Intraoperative aberrant bispectral index values due to facial nerve monitoring

被引:12
|
作者
Chakrabarti, Dhritiman [1 ]
Surve, Rohini Mrugank [1 ]
Deepti, B. S. [1 ]
Masapu, Dheeraj [1 ]
机构
[1] Natl Inst Mental Hlth & Neurosci, Dept Neuroanaesthesia, 3rd Floor,Fac Block,Hosur Rd, Bangalore 560029, Karnataka, India
关键词
Bispectral index; Artefact; Facial nerve monitoring;
D O I
10.1016/j.jclinane.2016.11.005
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Bispectral index is an accepted depth of anaesthesia monitor for guiding intraoperative hypnotic agent administration. Frontalis EMG displayed on BIS monitor may increase due to twitching of frontalis muscle. EMG increases are also known to cause artefactual increases in BIS values. We report a case of artefactual increase of EMG and subsequently BIS values, due to electrical artefact from cranial nerve stimulator being used to identify the facial nerve. An explanation of the effect of stimulator signal on BIS EMG and BIS values has been provided. (C) 2016 Elsevier Inc All rights reserved.
引用
收藏
页码:61 / 62
页数:2
相关论文
共 50 条
  • [1] Monitoring by the bispectral index: are the target values respected?
    Passot, S.
    Heritier, F.
    Billard, V.
    Mavoungou, P.
    Payen, J. -F.
    Sztark, F.
    Molliex, S.
    [J]. ANNALES FRANCAISES D ANESTHESIE ET DE REANIMATION, 2014, 33 : A143 - A144
  • [2] Practice patterns in the intraoperative use of bispectral index monitoring
    Gelfand, Melanie E.
    Gabriel, Rodney A.
    Gimlich, Robert
    Beutler, Sascha S.
    Urman, Richard D.
    [J]. JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2017, 31 (02) : 281 - 289
  • [3] Practice patterns in the intraoperative use of bispectral index monitoring
    Melanie E. Gelfand
    Rodney A. Gabriel
    Robert Gimlich
    Sascha S. Beutler
    Richard D. Urman
    [J]. Journal of Clinical Monitoring and Computing, 2017, 31 : 281 - 289
  • [4] INTRAOPERATIVE FACIAL-NERVE MONITORING
    GANTZ, BJ
    [J]. AMERICAN JOURNAL OF OTOLOGY, 1985, : 58 - 61
  • [5] INTRAOPERATIVE FACIAL-NERVE MONITORING
    SILVERSTEIN, H
    ROSENBERG, S
    [J]. OTOLARYNGOLOGIC CLINICS OF NORTH AMERICA, 1991, 24 (03) : 709 - 725
  • [6] INTRAOPERATIVE MONITORING OF THE FACIAL-NERVE
    HARNER, SG
    DAUBE, JR
    BEATTY, CW
    EBERSOLD, MJ
    [J]. LARYNGOSCOPE, 1988, 98 (02): : 209 - 212
  • [7] Spurious bispectral index values due to electromyographic activity
    Baldesi, O
    Bruder, N
    Velly, L
    Gouin, E
    [J]. EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2004, 21 (04) : 324 - 325
  • [8] Continuous intraoperative bispectral index monitoring and perioperative outcome in children
    Johansen, JW
    [J]. ANESTHESIA AND ANALGESIA, 1998, 86 (2S):
  • [9] Intraoperative video-monitoring of the facial nerve
    De Seta, E
    Bertoli, GA
    Pichi, B
    Nostro, G
    Filipo, R
    [J]. 4TH EUROPEAN CONGRESS OF OTO-RHINO-LARYNGOLOGY HEAD AND NECK SURGERY, VOLS 1 AND 2, 2000, : 299 - 303
  • [10] Facial nerve dysfunction following parotidectomy: role of intraoperative facial nerve monitoring
    José João Ruas
    Jorge Rodrigues
    Manuel Ribeiro
    Carla Pinto Moura
    [J]. European Archives of Oto-Rhino-Laryngology, 2023, 280 : 1479 - 1484