Bispectral index is an accepted depth of anaesthesia monitor for guiding intraoperative hypnotic agent administration. Frontalis EMG displayed on BIS monitor may increase due to twitching of frontalis muscle. EMG increases are also known to cause artefactual increases in BIS values. We report a case of artefactual increase of EMG and subsequently BIS values, due to electrical artefact from cranial nerve stimulator being used to identify the facial nerve. An explanation of the effect of stimulator signal on BIS EMG and BIS values has been provided. (C) 2016 Elsevier Inc All rights reserved.
引用
下载
收藏
页码:61 / 62
页数:2
相关论文
共 50 条
  • [21] Intraoperative facial nerve monitoring: Personal experience and considerations
    Piccirillo, E
    Nakao, Y
    Taibah, A
    Falcioni, M
    Burdo, S
    Mancini, F
    Sanna, M
    THIRD INTERNATIONAL CONFERENCE ON ACOUSTIC NEURINOMA AND OTHER CPA TUMORS, 1999, : 429 - 432
  • [22] Falsely increased Bispectral Index™ due to neuromuscular transmission monitoring
    Katerenchuk, V.
    Calcada, A.
    Batista, A. C.
    Cordeiro, L.
    ANAESTHESIA REPORTS, 2023, 11 (02)
  • [23] Unusual case of low bispectral index values due to electrocardiographic interferences
    Hemmerling, T. M.
    Charabati, S.
    Bracco, D.
    BRITISH JOURNAL OF ANAESTHESIA, 2008, 101 (06) : 877 - 878
  • [24] Bispectral Index Monitoring
    Akinci, Seda Banu
    Celebioglu, Bilge
    JOURNAL OF THE TURKISH SOCIETY OF INTENSIVE CARE-TURK YOGUN BAKM DERNEGI DERGISI, 2006, 4 (02): : 85 - 90
  • [25] Bispectral Index monitoring
    Dawson, JS
    ANAESTHESIA AND INTENSIVE CARE, 2004, 32 (05) : 721 - 721
  • [26] Narcotrend and bispectral index for monitoring intraoperative anesthetic depth in patients with severe burns
    Zhang, Yongqiang
    Yang, Shuangmei
    Liu, Jun
    Wang, Yunfei
    Gao, Baoqin
    Cui, Wenjing
    Cao, Shu'e
    Zhang, Wei
    INTERNATIONAL JOURNAL OF CLINICAL AND EXPERIMENTAL MEDICINE, 2016, 9 (11): : 22020 - 22026
  • [27] Useful Method for Intraoperative Monitoring of Facial Nerve in a Scarred Bed
    Aysal, Bilge Kagan
    Yapici, Abdulkerim
    Bayram, Yalcin
    Zor, Fatih
    JOURNAL OF CRANIOFACIAL SURGERY, 2016, 27 (07) : 1804 - 1805
  • [28] INTRAOPERATIVE FACIAL-NERVE MONITORING IN ACOUSTIC NEUROMA SURGERY
    SILVERSTEIN, H
    ROSENBERG, SI
    FLANZER, J
    SEIDMAN, MD
    AMERICAN JOURNAL OF OTOLOGY, 1993, 14 (06): : 524 - 532
  • [29] INTRAOPERATIVE ELECTROPHYSIOLOGICAL MONITORING OF THE FACIAL-NERVE - IS IT STANDARD OF PRACTICE
    ROLAND, PS
    MEYERHOFF, WL
    AMERICAN JOURNAL OF OTOLARYNGOLOGY, 1994, 15 (04) : 267 - 270
  • [30] Intraoperative monitoring of the facial nerve. Vestibular schwannoma surgery
    Prell, J.
    Strauss, C.
    Plontke, S. K.
    Rampp, S.
    HNO, 2017, 65 (05) : 404 - 412