Development and Validation of High-Throughput Bioanalytical Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS) Method for the Quantification of Newly Synthesized Antitumor Carbonic Anhydrase Inhibitors in Human Plasma

被引:2
|
作者
Abdel-Megied, Ahmed M. [1 ,2 ]
Eldehna, Wagdy M. [3 ]
Abdelrahman, Mohamed A. [4 ]
Elbarbry, Fawzy A. [2 ]
机构
[1] Kafrelsheikh Univ, Pharmaceut Analyt Chem Dept, Fac Pharm, Kafrelsheikh City 33516, Egypt
[2] Pacific Univ Oregon, Sch Pharm, Hillsboro, OR 97123 USA
[3] Kafrelsheikh Univ, Fac Pharm, Dept Pharmaceut Chem, Kafrelsheikh City 33516, Egypt
[4] Egyptian Russian Univ, Fac Pharm, Dept Pharmaceut Chem, Cairo 11829, Egypt
来源
MOLECULES | 2020年 / 25卷 / 23期
关键词
LC-MS; MS; tandem mass; carbonic anhydrase inhibitors; antitumor; human plasma; bioanalytical validation;
D O I
10.3390/molecules25235753
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
In the present study, a sensitive and fully validated bioanalytical high-performance liquid chromatography-tandem mass spectrometry (LC-MS/MS) method has been developed for the quantitative determination of three newly synthesized carbonic anhydrases inhibitors (CAIs) with potential antitumor activity in human plasma. The analytes and the internal standard (IS) were extracted using 1.5 mL acetonitrile from only 450 mu L aliquots of human plasma to achieve the desired protein precipitation. Chromatographic separations were achieved on Phenomenex Kinetex(R) C-18 column (100 x 4.6 mm, 2.6 mu m) using a binary gradient elution mode with a run time of less than 6 min. The mobile phase consisted of solvent (A): 0.1% formic acid in 50% methanol and solvent B: 0.1% formic acid in acetonitrile (30:70, v/v), pumped at a flow rate of 0.8 mL/min. Detection was employed using triple quadrupole tandem mass spectrometer (API 3500) equipped with an electrospray ionization (ESI) source in the positive ion mode. Multiple reaction monitoring (MRM) mode was selected for quantitation through monitoring the precursor-to-parent ion transition at m/z 291.9 (R) 173.0, m/z 396.9 (R) 225.1, m/z 388.9 (R) 217.0, and m/z 146.9 (R) 91.0 for AW-9a, WES-1, WES-2, and Coumarin (IS), respectively. Linearity was computed using the weighted least-squares linear regression method (1/x(2)) over a concentration range of 1-1000, 2.5-800, and 5-500 ng/mL for AW-9a, WES-1, and WES-2; respectively. The bioanalytical LC-MS/MS method was fully validated as per U.S. Food and Drug Administration (FDA) guidelines with all respect to linearity, accuracy, precision, carry-over, selectivity, dilution integrity, and stability. The proposed LC-MS/MS method was applied successfully for the determination of all investigated drugs in spiked human plasma with no significant matrix effect, which is a crucial cornerstone in further therapeutic drug monitoring of newly developed therapeutic agents.
引用
收藏
页数:12
相关论文
共 50 条
  • [41] KINETICS OF PLASMA APOLIPOPROTEIN E ISOFORMS IN HUMANS BY LIQUID CHROMATOGRAPHY-TANDEM MASS SPECTROMETRY (LC-MS/MS)
    Croyal, M.
    Blanchard, V.
    Ramin-Mangata, S.
    Ouguerram, K.
    Lambert, G.
    Krempf, M.
    ATHEROSCLEROSIS, 2018, 275 : E29 - E29
  • [42] Liquid Chromatography Tandem Mass Spectrometry (LC-MS/MS) Method for the Determination of Vildagliptin in Rat Plasma
    Sakthimanigandan, K.
    Ganesh, M.
    Kanthikiran, V. S.
    Sivakumar, T.
    Jang, H. T.
    ACTA CHROMATOGRAPHICA, 2015, 27 (02) : 295 - 307
  • [43] Quantitation of Phenolic Benzotriazole Class Compounds in Plasma by Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS)
    Mutlu, Esra
    South, Natalie
    Pierfelice, Jessica
    Djonabaye, Alison
    Pauff, Mindy
    Burback, Brian
    Waidyanatha, Suramya
    ANALYTICAL LETTERS, 2022, 55 (13) : 2074 - 2088
  • [44] Sensitive and simple liquid chromatography-tandem mass spectrometric (LC-MS/MS) method for estimation of valsartan in human plasma
    Chandramohan, K.
    Rathnakar, P.
    Savale, S. S.
    Hemanthkumar, K.
    Sam, V. A.
    Sundar, G.
    Singh, T.
    Nagraj, S.
    Arvind, A. S.
    JOURNAL OF PHARMACY AND PHARMACOLOGY, 2007, 59 : A29 - A29
  • [45] Rapid and accurate method for quantifying busulfan in plasma samples by isocratic liquid chromatography-tandem mass spectrometry (LC-MS/MS)
    Villena-Ortiz, Yolanda
    Castellote-Belles, Laura
    Martinez-Sanchez, Luisa
    Benitez-Carabante, Maria I.
    Miarons, Marta
    Vima-Bofarull, Jaume
    Barquin-DelPino, Raquel
    Paciucci, Rosanna
    Rodriguez-Frias, Francisco
    Ferrer-Costa, Roser
    Casis, Ernesto
    Lopez-Hellin, Joan
    ADVANCES IN LABORATORY MEDICINE-AVANCES EN MEDICINA DE LABORATORIO, 2022, 3 (03): : 263 - 271
  • [46] Development of rapid and high-throughput LC-MS/MS method for quantification of olprinone in rabbit plasma
    Kertys, Martin
    Kosutova, Petra
    Mokra, Daniela
    Mokry, Juraj
    BIOMEDICAL CHROMATOGRAPHY, 2019, 33 (10)
  • [47] Bioanalytical method development, validation and quantification of flupirtine maleate in rat plasma by liquid chromatography-tandem mass spectrometry
    Kandasamy, Karthikeyan
    Gowdra, Vasantharaju Surenahalli
    Nammalvar, Hariprabhu
    Govindarajan, Arul Kumaran Kottur S.
    ARZNEIMITTELFORSCHUNG-DRUG RESEARCH, 2011, 61 (12): : 693 - 699
  • [48] A liquid chromatography-tandem mass spectrometry (LC-MS/MS)-based assay for simultaneous quantification of aldosterone, renin activity, and angiotensin II in human plasma
    Chen, Fangjun
    Cheng, Ziyun
    Peng, Yingfei
    Wang, Zhenxin
    Huang, Chao
    Liu, Dan
    Wang, Beili
    Pan, Baishen
    Guo, Wei
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2021, 1179
  • [49] Quantification of Ptaquiloside and Pterosin B in Soil and Groundwater Using Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS)
    Jensen, Pia H.
    Jacobsen, Ole S.
    Hansen, Hans Christian B.
    Juhler, Rene K.
    JOURNAL OF AGRICULTURAL AND FOOD CHEMISTRY, 2008, 56 (21) : 9848 - 9854
  • [50] Rapid and sensitive method for quantification of gestodene in human plasma as the oxime derivative by liquid chromatography-tandem mass spectrometry (LC-MS/MS) and its application to bioequivalence study
    Saxena, Ashish
    Gupta, Arun
    Kasibhatta, Ravisekhar
    Bob, Manoj
    Kumar, V. Praveen
    Purwar, Bipin
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2014, 945 : 240 - 246