Pneumatic tube transport of blood samples affects global hemostasis and platelet function assays

被引:9
|
作者
Lorenzen, Henriette [1 ]
Frostrup, Ann-Britt [2 ,3 ]
Larsen, Anja S. [4 ]
Fenger, Michelle S. [1 ,2 ]
Dahdouh, Sanne [1 ,4 ]
Zoel-Ghina, Randa [2 ]
Nielsen, Leif K. [1 ]
机构
[1] Univ Coll Copenhagen, Fac Hlth, Sigurdsgade 26, DK-2200 Copenhagen N, Denmark
[2] Rigshosp, Dept Clin Immunol, Copenhagen, Denmark
[3] Zealand Univ Hosp, Dept Clin Immunol, Roskilde, Denmark
[4] Nordsjaellands Hosp, Dept Clin Immunol, Hillerod, Denmark
关键词
global hemostasis testing; platelet aggregation (Multiplate); pneumatic tube transport; thrombelastography (TEG); MULTIPLE ELECTRODE AGGREGOMETRY; ROUTINE HEMATOLOGY; SYSTEM TRANSPORT; DELIVERY-SYSTEM; SHEAR-STRESS; WHOLE-BLOOD; HEMOLYSIS; PARAMETERS; THROMBELASTOGRAPHY; COAGULATION;
D O I
10.1111/ijlh.13470
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction Pneumatic tube systems (PTS) are frequently used for rapid and cost-effective transportation of blood samples to the clinical laboratory. The impact of PTS transport on platelet function measured by the Multiplate system and global hemostasis measured by the TEG 5000 was evaluated. Methods Paired samples from healthy adult individuals were obtained at two study sites: Rigshospitalet (RH) and Nordsj AE llands Hospital (NOH). One sample was transported by PTS and one manually (non-PTS). Platelet function was assessed by platelet aggregation (Multiplate) and global hemostasis was assessed by a variety of thrombelastography (TEG) assays. Multiplate (n = 39) and TEG (n = 32) analysis was performed at site RH, and Multiplate (n = 28) analysis was performed at site NOH. Results A significant higher agonist-induced platelet aggregation was found for PTS samples compared to manual transport at site NOH (P < .02, all agonists). No significant difference was found at site RH (P > .05, all agonists). For Kaolin TEG, samples transported by PTS showed a significant lower R-time and higher Angle (P < .001). No significant differences in MA and LY30 was found (P > .05). ACT of RapidTEG was significantly reduced (P = .001) and MA of Functional Fibrinogen TEG was significantly increased (P < .001) after PTS transport. No significant impact of PTS was observed for TEG assays with heparinase (P > .05). Conclusions Depending on the type of PTS, transportation by PTS affected platelet aggregation measured by Multiplate. Furthermore, PTS alters TEG parameters possibly reflecting coagulation factors. Clinical laboratories should evaluate the effect of the local PTS on Multiplate and TEG results.
引用
收藏
页码:1207 / 1215
页数:9
相关论文
共 50 条
  • [21] Hemostasis Laboratory: Assays for Platelet Function and Von Willebrand Disease
    Anupama Kohli
    Fauzia Khan
    L. Michael Snyder
    L. Pechet
    Journal of Thrombosis and Thrombolysis, 1998, 6 : 159 - 167
  • [22] Pneumatic tube system for blood products transport
    Bardiaux, L.
    Boulanger, E.
    des Floris, M. -F. Leconte
    Bourcier, V.
    Cottier, D.
    TRANSFUSION CLINIQUE ET BIOLOGIQUE, 2012, 19 (4-5) : 195 - 198
  • [23] Platelet concentrate transport in pneumatic tube systems - does it work?
    Lance, M. D.
    Marcus, M. A. E.
    van Oerle, R.
    Theunissen, H. M. S.
    Henskens, Y. M. C.
    VOX SANGUINIS, 2012, 103 (01) : 79 - 82
  • [24] EFFECT OF TRANSPORT BY TUBE ON BLOOD SAMPLES
    COLLOMBEL, C
    BUREAU, J
    BALTASSAT, P
    ANNALES DE BIOLOGIE CLINIQUE, 1975, 33 (03) : 241 - 241
  • [25] Pneumatic tube system for transport of laboratory samples: preanalytical aspects
    Toth Judit
    Lenkey Agota
    Olah, V. Anna
    Koeteles Julianna
    Sziraki, Kissne Valeria
    Kerenyi Adrienne
    Kappelmayer Janos
    ORVOSI HETILAP, 2014, 155 (28) : 1113 - 1120
  • [26] THE EFFECT OF PNEUMATIC TUBE SYSTEMS ON THE HEMOLYSIS OF BIOCHEMISTRY BLOOD SAMPLES
    Cakirca, Gokhan
    Erdal, Huseyin
    JOURNAL OF EMERGENCY NURSING, 2017, 43 (03) : 255 - 258
  • [27] Changes in blood gas samples produced by a pneumatic tube system
    Collinson, PO
    John, CM
    Gaze, DC
    Ferrigan, LF
    Cramp, DG
    JOURNAL OF CLINICAL PATHOLOGY, 2002, 55 (02) : 105 - 107
  • [28] Impact of Pneumatic Tube System Transport of Whole Blood Samples on Free DNA Concentration in Human Plasma
    Kriegshaeuser, Gernot
    Wagner, Carina
    Mangge, Harald
    Halwachs-Baumann, Gabriele
    Enko, Dietmar
    CLINICAL CHEMISTRY, 2016, 62 (12) : 1672 - 1673
  • [29] Comparison of standard and global hemostasis assays in cord and peripheral blood of newborns
    Arutunyan, Belinda V.
    Koltsova, Ekaterina M.
    Shpilyuk, Margarita A.
    Lipets, Elena N.
    Timofeeva, Leila A.
    Karavaeva, Anna L.
    Balashova, Ekaterina N.
    Krechetova, Liubov V.
    Ataullakhanov, Fazoil I.
    Zubkov, Viktor V.
    Degtyarev, Dmitry N.
    Sukhikh, Gennady T.
    Tetruashvili, Nana K.
    Balandina, Anna N.
    PEDIATRIC RESEARCH, 2024,
  • [30] Validation of a pneumatic tube system to transport surgical pathology biopsy samples
    Guibourg, Briac
    Marcorelles, Pascale
    Uguen, Arnaud