Magnetic resonance imaging of the cervical spine: comparison of 2D T2-weighted turbo spin echo, 2D T2*weighted gradient-recalled echo and 3D T2-weighted variable flip-angle turbo spin echo sequences

被引:52
|
作者
Meindl, T. [1 ]
Wirth, S. [1 ]
Weckbach, S. [3 ]
Dietrich, O. [2 ]
Reiser, M. [2 ]
Schoenberg, S. O. [3 ]
机构
[1] Univ Hosp Munich, Dept Clin Radiol, D-80336 Munich, Germany
[2] Univ Hosp Munich, Dept Clin Radiol, D-81377 Munich, Germany
[3] Univ Heidelberg, Med Fac Mannheim, Univ Hosp Mannheim, Dept Clin Radiol, D-68167 Mannheim, Germany
关键词
3D MRI; Cervical spine; T2w; SPACE; SUSCEPTIBILITY ARTIFACTS; SINGLE-SLAB; MR-IMAGES; EFFICIENCY; QUALITY; PULSE; CT;
D O I
10.1007/s00330-008-1175-7
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To compare an isotropic three-dimensional (3D) high-resolution T2-weighted (w) MR sequence and its reformations with conventional sequences for imaging of the cervical spine. Fifteen volunteers were examined at 1.5 T using sagittal and axial 3D T2-w, sagittal and axial 2D T2w, and axial 2D T2*w MR sequences. Axial reformations of the sagittal 3D dataset were generated (3D MPR T2w). Signal-to-noise and image homogeneity were evaluated in a phantom and in vivo. Visibility of ten anatomical structures of the cervical spine was evaluated. Artifacts were assessed. For statistical analysis, Cohen's kappa, Wilcoxon matched pairs, and t-testing were utilized. There were no significant differences in homogeneity between the sequences. Sagittal 3D T2w enabled better delineation of nerve roots, neural foramina, and intraforaminal structures compared to sagittal 2D T2w. Axial 3D T2w and axial 3D MPR T2w resulted in superior visibility of most anatomical structures compared to axial 2D T2w and comparable results to 2D T2*w concerning the spinal cord, nerve roots, intraforaminal structures, and fat. Artifacts were most pronounced in axial 2D T2w and axial 3D T2w. Acquisition of a 3D T2w data set is feasible in the cervical spine with superior delineation of anatomical structures compared to 2D sequences.
引用
收藏
页码:713 / 721
页数:9
相关论文
共 50 条
  • [1] Magnetic resonance imaging of the cervical spine: comparison of 2D T2-weighted turbo spin echo, 2D T2*weighted gradient-recalled echo and 3D T2-weighted variable flip-angle turbo spin echo sequences
    T. Meindl
    S. Wirth
    S. Weckbach
    O. Dietrich
    M. Reiser
    S. O. Schoenberg
    [J]. European Radiology, 2009, 19 : 713 - 721
  • [2] A comparison study between 3D T2-weighted SPACE and conventional 2D T2-weighted turbo spin echo in assessment of carotid plaque
    Lv, Peng
    Dai, Yuanyuan
    Lin, Jiang
    Zhang, Weisheng
    Liu, Hao
    Liu, Hui
    Tang, Xiao
    [J]. INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2017, 33 (03): : 395 - 400
  • [3] A comparison study between 3D T2-weighted SPACE and conventional 2D T2-weighted turbo spin echo in assessment of carotid plaque
    Peng Lv
    Yuanyuan Dai
    Jiang Lin
    Weisheng Zhang
    Hao Liu
    Hui Liu
    Xiao Tang
    [J]. The International Journal of Cardiovascular Imaging, 2017, 33 : 395 - 400
  • [4] 3D T2-Weighted Spin Echo Imaging in the Breast
    Moran, Catherine J.
    Hargreaves, Brian A.
    Saranathan, Manojkumar
    Lipson, Jafi A.
    Kao, Jennifer
    Ikeda, Debra M.
    Daniel, Bruce L.
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2014, 39 (02) : 332 - 338
  • [5] Abdominal T2-Weighted Imaging and T2 Mapping Using a Variable Flip Angle Radial Turbo Spin-Echo Technique
    Keerthivasan, Mahesh B.
    Galons, Jean-Philippe
    Johnson, Kevin
    Umapathy, Lavanya
    Martin, Diego R.
    Bilgin, Ali
    Altbach, Maria, I
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2022, 55 (01) : 289 - 300
  • [6] 3D isotropic T2-weighted fast spin echo (VISTA) versus 2D T2-weighted fast spin echo in evaluation of the calcaneofibular ligament in the oblique coronal plane
    Park, H. J.
    Lee, S. Y.
    Choi, Y. J.
    Hong, H. P.
    Park, S. J.
    Park, J. H.
    Kim, E.
    [J]. CLINICAL RADIOLOGY, 2017, 72 (02) : 176.e1 - 176.e7
  • [7] MRI of the regions of inner ear and cerebellopontine angle using a 3D T-2-weighted turbo spin-echo sequence: Comparison with conventional 2D T-2-weighted turbo spin-echo sequences and T-1-weighted spin-echo sequences.
    Czerny, C
    Trattnig, S
    Baumgartner, WD
    Gstottner, W
    Imhof, H
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1997, 167 (04): : 377 - 383
  • [8] Uterine Tumors: Comparison of 3D versus 2D T2-weighted Turbo Spin-Echo MR Imaging at 3.0 T-Initial Experience
    Hori, Masatoshi
    Kim, Tonsok
    Onishi, Hiromitsu
    Ueguchi, Takashi
    Tatsumi, Mitsuaki
    Nakamoto, Atsushi
    Tsuboyama, Takahiro
    Tomoda, Kaname
    Tomiyama, Noriyuki
    [J]. RADIOLOGY, 2011, 258 (01) : 154 - 163
  • [9] Preoperative detection of malignant liver tumors: Comparison of 3D-T2-weighted sequences with T2-weighted turbo spin-echo and single shot T2 at 1.5 T
    Barat, Maxime
    Soyer, Philippe
    Dautry, Raphael
    Pocard, Marc
    Lo-Dico, Rea
    Najah, Haythem
    Eveno, Clarisse
    Cassinotto, Christophe
    Dohan, Anthony
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2018, 100 : 7 - 13
  • [10] An efficient 3D stack-of-stars turbo spin echo pulse sequence for simultaneous T2-weighted imaging and T2 mapping
    Keerthivasan, Mahesh Bharath
    Saranathan, Manojkumar
    Johnson, Kevin
    Fu, Zhiyang
    Weinkauf, Craig C.
    Martin, Diego R.
    Bilgin, Ali
    Altbach, Maria I.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2019, 82 (01) : 326 - 341