Radiochromic film dosimetry of a high dose rate beta source for intravascular brachytherapy

被引:27
|
作者
Duggan, DM [1 ]
Coffey, CW
Lobdell, JL
Schell, MC
机构
[1] Vanderbilt Univ, Med Ctr, Ctr Radiat Oncol, Nashville, TN 37232 USA
[2] Novoste Corp, Norcross, GA 30093 USA
[3] Univ Rochester, Med Ctr, Dept Radiat Oncol, Rochester, NY 14642 USA
关键词
intravascular brachytherapy; Sr-90; radiochromic film; brachytherapy dosimetry; quality assurance;
D O I
10.1118/1.598814
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Good clinical physics practice requires that dose rates of brachytherapy sources be checked by the institution using them, as recommended by American Association of Physicists in Medicine Task Group 56 and The American College of Radiology. For intravascular brachytherapy with catheter-based systems, AAPM Task Group 60 recommends that the dose rate be measured at a reference point located at a radial distance of 2 mm from the center of the catheter axis. AAPM Task Group 60 also recommends that the dose rate along the catheter axis at a radial distance of 2 mm should be uniform to within +/- 10% in the center two-thirds of the treated length, and the relative dose rate in the plane perpendicular to the catheter axis through the center of the source should be measured at distances from 0.5 mm to R-90 (the distance from a point source within which 90% of the energy is deposited) at intervals of 0.5 mm. Radiochromic him dosimetry has been used to measure the dose distribution in a plane parallel to and at a radial distance of 2 mm from the axis of a novel, catheter-based, beta source for intravascular brachytherapy. The dose rate was averaged along a line parallel to the catheter axis at a radial distance of 2 mm, in the centered 24.5 mm of the treated length. This average dose rate agreed with the dose rate measured with a well ionization chamber by the replacement method using source trains calibrated with an extrapolation chamber at the National institute of Standards and Technology. Ail of the dose rates in the centered 24.5 mm of a line parallel to the axis at a distance of 2 mm were within +/- 10% of the average. (C) 1999 American Association of Physicists in Medicine. [S0094-2405(99)01711-3].
引用
收藏
页码:2461 / 2464
页数:4
相关论文
共 50 条
  • [1] In vivo dosimetry of high dose rate brachytherapy using an improved radiochromic film
    Parsai, E
    Feldmeier, J
    Ye, S
    [J]. MEDICAL PHYSICS, 2002, 29 (06) : 1206 - 1206
  • [2] Radiochromic film dosimetry for a new electronic brachytherapy source
    Chiu-Tsao, S
    Rusch, T
    Axelrod, S
    Tsao, H
    Harrison, L
    [J]. MEDICAL PHYSICS, 2004, 31 (06) : 1913 - 1913
  • [3] Radiochromic film dosimetry of rectal inhomogeneity and applicator attenuation in high dose rate brachytherapy of uterine cervix
    Uniyal, Satish C.
    Naithani, Umesh C.
    Sharma, Sunil D.
    Srivastava, Anoop K.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2012, 13 (01): : 66 - 75
  • [4] High dose-rate brachytherapy source position quality assurance using radiochromic film
    Evans, M. D. C.
    Devic, S.
    Podgorsak, E. B.
    [J]. MEDICAL DOSIMETRY, 2007, 32 (01) : 13 - 15
  • [5] Application of a color scanner for 60Co high dose rate brachytherapy dosimetry with EBT radiochromic film
    Ghorbani, Mahdi
    Toossi, Mohammad Taghi Bahreyni
    Mowlavi, Ali Asghar
    Roodi, Shahram Bayani
    Meigooni, Ali Soleimani
    [J]. RADIOLOGY AND ONCOLOGY, 2012, 46 (04) : 363 - 369
  • [6] RADIOCHROMIC FILM DOSIMETRY OF A NEW HIGH DOSE RATE (HDR) APPLICATOR
    Hariri, S.
    Aghamiri, M. R.
    Najarian, S.
    Jaberi, R.
    Dehghan, H. R.
    Azmi, M.
    Hajizadeh, S.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2009, 91 : S45 - S45
  • [7] Application of Gafchromic® film in the dosimetry of an intravascular brachytherapy source
    Song, Haijun
    Roa, D. Eduardo
    Yue, Ning
    d'Errico, Francesco
    Chen, Zhe
    Nath, Ravinder
    [J]. MEDICAL PHYSICS, 2006, 33 (07) : 2519 - 2524
  • [8] A comparison of polyacrylamide gels and radiochromic film for source measurements in intravascular brachytherapy
    Amin, MN
    Horsfield, MA
    Bonnett, DE
    Dunn, MJ
    Poulton, M
    Harding, PF
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2003, 76 (911): : 824 - 831
  • [9] Radiochromic film measurement of anisotropy function for high-dose-rate Ir-192 brachytherapy source
    Sharma, SD
    Bianchi, C
    Conte, L
    Novario, R
    Bhatt, BC
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (17): : 4065 - 4072
  • [10] The use of improved radiochromic film for in vivo quality assurance of high dose rate brachytherapy
    Pai, S
    Reinstein, LE
    Gluckman, G
    Xu, ZG
    Weiss, T
    [J]. MEDICAL PHYSICS, 1998, 25 (07) : 1217 - 1221