When local impedance meets contact force: preliminary experience from the CHARISMA registry

被引:9
|
作者
Solimene, Francesco [1 ]
De Sanctis, Valerio [2 ]
Maggio, Ruggero [3 ]
Malacrida, Maurizio [4 ]
Segreti, Luca [5 ]
Anselmino, Matteo [6 ]
Schillaci, Vincenzo [1 ]
Mantica, Massimo [2 ]
Scaglione, Marco [7 ]
Dello Russo, Antonio [8 ]
Cauti, Filippo Maria [9 ]
Zingarini, Gianluca [10 ]
Pandozi, Claudio [11 ]
Cavaiani, Marco [4 ]
Ferraro, Anna [3 ]
Maglia, Giampiero [12 ]
Stabile, Giuseppe [1 ,13 ]
机构
[1] Clin Montevergine, Avellino, Mercogliano, Italy
[2] Ist Clin St Ambrogio, Dept Cardiac Electrophysiol & Pacing, Via Luigi Giuseppe Faravelli 16, I-20149 Milan, Italy
[3] Infermi Hosp, Rivoli, Italy
[4] Boston Sci, Milan, Italy
[5] Azienda Osped Univ Pisana, New Santa Chiara Hosp, Cardiac Thorac Vasc Dept, Div Cardiol 2, Pisa, Italy
[6] Univ Turin, Citta Salute & Sci Torino Hosp, Dept Med Sci, Div Cardiol, Turin, Italy
[7] Cardinal Massaia Hosp, Asti, Italy
[8] Marche Polytech Univ, Cardiol & Arrhythmol Clin, Ancona, Italy
[9] Osped San Giovanni Calibita, Fatebefratelli, Arrhythmol Unit, Rome, Italy
[10] Osped Santa Maria Misericordia, Perugia, Italy
[11] San Filippo Neri Hosp, Div Cardiol, Rome, Italy
[12] Azienda Osped Pugliese Ciaccio, Catanzaro, Italy
[13] Anthea Hosp, Bari, Italy
关键词
Atrial fibrillation; Catheter ablation; Local impedance; Contact force; Lesion formation; PVI; RADIOFREQUENCY CATHETER ABLATION; ATRIAL-FIBRILLATION;
D O I
10.1007/s10840-022-01163-7
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Purpose Highly localized impedance (LI) measurements during atrial fibrillation (AF) ablation have emerged as a viable real-time indicator of tissue characteristics and the consequent durability of the lesions created. We investigated the impact of catheter-tissue contact force (CF) on LI behavior during pulmonary vein isolation (PVI). Methods Forty-five consecutive patients of the CHARISMA registry undergoing de novo AF radiofrequency (RF) catheter ablation with a novel open-irrigated-tip catheter endowed with CF and LI measurement capabilities (Stablepoint (TM) catheter, Boston Scientific) were included. Results A total of 2895 point-by-point RF applications were analyzed (RF delivery time (DT) = 8.7 +/- 4s, CF = 13 +/-+/- 8 g, LI drop = 23 +/-+/- 7 Omega). All PVs were successfully isolated in an overall procedure time of 118 +/-+/- 34 min (fluoroscopy time = 13 +/-+/- 8 min). The magnitude of LI drop weakly correlated with CF (r = 0.13, 95% confidence interval (CI): 0.09 to 0.16, p < 0.0001), whereas both CF and LI drop inversely correlated with DT (r = -0.26, 95%CI: -0.29 to -0.22, p < 0.0001 for CF; r = -0.36, 95%CI: -0.39 to -0.33, p < 0.0001 for LI). For each 10 g of CF, LI drop markedly increased from 22.4 +/- 7 Omega to 24.0 +/- 8 Omega at 5 to 25 g CF intervals (5-14 g of CF vs 15-24 g of CF, p < 0.0001), whereas it showed smooth transition over 25 g (24.8 +/- 7 Omega at >= 25 g CF intervals, p = 0.0606 vs 15-24 g of CF). No major complications occurred during the procedures or within 30 days. Conclusions CF significantly affects LI drop and probable consequent lesion formation during RF PVI. The benefit of higher contact (> 25 g) between the catheter and the tissue appears to have less impact on LI drop.
引用
收藏
页码:749 / 758
页数:10
相关论文
共 50 条
  • [21] Local impedance measurements during contact force-guided cavotricuspid isthmus ablation for predicting an effective radiofrequency ablation
    Sasaki, Takehito
    Nakamura, Kohki
    Minami, Kentaro
    Take, Yutaka
    Nakatani, Yosuke
    Miki, Yuko
    Goto, Koji
    Kaseno, Kenichi
    Yamashita, Eiji
    Koyama, Keiko
    Naito, Shigeto
    JOURNAL OF ARRHYTHMIA, 2022, 38 (02) : 245 - 252
  • [22] When Do Citizens Respond Politically to the Local Economy? Evidence from Registry Data on Local Housing Markets
    Larsen, Martin Vinaes
    Hjorth, Frederik
    Dinesen, Peter Thisted
    Sonderskov, Kim Mannemar
    AMERICAN POLITICAL SCIENCE REVIEW, 2019, 113 (02) : 499 - 516
  • [23] Complementary use of contact force and local catheter impedance during RF ablation reduces ablation time in an in vivo swine model
    Garrott, K.
    Sugrue, A.
    Laughner, J.
    Bush, J.
    Gutbrod, S.
    Sulkin, M.
    Yasin, O.
    Meyers, J.
    Kapa, S.
    EUROPEAN HEART JOURNAL, 2019, 40 : 1805 - 1805
  • [24] Higher contact force, energy setting, and impedance rise during radiofrequency ablation predicts charring: New insights from contact force-guided in vivo ablation
    Makimoto, Hisaki
    Metzner, Andreas
    Tilz, Roland Richard
    Lin, Tina
    Heeger, Christian-H
    Rillig, Andreas
    Mathew, Shibu
    Lemes, Christine
    Wissner, Erik
    Kuck, Karl-Heinz
    Ouyang, Feifan
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2018, 29 (02) : 227 - 235
  • [25] Contact force and impedance decrease during ablation depends on catheter location and orientation: insights from pulmonary vein isolation using a contact force-sensing catheter
    Knecht, Sven
    Reichlin, Tobias
    Pavlovic, Nikola
    Schaer, Beat
    Osswald, Stefan
    Sticherling, Christian
    Kuehne, Michael
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2015, 43 (03) : 297 - 306
  • [26] Contact force and impedance decrease during ablation depends on catheter location and orientation: insights from pulmonary vein isolation using a contact force-sensing catheter
    Sven Knecht
    Tobias Reichlin
    Nikola Pavlovic
    Beat Schaer
    Stefan Osswald
    Christian Sticherling
    Michael Kühne
    Journal of Interventional Cardiac Electrophysiology, 2015, 43 : 297 - 306
  • [27] Performance of the ALSFRSr when administered by phone to patients and to caregivers: Experience from the VA ALS registry
    Kasarskis, EJ
    Allen, KD
    Lindquist, JH
    Coffman, CJ
    Oddone, EZ
    NEUROLOGY, 2006, 66 (05) : A246 - A246
  • [28] A novel local impedance algorithm to guide effective pulmonary vein isolation in AF patients: 1-year outcome from the CHARISMA Study
    Segreti, L.
    De Simone, A.
    Schillaci, V
    Zucchelli, G.
    Pandozi, C.
    Pecora, D.
    Pelargonio, G.
    Arestia, A.
    De Lucia, R.
    Shopova, G.
    Viani, S.
    Ferrari, C.
    Casati, F.
    Malacrida, M.
    Solimene, F.
    EUROPEAN HEART JOURNAL, 2020, 41 : 600 - 600
  • [29] Pulmonary vein isolation in atrial fibrillation patients guided by a novel local impedance algorithm: 1-year outcome from the CHARISMA study
    Solimene, Francesco
    Santoro, Mario Giannotti
    De Simone, Antonio
    Malacrida, Maurizio
    Stabile, Giuseppe
    Pandozi, Claudio
    Pelargonio, Gemma
    Cauti, Filippo Maria
    Scaglione, Marco
    Pecora, Domenico
    Bongiorni, Maria Grazia
    Arestia, Alberto
    Grimaldi, Gabriella
    Russo, Maurizio
    Narducci, Maria Lucia
    Segreti, Luca
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2021, 32 (06) : 1540 - 1548
  • [30] Initial impedance decrease as an indicator of good catheter contact: Insights from radiofrequency ablation with force sensing catheters
    Reichlin, Tobias
    Knecht, Sven
    Lane, Christopher
    Kuehne, Michael
    Nof, Eyal
    Chopra, Nagesh
    Tadros, Thomas M.
    Reddy, Vivek Y.
    Schaer, Beat
    John, Roy M.
    Osswald, Stefan
    Stevenson, William G.
    Sticherling, Christian
    Michaud, Gregory F.
    HEART RHYTHM, 2014, 11 (02) : 194 - 201