Multi-slit prompt-gamma camera for locating of distal dose falloff in proton therapy

被引:2
|
作者
Park, Jong Hoon [1 ]
Kim, Sung Hun [1 ]
Ku, Youngmo [1 ]
Kim, Chan Hyeong [1 ]
Lee, Han Rim [2 ]
Jeong, Jong Hwi [3 ]
Lee, Se Byeong [3 ]
Shin, Dong Ho [4 ]
机构
[1] Hanyang Univ, Dept Nucl Engn, 222 Wangsimni Ro, Seoul 04763, South Korea
[2] Korea Atom Energy Res Inst, Neutron Utilizat Technol Div, Daejeon 34129, South Korea
[3] Natl Canc Ctr, Proton Therapy Ctr, Goyang Si 10408, Gyonggi Do, South Korea
[4] Elect & Telecommun Res Inst, IT Convergence Technol Res Lab, Daejeon 34129, South Korea
关键词
Proton therapy; Prompt gamma; Beam range verification; Distal dose falloff; BEAM RANGE VERIFICATION; COMPTON CAMERA; RADIATION HARDNESS; PENCIL BEAMS; SLIT CAMERA; TARGETS; SYSTEM;
D O I
10.1016/j.net.2019.03.008
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
In this research, a multi-slit prompt-gamma camera was developed to locate the distal dose falloff of the proton beam spots in spot scanning proton therapy. To see the performance of the developed camera, therapeutic proton beams were delivered to a solid plate phantom and then the prompt gammas from the phantom were measured using the camera. Our results show that the camera locates the 90% distal dose falloff (= d90%), within about 2-3 mm of error for the spots which are composed 3.8 x 10(8) protons or more. The measured location of d90% is not very sensitive to the irradiation depth of the proton beam (i.e., the depth of proton beam from the phantom surface toward which the camera is located). Considering the number of protons per spot for the most distal spots in typical treatment cases (i.e., 2 Gy dose divided in 2 fields), the camera can locate d90% only for a fraction of the spots depending on the treatment cases. However, the information of those spots is still valuable in that, in the multi-slit prompt-gamma camera, the distal dose falloff of the spots is located solely based on prompt gamma measurement, i.e., not referring to Monte Carlo simulation. (C) 2019 Korean Nuclear Society, Published by Elsevier Korea LLC.
引用
收藏
页码:1406 / 1416
页数:11
相关论文
共 50 条
  • [31] Time-resolved imaging of prompt-gamma rays for proton range verification using a knife-edge slit camera based on digital photon counters
    Lopes, Patricia Cambraia
    Clementel, Enrico
    Crespo, Paulo
    Henrotin, Sebastien
    Huizenga, Jan
    Janssens, Guillaume
    Parodi, Katia
    Prieels, Damien
    Roellinghoff, Frauke
    Smeets, Julien
    Stichelbaut, Frederic
    Schaart, Dennis R.
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (15): : 6063 - 6085
  • [32] Assessment of Geant4 Prompt-Gamma Emission Yields in the Context of Proton Therapy Monitoring
    Pinto, Marco
    Dauvergne, Denis
    Freud, Nicolas
    Krimmer, Jochen
    Letang, Jean M.
    Testa, Etienne
    FRONTIERS IN ONCOLOGY, 2016, 6
  • [33] Prompt gamma imaging of passively shaped proton fields with a knife-edge slit camera
    Priegnitz, M.
    Barczyk, S.
    Keitz, I.
    Mein, S.
    Stappen, F. V.
    Janssens, G.
    Hotoiu, L.
    Smeets, J.
    Fiedler, F.
    Prieels, D.
    Enghardt-, W.
    Pausch, G.
    Richter, C.
    RADIOTHERAPY AND ONCOLOGY, 2016, 118 : S86 - S87
  • [34] Prompt Gamma Measurements for the Verification of Dose Deposition in Proton Therapy
    Kim, J.
    Kubo, H.
    Tanimori, T.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [35] Prompt-Gamma Spatial Energy-Distribution in a Proton Therapy Plan for Beam Range Verification
    Vella, A.
    Fiorini, F.
    Van den Heuvel, F.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2020, 108 (03): : E281 - E281
  • [36] Computational model for detector timing effects in Compton-camera based prompt-gamma imaging for proton radiotherapy
    Maggi, Paul
    Peterson, Steve
    Panthi, Rajesh
    Mackin, Dennis
    Yang, Hao
    He, Zhong
    Beddar, Sam
    Polf, Jerimy
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (12):
  • [37] Prompt-gamma imaging for prostate cancer proton therapy: CNN-based detection of anatomical changes
    Pietsch, J.
    Piplack, N.
    Berthold, J.
    Khamfongkhruea, C.
    Thiele, J.
    Hoelscher, T.
    Traneus, E.
    Janssens
    Smeets, J.
    Stuetzer, K.
    Loeck, S.
    Richter, C.
    RADIOTHERAPY AND ONCOLOGY, 2022, 170 : S546 - S548
  • [38] Comparison of knife-edge and multi-slit camera for proton beam range verification by Monte Carlo simulation
    Park, Jong Hoon
    Kim, Sung Hun
    Ku, Youngmo
    Lee, Hyun Su
    Kim, Chan Hyeong
    Shin, Dong Ho
    Jeong, Jong Hwi
    NUCLEAR ENGINEERING AND TECHNOLOGY, 2019, 51 (02) : 533 - 538
  • [39] Experimental evaluation of depth-encoding absorber designs for prompt-gamma Compton imaging in proton therapy
    Tsai, Hui-Yu
    Chen, Hsin-Yu
    Lee, Ming-Wei
    Wang, Ze
    Tseng, Sheng-Pin
    Hong, Ji-Hong
    Jan, Meei-Ling
    RADIATION MEASUREMENTS, 2019, 127
  • [40] Development of a Compton Camera for Online Range Monitoring of Laser-Accelerated Proton Beams via Prompt-Gamma Detection
    Thirolf, P. G.
    Lang, C.
    Aldawood, S.
    v.d. Kolff, H. G.
    Maier, L.
    Schaart, D. R.
    Parodi, K.
    INPC 2013 - INTERNATIONAL NUCLEAR PHYSICS CONFERENCE, VOL. 2, 2014, 66