Feasibility of spinal stereotactic body radiotherapy in Elekta Unity® MR-Linac

被引:0
|
作者
Han, Eun Young [1 ]
Aima, Manik [1 ]
Hughes, Neil [2 ]
Briere, Tina M. [1 ]
Yeboa, Debra N. [2 ]
Castillo, Pam [2 ]
Wang, Jihong [1 ]
Yang, Jinzhong [1 ]
Vedam, Sastry [1 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Unit 94,1515 Holcomb Blvd, Houston, TX 77030 USA
[2] Univ Texas MD Anderson Canc Ctr, Dept Radiat Oncol, Houston, TX 77030 USA
来源
JOURNAL OF RADIOSURGERY AND SBRT | 2020年 / 7卷 / 02期
关键词
Spine SBRT; Unity MR-Linac; TrueBeam; MAGNETIC-FIELD; MONTE-CARLO; REFERENCE DOSIMETRY; ACCELERATOR; OPTIMIZATION; METASTASES; SCANNER; IMPACT;
D O I
暂无
中图分类号
R61 [外科手术学];
学科分类号
摘要
The Elekta Unity MR-Linac (MRL) is expected to benefit spine stereotactic body radiotherapy (SBRT) due to the improved soft tissue contrast available with onboard MR imaging. However, the irradiation geometry and beam configuration of the MRL deviates from the conventional linear accelerator (Linac). The purpose of the study was to investigate the feasibility of spine SBRT on the MRL. Treatment plans were generated for lumbar and thoracic spines. Target and spinal cord doses were measured with two cylindrical ion chambers inserted into an anthropomorphic spine phantom. Our study indicated that the Monaco treatment planning system (TPS) could generate clinical treatment plans for the MRL that were of comparable quality to the RayStation TPS with a conventional Linac. For both Linacs the planned dose within the gross tumor volume agreed with measurements within +/- 3%. For the spinal cord, while the measured doses from the TrueBeam were 1.8% higher for the lumbar spine plan and 6.9% higher for thoracic spine plan, the measured doses from MRL were 0.6% lower for the lumbar spine plan and 3.9% higher for the thoracic spine plan. In conclusion, the feasibility of spine SBRT in Elekta Unity MRL has been demonstrated, however, more effort is needed for such as optimizing the online plan adaptation method.
引用
收藏
页码:127 / 134
页数:8
相关论文
共 50 条
  • [11] Stereotactic body radiotherapy (SBRT) for localised prostate cancer on the MR-Linac
    Pathmanathan, A.
    Mitchell, A.
    Thomas, K.
    Henderson, D.
    Nill, S.
    Oelfke, U.
    Huddart, R.
    Van As, N.
    Tree, A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S445 - S445
  • [12] Markerless liver stereotactic body radiotherapy on a 1.5 T MR-Linac
    Gani, C.
    Simon, B.
    Nachbar, M.
    Stolte, A.
    Boldt, J.
    Marks, C.
    Thorwarth, D.
    Zips, D.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S907 - S908
  • [13] Secondary MU Check Using RadCalc for Elekta Unity MR-Linac
    Yang, J.
    Vedam, S.
    Brink, C.
    Nielsen, M.
    Ji, M.
    Bist, L.
    Ohrt, J.
    Wang, J.
    Balter, P.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E267 - E267
  • [14] Feasibility of stereotactic body radiotherapy of (ultra)central lung tumors using an 1.5 T MR-linac
    Merckel, L.
    Hackett, S.
    van Lier, A.
    van den Dobbelsteen, M.
    Rasing, M.
    Snoeren, L.
    van Es, C.
    Fast, M.
    van Rossum, P.
    Verhoeff, J.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S963 - S964
  • [15] Use of an a-Si EPID for routine QC of the Elekta Unity MR-Linac
    Berresford, J.
    Agnew, J.
    Harriden, T.
    Budgell, G.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S1159 - S1160
  • [16] Clinical Experience of Patient-Specific QA for Online Adaptive Radiotherapy Using Elekta Unity MR-Linac
    Yang, J.
    Vedam, S.
    Wang, J.
    Fuller, C.
    Choi, S.
    Chung, C.
    McAleer, M.
    Lee, B.
    Hughes, N.
    Ungchusri, G.
    Gillin, M.
    Martel, M.
    Balter, P.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E204 - E204
  • [17] Initial experience of rectum SIB radiotherapy on the Unity MR-Linac
    Stankiewicz, H.
    Chick, J.
    Ingle, M.
    Mitchell, A.
    Barnes, H.
    Herbert, T.
    Nill, S.
    Oelfke, U.
    Ng-Cheng-Hin, B.
    Bhide, S.
    Dunlop, A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2023, 182 : S1129 - S1130
  • [18] Elekta Unity MR-Linac Daily Automated End-To-End Quality Control
    Malkov, V.
    Winter, J.
    Letourneau, D.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E274 - E275
  • [19] Commissioning of a 1.5T Elekta Unity MR-linac: A single institution experience
    Snyder, Jeffrey E.
    St-Aubin, Joel
    Yaddanapudi, Sridhar
    Boczkowski, Amanda
    Dunkerley, David A. P.
    Graves, Stephen A.
    Hyer, Daniel E.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (07): : 160 - 172
  • [20] Output factors measurement with a new plastic scintillator detector in an Elekta Unity MR-Linac
    Ferrer, C.
    Huertas, C.
    Feijoo, M.
    Garcia, D.
    Saez, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2022, 170 : S1362 - S1362