Electron beam therapy at extended SSDs: an analysis of output correction factors for a Mitsubishi linear accelerator
被引:3
|
作者:
Rajasekar, D
论文数: 0引用数: 0
h-index: 0
机构:
Sanjay Gandhi Postgrad Inst Med Sci, Dept Radiotherapy, Lucknow, Uttar Pradesh, IndiaSanjay Gandhi Postgrad Inst Med Sci, Dept Radiotherapy, Lucknow, Uttar Pradesh, India
Rajasekar, D
[1
]
Datta, NR
论文数: 0引用数: 0
h-index: 0
机构:
Sanjay Gandhi Postgrad Inst Med Sci, Dept Radiotherapy, Lucknow, Uttar Pradesh, IndiaSanjay Gandhi Postgrad Inst Med Sci, Dept Radiotherapy, Lucknow, Uttar Pradesh, India
Datta, NR
[1
]
Das, KJM
论文数: 0引用数: 0
h-index: 0
机构:
Sanjay Gandhi Postgrad Inst Med Sci, Dept Radiotherapy, Lucknow, Uttar Pradesh, IndiaSanjay Gandhi Postgrad Inst Med Sci, Dept Radiotherapy, Lucknow, Uttar Pradesh, India
Das, KJM
[1
]
Ayyagari, S
论文数: 0引用数: 0
h-index: 0
机构:
Sanjay Gandhi Postgrad Inst Med Sci, Dept Radiotherapy, Lucknow, Uttar Pradesh, IndiaSanjay Gandhi Postgrad Inst Med Sci, Dept Radiotherapy, Lucknow, Uttar Pradesh, India
Ayyagari, S
[1
]
机构:
[1] Sanjay Gandhi Postgrad Inst Med Sci, Dept Radiotherapy, Lucknow, Uttar Pradesh, India
来源:
PHYSICS IN MEDICINE AND BIOLOGY
|
2002年
/
47卷
/
18期
关键词:
D O I:
10.1088/0031-9155/47/18/303
中图分类号:
R318 [生物医学工程];
学科分类号:
0831 ;
摘要:
The effects of extended source-to-surface distance (SSD) on the electron beam dose profiles were evaluated for various electron beam energies-6, 9, 12, 15 and 20 MeV-and the accuracy of various output correction methods was analysed on a Mitsubishi linear accelerator using a radiation field analyser (RFA). The dose fall-off region of the central axis depth-dose curves was nearly independent for SSDs up to 120 cm where as in the build-up region, a marginal reduction of surface dose was observed, particularly for lower energies and for smaller field sizes. Effective SSDs and virtual source distances were evaluated for field sizes ranging from 5 x 5 to 15 x 15 cm(2) for various energies. Curve fitting was done with the measured outputs with various equations and coefficients were evaluated. The accuracy of the derived output correction factors by effective SSD, virtual source distance and curve-fit methods was assessed by evaluating correlation coefficients between the calculated and the measured values. The correlation coefficient was best with the linear-quadratic equation followed by the effective SSD method and the virtual source method. The output correction based on the linear-quadratic equation showed the best estimate of electron beam output at extended SSDs with an accuracy well within +/-1%. The rapid reduction of dose due to the applicator-scattered component at d(max) point with an extended SSD was significant for the 5 x 5 cm(2) applicator and lower energies.
机构:
Qingdao West Coast New Area Peoples Hosp, Qingdao 266400, Shandong, Peoples R ChinaQingdao West Coast New Area Peoples Hosp, Qingdao 266400, Shandong, Peoples R China
Wang, Tao
Wang, Zhen
论文数: 0引用数: 0
h-index: 0
机构:
Qingdao West Coast New Area Peoples Hosp, Qingdao 266400, Shandong, Peoples R ChinaQingdao West Coast New Area Peoples Hosp, Qingdao 266400, Shandong, Peoples R China
Wang, Zhen
Xu, Lei
论文数: 0引用数: 0
h-index: 0
机构:
Cent Hosp Shengli Oil Field, Med Imaging Dept, Dongying 257035, Shandong, Peoples R ChinaQingdao West Coast New Area Peoples Hosp, Qingdao 266400, Shandong, Peoples R China