Agreement and repeatability for standard automated perimetry and confocal scanning laser ophthalmoscopy in the Diagnostic Innovations in Glaucoma Study

被引:7
|
作者
Ng, Diana [1 ]
Zangwill, Linda M. [1 ]
Racette, Lyne [1 ]
Bowd, Christopher [1 ]
Pascual, John P. [1 ]
Bourne, Rupert R. A. [1 ]
Boden, Catherine [1 ]
Weinreb, Robert N. [1 ]
Sample, Pamela A. [1 ]
机构
[1] Univ Calif San Diego, Hamilton Glaucoma Ctr, Dept Ophthalmol, La Jolla, CA 92093 USA
关键词
D O I
10.1016/j.ajo.2006.04.027
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
center dot PURPOSE: To assess agreement between structural and functional testing in classifying eyes as normal or abnormal and their repeatability on two consecutive visits by means of standard automated perimetry (SAP) and confocal scanning laser ophthalmoscopy. center dot DESIGN: Analysis of selected data obtained from a prospective longitudinal observational cohort study, the Diagnostic Innovations in Glaucoma Study. center dot METHODS: One hundred fifty,one participants with a SAP and a Heidelberg Retina Tomograph (HRT) test within a three,month window at two visits within 15 months were included. Eyes were classified by SAP and HRT at each visit. Agreement and repeatability were assessed by kappa statistics. center dot RESULTS: At visit 1, 33 (22%) of 151 eyes had only SAP defects, 11 (7%) eyes had only HRT defects, and 35 (23%) eyes had both SAP and HRT defects. Seventy-two eyes (48%) were classified as normal by both tests. Similar results were obtained for visit 2. The agreement between SAP and HRT in classifying eyes was 70% (kappa = 0.393) at visit one and 68% (kappa = 0.363) at visit 2. Repeatability of classification by SAP alone, HRT alone, and both SAP and HRT between visits 1 and 2 was 84% (kappa = 0.693), 89% (kappa = 0.752), and 90% (kappa = 0.723), respectively. center dot CONCLUSIONS: Agreement between SAP and HRT was only fair at both visits. Repeatability across visits was substantial for SAP alone, HRT alone, and for the combination of SAP and HRT. These results suggest that detection of particular features of glaucomatous damage depends on the technique used. Structural and functional tests appear complementary, and both should be used for early detection of glaucoma.
引用
收藏
页码:381 / 386
页数:6
相关论文
共 50 条
  • [1] Mapping structural to functional damage in glaucoma with standard automated perimetry and confocal scanning laser ophthalmoscopy
    Anton, A
    Yamagishi, N
    Zangwill, L
    Sample, PA
    Weinreb, RN
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 1998, 125 (04) : 436 - 446
  • [2] Agreement between standard automated perimetry and Heidelberg retina tomograph for detecting glaucoma in the diagnostic innovations in glaucoma study (DIGS)
    Ng, D
    Zangwill, LM
    Pascual, JP
    Bourne, RR
    Bowd, C
    Boden, C
    Weinreb, RN
    Sample, PA
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46
  • [3] Relationships between standard automated perimetry, HRT confocal scanning laser ophthalmoscopy, and GDx VCC scanning laser polarimetry
    Reus, NJ
    Lemij, HG
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46 (11) : 4182 - 4188
  • [4] The effects of study design and spectrum bias on the evaluation of diagnostic accuracy of confocal scanning laser ophthalmoscopy in glaucoma
    Medeiros, Felipe A.
    Ng, Diana
    Zangwill, Linda M.
    Sample, Pamela A.
    Bowd, Christopher
    Weinreb, Robert N.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2007, 48 (01) : 214 - 222
  • [5] Scanning laser polarimetry and confocal scanning laser ophthalmoscopy: technical notes on their use in glaucoma
    Ferreri, Felicia
    Aragona, Pasquale
    Ferreri, Giuseppe
    GLAUCOMA: AN OPEN WINDOW TO NEURODEGENERATION AND NEUROPROTECTION, 2008, 173 : 125 - 138
  • [6] Combination of structural testing (GDx VCC scanning laser polarimetry, stratus optical coherence tomography [OCT], HRT II confocal scanning laser ophthalmoscopy) and functional testing (frequency doubling technology [FDT] perimetry) for detection of glaucoma in the diagnostic innovations in glaucoma study (DIGS)
    Shah, NN
    Medeiros, FA
    Sample, PA
    Bowd, C
    Hoffmann, EM
    Weinreb, RN
    Zangwill, LM
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46
  • [7] Diagnostic Ability and Repeatability of a New Supra-Threshold Glaucoma Screening Program in Standard Automated Perimetry
    Takahashi, Natsumi
    Hirasawa, Kazunori
    Hoshina, Miki
    Kasahara, Masayuki
    Matsumura, Kazuhiro
    Shoji, Nobuyuki
    TRANSLATIONAL VISION SCIENCE & TECHNOLOGY, 2017, 6 (03):
  • [8] Combined Evaluation of Frequency Doubling Technology Perimetry and Scanning Laser Ophthalmoscopy for Glaucoma Detection Using Automated Classification
    Horn, Folkert K.
    Laemmer, Robert
    Mardin, Christian Y.
    Juenemann, Anselm G.
    Michelson, Georg
    Lausen, Berthold
    Adler, Werner
    JOURNAL OF GLAUCOMA, 2012, 21 (01) : 27 - 34
  • [9] Correlation between confocal scanning laser ophthalmoscopy and scanning laser polarimetry in open angle glaucoma
    Sihota, R
    Gulati, V
    Saxena, R
    Agarwal, HC
    Sharma, AK
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2003, 13 (03) : 266 - 275
  • [10] Repeatability and correlation of a virtual reality perimeter with standard automated perimetry in glaucoma patients
    Chaudhry, Anisa Fatima
    Berneshawi, Andrew Raphael
    Liu, Jocelyn
    Shue, Ann
    Chang, Dolly
    Kim, Julia
    Chang, Robert T.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2022, 63 (07)