Pneumatic tube transport of blood samples affects global hemostasis and platelet function assays

被引:9
|
作者
Lorenzen, Henriette [1 ]
Frostrup, Ann-Britt [2 ,3 ]
Larsen, Anja S. [4 ]
Fenger, Michelle S. [1 ,2 ]
Dahdouh, Sanne [1 ,4 ]
Zoel-Ghina, Randa [2 ]
Nielsen, Leif K. [1 ]
机构
[1] Univ Coll Copenhagen, Fac Hlth, Sigurdsgade 26, DK-2200 Copenhagen N, Denmark
[2] Rigshosp, Dept Clin Immunol, Copenhagen, Denmark
[3] Zealand Univ Hosp, Dept Clin Immunol, Roskilde, Denmark
[4] Nordsjaellands Hosp, Dept Clin Immunol, Hillerod, Denmark
关键词
global hemostasis testing; platelet aggregation (Multiplate); pneumatic tube transport; thrombelastography (TEG); MULTIPLE ELECTRODE AGGREGOMETRY; ROUTINE HEMATOLOGY; SYSTEM TRANSPORT; DELIVERY-SYSTEM; SHEAR-STRESS; WHOLE-BLOOD; HEMOLYSIS; PARAMETERS; THROMBELASTOGRAPHY; COAGULATION;
D O I
10.1111/ijlh.13470
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction Pneumatic tube systems (PTS) are frequently used for rapid and cost-effective transportation of blood samples to the clinical laboratory. The impact of PTS transport on platelet function measured by the Multiplate system and global hemostasis measured by the TEG 5000 was evaluated. Methods Paired samples from healthy adult individuals were obtained at two study sites: Rigshospitalet (RH) and Nordsj AE llands Hospital (NOH). One sample was transported by PTS and one manually (non-PTS). Platelet function was assessed by platelet aggregation (Multiplate) and global hemostasis was assessed by a variety of thrombelastography (TEG) assays. Multiplate (n = 39) and TEG (n = 32) analysis was performed at site RH, and Multiplate (n = 28) analysis was performed at site NOH. Results A significant higher agonist-induced platelet aggregation was found for PTS samples compared to manual transport at site NOH (P < .02, all agonists). No significant difference was found at site RH (P > .05, all agonists). For Kaolin TEG, samples transported by PTS showed a significant lower R-time and higher Angle (P < .001). No significant differences in MA and LY30 was found (P > .05). ACT of RapidTEG was significantly reduced (P = .001) and MA of Functional Fibrinogen TEG was significantly increased (P < .001) after PTS transport. No significant impact of PTS was observed for TEG assays with heparinase (P > .05). Conclusions Depending on the type of PTS, transportation by PTS affected platelet aggregation measured by Multiplate. Furthermore, PTS alters TEG parameters possibly reflecting coagulation factors. Clinical laboratories should evaluate the effect of the local PTS on Multiplate and TEG results.
引用
收藏
页码:1207 / 1215
页数:9
相关论文
共 50 条
  • [31] The effect of a pneumatic tube transport system on PFA-100™ closure time and whole blood platelet aggregation
    Dyszkiewicz-Korpanty, A
    Quinton, R
    Yassine, J
    Sarode, R
    JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2004, 2 (02) : 354 - 356
  • [32] Validation of blood components transport through a pneumatic tube system
    Garcia, Lais Oliveira
    Speransa, Daniela Michelim Rodriguez
    Rodrigues, Cristiane Braga
    Benites, Rosimari Melgarejo
    Garcia, Mariana Timm
    Sekine, Leo
    Franz, Juliana Pires Marafon
    HEMATOLOGY TRANSFUSION AND CELL THERAPY, 2022, 44 (04) : 519 - 525
  • [33] Validation of blood component transport through a pneumatic tube system
    Novak, D.
    Bakhirev, A.
    Crookston, K.
    TRANSFUSION, 2008, 48 (02) : 267A - 267A
  • [34] IMPLEMENTATION OF A PNEUMATIC-TUBE SYSTEM FOR TRANSPORT OF BLOOD SPECIMENS
    POZNANSKI, W
    SMITH, F
    BODLEY, F
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1978, 70 (02) : 291 - 295
  • [35] EFFECT OF SAMPLE TRANSPORT BY TEMPUS600 PNEUMATIC TUBE SYSTEM ON PLATELETS, PLATELET FUNCTION TESTS AND THROMBOELASTOMETRY
    De Kesel, Pieter
    Devreese, Katrien
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2019, 41 : 49 - 49
  • [36] Effects of Pneumatic Tube Transport on Blood Gas and Supplemental Analytes
    Jensen, Phillip R.
    Aramaki, Ryan L.
    Mannebach, Steve C.
    Markewitz, Boaz A.
    RESPIRATORY CARE, 2021, 66 (10) : 1567 - 1571
  • [37] IMMUNOLOGICAL STUDIES OF BLOOD-SAMPLES TRANSPORTED BY A PNEUMATIC TUBE SYSTEM
    GREENDYK, RM
    BANZHAF, JC
    PELYSKO, S
    BAUMAN, B
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1977, 68 (04) : 508 - 510
  • [38] Pre-analytical effects of pneumatic tube transport on impedance platelet aggregometry
    Bolliger, Daniel
    Seeberger, Manfred D.
    Tanaka, Kenichi A.
    Dell-Kuster, Salome
    Gregor, Michael
    Zenklusen, Urs
    Grapow, Martin
    Tsakiris, Dimitrios A.
    Filipovic, Miodrag
    PLATELETS, 2009, 20 (07) : 458 - 465
  • [39] Pre-analytical effects of pneumatic tube transport on impedance platelet aggregometry
    Bolliger, D.
    Seeberger, M. D.
    Gregor, M.
    Zenklusen, U.
    Grapow, M.
    Filipovic, M.
    SWISS MEDICAL WEEKLY, 2009, 139 (41-42) : 18S - 18S
  • [40] Gentle blood aspiration and tube cushioning reduce pneumatic tube system interference in lactate dehydrogenase assays
    Strubi-Vuillaume, Isabelle
    Carlier, Valentine
    Obeuf, Catherine
    Vasseur, Francis
    Maury, J-Claude
    Maboudou, Patrice
    Mangalaboyi, Jacques
    Durocher, Alain
    Launay, David
    Noel, Christian
    Brousseau, Thierry
    ANNALS OF CLINICAL BIOCHEMISTRY, 2016, 53 (02) : 295 - 297