Novel blood pressure and pulse pressure estimation based on pulse transit time and stroke volume approximation

被引:19
|
作者
Lee, Joonnyong [1 ]
Sohn, JangJay [1 ]
Park, Jonghyun [1 ]
Yang, SeungMan [1 ]
Lee, Saram [2 ]
Kim, Hee Chan [3 ,4 ]
机构
[1] Seoul Natl Univ, Grad Sch, Interdisciplinary Program Bioengn, Suite 321,Bldg 8,101 Daehak Ro, Seoul 03080, South Korea
[2] Seoul Natl Univ Hosp, Biomed Res Inst, Suite 1203-1,71 Daehak Ro, Seoul 03082, South Korea
[3] Seoul Natl Univ, Dept Biomed Engn, Coll Med, Suite 11315,101 Daehak Ro, Seoul 03080, South Korea
[4] Seoul Natl Univ, Inst Med & Biol Engn, Med Res Ctr, Suite 11315,101 Daehak Ro, Seoul 03080, South Korea
来源
基金
新加坡国家研究基金会;
关键词
Blood pressure; Pulse pressure; Stroke volume; Pre-ejection period; Pulse transit time; Ubiquitous healthcare; TOTAL ARTERIAL COMPLIANCE; CORONARY-HEART-DISEASE; CARDIOVASCULAR RISK; ARRIVAL-TIME; HYPERTENSION; BALLISTOCARDIOGRAM; VARIABILITY; RATIO; ABNORMALITIES; POPULATION;
D O I
10.1186/s12938-018-0510-8
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Non-invasive continuous blood pressure monitors are of great interest to the medical community due to their value in hypertension management. Recently, studies have shown the potential of pulse pressure as a therapeutic target for hypertension, but not enough attention has been given to non-invasive continuous monitoring of pulse pressure. Although accurate pulse pressure estimation can be of direct value to hypertension management and indirectly to the estimation of systolic blood pressure, as it is the sum of pulse pressure and diastolic blood pressure, only a few inadequate methods of pulse pressure estimation have been proposed. Methods: We present a novel, non-invasive blood pressure and pulse pressure estimation method based on pulse transit time and pre-ejection period. Pre-ejection period and pulse transit time were measured non-invasively using electrocardiogram, seismocardiogram, and photoplethysmogram measured from the torso. The proposed method used the 2-element Windkessel model to model pulse pressure with the ratio of stroke volume, approximated by pre-ejection period, and arterial compliance, estimated by pulse transit time. Diastolic blood pressure was estimated using pulse transit time, and systolic blood pressure was estimated as the sum of the two estimates. The estimation method was verified in 11 subjects in two separate conditions with induced cardiovascular response and the results were compared against a reference measurement and values obtained from a previously proposed method. Results: The proposed method yielded high agreement with the reference (pulse pressure correlation with reference R >= 0.927, diastolic blood pressure correlation with reference R >= 0.854, systolic blood pressure correlation with reference R >= 0.914) and high estimation accuracy in pulse pressure (mean root-mean-squared error <= 3.46 mmHg) and blood pressure (mean root-mean-squared error <= 6.31 mmHg for diastolic blood pressure and <= 8.41 mmHg for systolic blood pressure) over a wide range of hemodynamic changes. Conclusion: The proposed pulse pressure estimation method provides accurate estimates in situations with and without significant changes in stroke volume. The proposed method improves upon the currently available systolic blood pressure estimation methods by providing accurate pulse pressure estimates.
引用
收藏
页数:20
相关论文
共 50 条
  • [21] BLOOD VOLUME AND PULSE PRESSURE
    GOLDMAN, H
    NEW ENGLAND JOURNAL OF MEDICINE, 1967, 277 (12): : 661 - &
  • [22] Estimation of systolic and diastolic pressure using the pulse transit time
    Ye, Soo-Young
    Kim, Gi-Ryon
    Jung, Dong-Keun
    Baik, Seong-Wan
    Jeon, Gye-Rok
    World Academy of Science, Engineering and Technology, 2010, 43 : 726 - 731
  • [23] Noninvasive Cuffless Blood Pressure Estimation Using Pulse Transit Time and Impedance Plethysmography
    Toan Huu Huynh
    Jafari, Roozbeh
    Chung, Wan-Young
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2019, 66 (04) : 967 - 976
  • [24] Estimation of pulse transit time using two diametric blood pressure waveform measurements
    Hahn, Jin-Oh
    Reisner, Andrew T.
    Asada, H. Harry
    MEDICAL ENGINEERING & PHYSICS, 2010, 32 (07) : 753 - 759
  • [25] Optimizing Non-Invasive Blood Pressure Estimation Using Pulse Transit Time
    Jeong, In Cheol
    Finkelstein, Joseph
    MEDINFO 2013: PROCEEDINGS OF THE 14TH WORLD CONGRESS ON MEDICAL AND HEALTH INFORMATICS, PTS 1 AND 2, 2013, 192 : 1198 - 1198
  • [26] Cuffless Blood Pressure Estimation Using Pulse Transit Time and Photoplethysmogram Intensity Ratio
    Gholamhosseini, Hamid
    Baig, Mirza
    Rastegar, Solmaz
    Linden, Maria
    PROCEEDINGS OF THE 15TH INTERNATIONAL CONFERENCE ON WEARABLE MICRO AND NANO TECHNOLOGIES FOR PERSONALIZED HEALTH (PHEALTH 2018), 2018, 249 : 77 - 83
  • [27] Pulse Transit Time Based Continuous Cuffless Blood Pressure Estimation: A New Extension and A Comprehensive Evaluation
    Xiaorong Ding
    Bryan P. Yan
    Yuan-Ting Zhang
    Jing Liu
    Ni Zhao
    Hon Ki Tsang
    Scientific Reports, 7
  • [28] Pulse Transit Time Based Continuous Cuffless Blood Pressure Estimation: A New Extension and A Comprehensive Evaluation
    Ding, Xiaorong
    Yan, Bryan P.
    Zhang, Yuan-Ting
    Liu, Jing
    Zhao, Ni
    Tsang, Hon Ki
    SCIENTIFIC REPORTS, 2017, 7
  • [29] Pulse Transit Time-based Blood Pressure Estimation Using Hilbert-Huang Transform
    Zhang, Qiao
    Shi, Yang
    Teng, Daniel
    Dinh, Anh
    Ko, Seok-Bum
    Chen, Li
    Basran, Jenny
    Dal Bello-Haas, Vanina
    Choi, Younhee
    2009 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-20, 2009, : 1785 - +
  • [30] Continuous estimation of stroke volume with mean pulse pressure.
    Rhee, KH