Evaluation of a novel personal nanoparticle sampler

被引:7
|
作者
Zhou, Yue [1 ]
Irshad, Hammad [1 ]
Tsai, Chuen-Jinn [2 ]
Hung, Shao-Ming [2 ]
Cheng, Yung-Sung [1 ]
机构
[1] Lovelace Resp Res Inst, Albuquerque, NM 87108 USA
[2] NCTU, Hsinchu, Taiwan
关键词
INHALABLE AEROSOL SAMPLERS; ULTRAFINE PARTICLES; DEPOSITION; EXPOSURE; PERFORMANCE; AIRWAYS; CALIBRATION; WORKPLACES; EFFICIENCY; CHILDREN;
D O I
10.1039/c3em00497j
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This work investigated the performance in terms of collection efficiency and aspiration efficiency of a personal sampler capable of collecting ultrafine particles (nanoparticles) in the occupational environment. This sampler consists of a cyclone for respirable particle classification, micro-orifice impactor stages with an acceleration nozzle to achieve nanoparticle classification and a backup filter to collect nanoparticles. Collection efficiencies of the cyclone and impactor stages were determined using monodisperse polystyrene latex and silver particles, respectively. Calibration of the cyclone and impactor stages showed 50% cut-off diameters of 3.95 mu m and 94.7 nm meeting the design requirements. Aspiration efficiencies of the sampler were tested in a wind tunnel with wind speeds of 0.5, 1.0, and 1.5 m s(-1). The test samplers were mounted on a full size mannequin with three orientations toward the wind direction (0 degrees, 90 degrees, and 180 degrees). Monodisperse oleic acid aerosols tagged with sodium fluorescein in the size range of 2 to 10 mu m were used in the test. For particles smaller than 2 mu m, the fluorescent polystyrene latex particles were generated by using nebulizers. For comparison of the aspiration efficiency, a NIOSH two-stage personal bioaerosol sampler was also tested. Results showed that the orientation-averaged aspiration efficiency for both samplers was close to the inhalable fraction curve. However, the direction of wind strongly affected the aspiration efficiency. The results also showed that the aspiration efficiency was not affected by the ratio of free-stream velocity to the velocity through the sampler orifice. Our evaluation showed that the current design of the personal sampler met the designed criteria for collecting nanoparticles <= 100 nm in occupational environments.
引用
收藏
页码:203 / 210
页数:8
相关论文
共 50 条
  • [21] Field Application of a Newly Developed Personal Nanoparticle Sampler to Selected Metalworking Operations
    Young, Li-Hao
    Lin, Yun-Hua
    Lin, Tzu-Hsien
    Tsai, Perng-Jy
    Wang, Ying-Fang
    Hung, Shao-Ming
    Tsai, Chuen-Jinn
    Chen, Chun-Wan
    AEROSOL AND AIR QUALITY RESEARCH, 2013, 13 (03) : 849 - 861
  • [22] Development and evaluation of a new personal sampler for culturable airborne microorganisms
    Agranovski, IE
    Agranovski, V
    Reponen, T
    Willeke, K
    Grinshpun, SA
    ATMOSPHERIC ENVIRONMENT, 2002, 36 (05) : 889 - 898
  • [23] LABORATORY EVALUATION OF THE CIP-10 PERSONAL DUST SAMPLER
    GERO, A
    TOMB, T
    AMERICAN INDUSTRIAL HYGIENE ASSOCIATION JOURNAL, 1988, 49 (06): : 286 - 292
  • [24] Development of PM0.1 Personal Sampler for Evaluation of Personal Exposure to Aerosol Nanoparticles
    Thongyen, Thunyapat
    Hata, Mitsuhiko
    Toriba, Akira
    Ikeda, Takuji
    Koyama, Hiromi
    Otani, Yoshio
    Furuuchi, Masami
    AEROSOL AND AIR QUALITY RESEARCH, 2015, 15 (01) : 180 - 187
  • [25] SAMPLER OF PERSONAL OBSERVATORIES
    SCHRICKER, D
    WORRICK, C
    CORTES, M
    SODER, J
    HERS, J
    PRIMI, D
    SKY AND TELESCOPE, 1978, 56 (04): : 295 - 298
  • [26] PERSONAL SAMPLER FOR NOX
    PALMES, ED
    TOMCZYK, C
    AMERICAN INDUSTRIAL HYGIENE ASSOCIATION JOURNAL, 1979, 40 (07): : 588 - 591
  • [27] Development and Performance Evaluation of Air Sampler with Inertial Filter for Nanoparticle Sampling
    Furuuchi, Masami
    Eryu, Kazunobu
    Nagura, Mizuki
    Hata, Mitsuhiko
    Kato, Takaharu
    Tajima, Naoko
    Sekiguchi, Kazuhiko
    Ehara, Kensei
    Seto, Takafumi
    Otani, Yoshio
    AEROSOL AND AIR QUALITY RESEARCH, 2010, 10 (02) : 185 - 192
  • [28] MEASUREMENT OF PERSONAL EXPOSURE TO NO2 IN SWEDEN - EVALUATION OF A PASSIVE SAMPLER
    BERGLUND, M
    VAHTER, M
    BYLIN, G
    JOURNAL OF EXPOSURE ANALYSIS AND ENVIRONMENTAL EPIDEMIOLOGY, 1992, 2 (03): : 295 - 307
  • [29] Development of a Sampler for Evaluation of Personal Exposure to PM0.1 particles
    Thongen, Thunyapat
    Kusunoki, Masaru
    Hata, Mitsuhiko
    Bao, Linfa
    Furuuchi, Masami
    Otani, Yoshio
    Ikeda, Takuji
    29TH SYMPOSIUM ON AEROSOL SCIENCE AND TECHNOLOGY, 2012, 2012, : 233 - 234
  • [30] PERSONAL AIR SAMPLER FOR PESTICIDES
    HILL, RH
    ARNOLD, JE
    ARCHIVES OF ENVIRONMENTAL CONTAMINATION AND TOXICOLOGY, 1979, 8 (05) : 621 - 628