Realization of the spectral responsivity scale based on an absolute cryogenic radiometer

被引:7
|
作者
Hong, Kee-Suk [1 ]
Park, Seongchong [1 ]
Kim, Seung-Kwan [1 ]
Park, Seung-Nam [1 ]
Lee, Dong-Hoon [1 ]
机构
[1] Korea Res Inst Stand & Sci, Ctr Photometry & Radiometry, Daejeon 34186, South Korea
关键词
Spectral responsivity; Cryogenic radiometer; Trap detector; Measurement uncertainty; TRAP DETECTORS; NM;
D O I
10.3938/jkps.67.2045
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
We present the concept of and the procedure for realizing the primary scale of the spectral responsivity based on an absolute cryogenic radiometer at the Korea Research Institute of Standards and Science (KRISS). A detailed error and uncertainty analysis is performed by calibrating the standard detectors in the wavelength range from 300 nm to 1000 nm. The relative expanded uncertainty of the spectral responsivity calibration for a Si trap detector is 0.06% at 632.8 nm and 0.4% for other wavelengths (k = 2). The calibration results are validated by comparing with the results from a theoretical model of quantum efficiency for a transmission-type trap detector.
引用
收藏
页码:2045 / 2058
页数:14
相关论文
共 50 条
  • [31] Development of a detector-based absolute spectral power responsivity scale in the spectral range of 300–1600 nm
    Abdallah M. Karmalawi
    Alaaeldin A. Abdelmageed
    Journal of Materials Science: Materials in Electronics, 2021, 32 : 5215 - 5221
  • [32] PROCEDURES FOR DETERMINING SPECTRAL RESPONSIVITY OF AN INFRARED RADIOMETER
    SCHNEIDE.WE
    GARVEY, JA
    APPLIED OPTICS, 1968, 7 (06): : 1141 - &
  • [33] High accuracy measurement of the absolute spectral responsivity of Ge and InGaAs trap detectors by direct calibration against an electrically calibrated cryogenic radiometer in the near-infrared
    Lopez, M.
    Hofer, H.
    Kueck, S.
    METROLOGIA, 2006, 43 (06) : 508 - 514
  • [34] A cryogenic radiometer for absolute neutron rate measurement
    Chowdhuri, Z
    Hansen, GL
    Jane, V
    Keith, CD
    Lozowski, WM
    Snow, WM
    Dewey, MS
    Gilliam, DM
    Greene, GL
    Nico, JS
    Thompson, AK
    Wietfeldtd, FE
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2003, 74 (10): : 4280 - 4293
  • [35] The new ultraviolet spectral responsivity scale based on cryogenic radiometry at Synchrotron Ultraviolet Radiation Facility III
    Shaw, PS
    Larason, TC
    Gupta, R
    Brown, SW
    Vest, RE
    Lykke, KR
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2001, 72 (05): : 2242 - 2247
  • [36] Realization of a photometric scale based on cryogenic radiometry
    Clare, JF
    Nield, KM
    METROLOGIA, 1998, 35 (04) : 251 - 254
  • [37] Realization of a photometric scale based on cryogenic radiometry
    Clare, J.F.
    Nield, K.M.
    Metrologia, 35 (04): : 251 - 254
  • [38] Design and calibration of electrical substitution radiometer traced to absolute cryogenic radiometer
    Zhang, Lei
    Xu, Wen-Bin
    Wang, Feng
    OPTICAL ENGINEERING, 2020, 59 (01)
  • [39] Development of a detector-based absolute spectral power responsivity scale in the spectral range of 300-1600 nm
    Karmalawi, Abdallah M.
    Abdelmageed, Alaaeldin A.
    JOURNAL OF MATERIALS SCIENCE-MATERIALS IN ELECTRONICS, 2021, 32 (04) : 5215 - 5221
  • [40] Absolute cryogenic radiometer for high accuracy optical radiant power measurement in a wide spectral range
    Gan, Haiyong
    He, Yingwei
    Liu, Xiangliang
    Xu, Nan
    Wu, Houping
    Feng, Guojin
    Liu, Wende
    Lin, Yandong
    CHINESE OPTICS LETTERS, 2019, 17 (09)