source size varies between 2.4 and 6.2 cm for the large-field options, and 1.0 and 2.0 cm for the small-field options. The pristine-peak energy spread decreases from 1.05% at the lowest range to 0.6% at the highest. The virtual SAD as well as the effective source size can be accurately described by a linear relationship as function of the inverse of the residual energy. An additional linear correction term as function of RM-step thickness is required for accurate parameterization of the effective SAD. The GBD energy spread is given by a linear function of the exponential of the beam energy. Except for a few outliers, the measured parameters match the GBD within the specified tolerances in all of the four rooms investigated. For a SOBP field with a range of 15 g/cm(2) and an air gap of 25 cm, the maximum difference in the 80%-20% lateral penumbra between the GBD-commissioned treatment-planning system and measurements in any of the four rooms is 0.5 mm. Conclusions: The beam model parameters of the double-scattering system can be parameterized with a limited set of equations and parameters. This GBD closely matches the measured dosimetric properties in four different rooms. (C) 2014 American Association of Physicists in Medicine.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Golden beam data for proton pencil-beam scanning
    Clasie, Benjamin
    Depauw, Nicolas
    Fransen, Maurice
    Goma, Carles
    Panahandeh, Hamid Reza
    Seco, Joao
    Flanz, Jacob B.
    Kooy, Hanne M.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (05): : 1147 - 1158
  • [2] Implementation of apertures in a proton pencil-beam dose algorithm
    Depauw, N.
    Kooy, H. M.
    Daartz, J.
    Bussiere, M.
    Batin, E.
    Madden, T.
    Williams, M.
    Schueman, J.
    Clasie, B. M.
    [J]. BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2022, 8 (02)
  • [3] Out-Of-Beam Dose for a Compact Double-Scattering Proton Beam Therapy System
    Islam, M.
    Ahmad, S.
    Jin, H.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3493 - 3493
  • [4] Incorporating partial shining effects in proton pencil-beam dose calculation
    Li, Yupeng
    Zhang, Xiaodong
    Lii, MingFwu
    Sahoo, Narayan
    Zhu, Ron X.
    Gillin, Michael
    Mohan, Radhe
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (03): : 605 - 616
  • [5] Extended collimator model for pencil-beam dose calculation in proton radiotherapy
    Kanematsu, Nobuyuki
    Akagi, Takashi
    Takatani, Yasuyuki
    Yonai, Shunsuke
    Sakamoto, Hidenobu
    Yamashita, Haruo
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (19): : 4807 - 4817
  • [6] ACCELERATION OF A PENCIL-BEAM DOSE CALCULATION ALGORITHM WITH GRAPHICS PROCESSING UNITS
    Arhjoul, L.
    Hissoiny, S.
    Ozell, B.
    Despres, P.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S70 - S70
  • [7] A Golden Beam Data Commissioning Framework of Monte Carlo Dose Calculation Algorithms of Two Pencil Beam Scanning Proton Therapy Treatment Planning Systems
    Chang, C.
    Harms, J.
    Zhang, R.
    Zhou, J.
    Lin, Y.
    Slopsema, R.
    Dhabaan, A.
    Liu, T.
    McDonald, M.
    Langen, K.
    Lin, L.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E167 - E167
  • [8] Commissioning of the world's first compact pencil-beam scanning proton therapy system
    Pidikiti, Rajesh
    Patel, Bijal C.
    Maynard, Matthew R.
    Dugas, Joseph P.
    Syh, Joseph
    Sahoo, Narayan
    Wu, Hsinshun Terry
    Rosen, Lane R.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (01): : 94 - 105
  • [9] Development of a Proton Beam Double-Scattering Apparatus for Radiobiology Research
    Egan, A.
    McDonough, J.
    Kiker, R.
    [J]. MEDICAL PHYSICS, 2010, 37 (06)
  • [10] PENCIL-BEAM REDEFINITION ALGORITHM FOR ELECTRON DOSE DISTRIBUTIONS
    SHIU, AS
    HOGSTROM, KR
    [J]. MEDICAL PHYSICS, 1991, 18 (01) : 7 - 18