Monte Carlo simulation of the effect of magnetic fields on brachytherapy dose distributions in lung tissue material

被引:3
|
作者
Moreno-Barbosa, Fernando [1 ]
de Celis-Alonso, Benito [1 ]
Moreno-Barbosa, Eduardo [1 ]
Hernandez-Lopez, Javier Miguel [1 ]
Geoghegan, Theodore [2 ]
Ramos-Mendez, Jose [3 ]
机构
[1] Benemerita Univ Autonoma Puebla, Fac Math & Phys Sci, Mexico City, DF, Mexico
[2] Univ Iowa Hosp & Clin, Dept Radiat Oncol, Iowa City, IA 52242 USA
[3] Univ Calif San Francisco, Dept Radiat Oncol, San Francisco, CA USA
来源
PLOS ONE | 2020年 / 15卷 / 10期
关键词
MV RADIOTHERAPY ACCELERATOR; MRI SCANNER; ELECTRON; DOSIMETRY; ALGORITHM; TOMOGRAPHY; DEPOSITION; GEANT4; TOPAS;
D O I
10.1371/journal.pone.0238704
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
The aim of this work was to use TOPAS Monte Carlo simulations to model the effect of magnetic fields on dose distributions in brachytherapy lung treatments, under ideal and clinical conditions. Idealistic studies were modeled consisting of either a monoenergetic electron source of 432 keV, or a polyenergetic electron source using the spectrum of secondary electrons produced by(192)Ir gamma-ray irradiation. The electron source was positioned in the center of a homogeneous, lung tissue phantom (rho = 0.26 g/cm(3)). Conversely, the clinical study was simulated using the VariSource VS2000 Ir-192 source in a patient with a lung tumor. Three contoured volumes were considered: the tumor, the planning tumor volume (PTV), and the lung. In all studies, dose distributions were calculated in the presence or absence of a constant magnetic field of 3T. Also, TG-43 parameters were calculated for the VariSource and compared with published data from EGS-brachy (EGSnrc) and PENELOPE. The magnetic field affected the dose distributions in the idealistic studies. For the monoenergetic and poly-energetic studies, the radial distance of the 10% iso-dose line was reduced in the presence of the magnetic field by 64.9% and 24.6%, respectively. For the clinical study, the magnetic field caused differences of 10% on average in the patient dose distributions. Nevertheless, differences in dose-volume histograms were below 2%. Finally, for TG-43 parameters, the dose-rate constant from TOPAS differed by 0.09% +/- 0.33% and 0.18% +/- 0.33% with respect to EGS-brachy and PENELOPE, respectively. The geometry and anisotropy functions differed within 1.2% +/- 1.1%, and within 0.0% +/- 0.3%, respectively. The Lorentz forces inside a 3T magnetic resonance machine during(192)Ir brachytherapy treatment of the lung are not large enough to affect the tumor dose distributions significantly, as expected. Nevertheless, large local differences were found in the lung tissue. Applications of this effect are therefore limited by the fact that meaningful differences appeared only in regions containing air, which is not abundant inside the human.
引用
收藏
页数:14
相关论文
共 50 条
  • [41] THE EFFECT OF RIB AND LUNG INHOMOGENEITY ON THE DOSE DELIVERY TO LUNG AS A RESULT OF IR-192 HDR BREAST BRACHYTHERAPY: A MONTE CARLO STUDY
    Yazdi, H. Salehi
    Jaberi, R.
    Shabani, H.
    Vaezzadeh, S. A.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S404 - S404
  • [42] A Monte Carlo investigation of lung brachytherapy treatment planning
    Sutherland, J. G. H.
    Furutani, K. M.
    Thomson, R. M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (14): : 4763 - 4780
  • [43] A Monte Carlo study on tissue dose enhancement in brachytherapy: a comparison between gadolinium and gold nanoparticles
    Mohammad Taghi Bahreyni Toossi
    Mahdi Ghorbani
    Mohammad Mehrpouyan
    Fateme Akbari
    Leila Sobhkhiz Sabet
    Ali Soleimani Meigooni
    Australasian Physical & Engineering Sciences in Medicine, 2012, 35 : 177 - 185
  • [44] Sensitivity of low energy brachytherapy Monte Carlo dose calculations to uncertainties in human tissue composition
    Landry, Guillaume
    Reniers, Brigitte
    Murrer, Lars
    Lutgens, Ludy
    Bloemen-Van Gurp, Esther
    Pignol, Jean-Philippe
    Keller, Brian
    Beaulieu, Luc
    Verhaegen, Frank
    MEDICAL PHYSICS, 2010, 37 (10) : 5188 - 5198
  • [45] A Monte Carlo study on tissue dose enhancement in brachytherapy: a comparison between gadolinium and gold nanoparticles
    Toossi, Mohammad Taghi Bahreyni
    Ghorbani, Mahdi
    Mehrpouyan, Mohammad
    Akbari, Fateme
    Sabet, Leila Sobhkhiz
    Meigooni, Ali Soleimani
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2012, 35 (02) : 177 - 185
  • [46] Monte Carlo simulation of radiated seed source brachytherapy
    School of the Nuclear Science and Technology, Nanhua University, Hengyang 421001, China
    不详
    Qinghua Daxue Xuebao, 2007, SUPPL. 1 (924-927):
  • [47] Patient-Specific Monte Carlo Simulation of Lung Brachytherapy: Metallic Artifact Reduction and Organ-Constrained Tissue Assignment
    Sutherland, J. G. H.
    Miksys, N.
    Furutani, K. M.
    Thomson, R. M.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [48] Uncertainties of Monte Carlo simulation for brachytherapy source dosimetry
    Rachabatthula, V
    Meigooni, A
    MEDICAL PHYSICS, 2004, 31 (06) : 1732 - 1732
  • [49] Monte Carlo simulation of multi-dimensional dose distribution for high dose-rate brachytherapy of cancer
    Lee, H
    Hsi, W
    Chu, J
    Saxena, A
    MEDICAL PHYSICS, 2002, 29 (06) : 1231 - 1231
  • [50] Dose Calculation in Gynecological Brachytherapy using Monte Carlo simulation for intracavitary treatment of Cervical Cancer
    Garcia Contreras, O. J.
    Cristancho, F.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, 2015, VOLS 1 AND 2, 2015, 51 : 452 - 455