Effects of bifurcation-specific and conventional stents on coronary bifurcation flow. An experimental and numerical study

被引:12
|
作者
Garcia Garcia, J. [1 ]
Garcia Carrascal, P. [2 ]
Castro Ruiz, F. [2 ]
Manuel Martin, F. [1 ]
Fernandez, J. A. [3 ]
机构
[1] Univ Politecn Madrid, Escuela Tecn Super Ingenieros Ind, Dept Ingn Energet, C Jose Gutierrez Abascal 2, E-28006 Madrid, Spain
[2] Univ Valladolid, Escuela Ingn Ind, Dept Ingn Energet & Fluidomecan, Paseo Cauce 59, E-47011 Valladolid, Spain
[3] Hosp Univ Puerta Hierro, Majadahonda, Spain
关键词
Computational fluid dynamics; Coronary bifurcation; Stenting techniques; Image-based reconstruction; Particle image velocimetry; WALL SHEAR-STRESS; MICRO-COMPUTED TOMOGRAPHY; IN-VITRO; FLUID-DYNAMICS; HEMODYNAMICS; IMPLANTATION; ANGIOPLASTY; APPOSITION;
D O I
10.1016/j.jbiomech.2017.01.043
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Our paper builds on existing research into conventional bare metal stents in order to assess new devices specifically designed for coronary bifurcation angioplasty. The first aim is to validate the numerical model against data from in vitro experiments on stented coronary phantoms. A surface mesh was built in accordance with micro-computed tomography images obtained from coronary stents implanted in silicone models and used for numerical analysis. Computational simulations for steady and unsteady cases generally agreed with their experimental counterparts. A second objective is to compare the hemodynamic performance of one of these new devices (Stentys) to that of conventional devices and stenting techniques in a simplified coronary bifurcation model. Four different coronary bifurcation stenting techniques were analyzed. We have focused on factors contributing to restenosis, such as wall shear stress (WSS), oscillatory shear index (OSI), pressure loss, and local normalized helicity (LNH). It was found that bifurcation-specific stents implanted in the side branch led to increased malapposition. This effect has proved to be more important than stent specific design characteristics such as strut size (different for conventional and Stentys stent). This conclusion is confirmed by means of drop in pressure and mechanical energy loss rate calculation; for the latter, the increase ranged from 9% to 17%, depending on the stenting technique, when dedicated stents were implanted in the side branch. The behavior patterns presented in this study should be double-checked against those obtained in more realistic geometries. (C) 2017 Elsevier Ltd. All rights reserved.
引用
收藏
页码:64 / 72
页数:9
相关论文
共 50 条
  • [41] Experimental study of the flow distribution uniformity in flow distributors having novel flow channel bifurcation structures
    Liu, Hong
    Li, Peiwen
    Van Lew, Jon
    Juarez-Robles, Daniel
    EXPERIMENTAL THERMAL AND FLUID SCIENCE, 2012, 37 : 142 - 153
  • [42] NUMERICAL SIMULATION OF TRANSIENT BLOOD FLOW THROUGH THE LEFT CORONARY ARTERY WITH VARYING DEGREES OF BIFURCATION ANGLES
    Zhang, Bo
    Jin, Yilun
    Wang, Xiaoran
    Zeng, Taisheng
    Wang, Liansheng
    JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY, 2017, 17 (01)
  • [43] Axisymmetric shape of gas bubbles in a uniform flow: Numerical study of bifurcation by analytic continuation
    Er-Riani, Mustapha
    Filali, Youness
    El Jarroudi, Mustapha
    APPLIED MATHEMATICS AND COMPUTATION, 2016, 274 : 306 - 317
  • [44] Numerical Study on Characteristics of Pulsitile Flow by Location of Stenosis in Blood Vessel with the Second Bifurcation
    Lee, In-sub
    Ryou, Hong-sun
    TRANSACTIONS OF THE KOREAN SOCIETY OF MECHANICAL ENGINEERS B, 2005, 29 (01) : 63 - 70
  • [45] PULSATILE NON-NEWTONIAN BLOOD-FLOW IN 3-DIMENSIONAL CAROTID BIFURCATION MODELS - A NUMERICAL STUDY OF FLOW PHENOMENA UNDER DIFFERENT BIFURCATION ANGLES
    PERKTOLD, K
    PETER, RO
    RESCH, M
    LANGS, G
    JOURNAL OF BIOMEDICAL ENGINEERING, 1991, 13 (06): : 507 - 515
  • [46] A numerical study of magnetic resonance images of pulsatile flow in a two dimensional carotid bifurcation - A numerical study of MR images
    Jou, LD
    Saloner, D
    MEDICAL ENGINEERING & PHYSICS, 1998, 20 (09) : 643 - 652
  • [47] Experimental and numerical study of the reflection of an oblique shock wave in a stationary hypersonic flow.
    Chpoun, A
    Passerel, D
    Lengrand, JC
    Li, H
    BenDor, G
    RECHERCHE AEROSPATIALE, 1996, (02): : 95 - 105
  • [48] A numerical and experimental analysis of the flow field in a two-dimensional model of the human carotid artery bifurcation
    Rindt, C.C.M.
    v.d. Vosse, F.N.
    v. Steenhoven, A.A.
    Janssen, J.D.
    1600, (20):
  • [49] Numerical model and experimental investigation of blood flow through a bifurcation: Interaction between an artery and a small prosthesis
    Department of Biomedical Engineering, Faculty of Engineering, Tel-Aviv University, Tel-Aviv, 69978, Israel
    不详
    ASAIO J., 4 (326-333):
  • [50] Numerical model and experimental investigation of blood flow through a bifurcation - Interaction between an artery and a small prosthesis
    Paz, D
    Einav, S
    Raderer, F
    Trubel, W
    ASAIO JOURNAL, 1997, 43 (04) : 326 - 333