Biochemical assay for serum creatinine detection through a 1-methylhydantoin and cobalt complex

被引:12
|
作者
Dasgupta, Pallavi [1 ]
Kumar, Vinay [2 ]
Krishnaswamy, Patnam R. [1 ]
Bhat, Navakanta [1 ]
机构
[1] Indian Inst Sci, Ctr Nanosci & Engn, Bengaluru, India
[2] PathShodh Healthcare Pvt Ltd, Bengaluru, India
关键词
BIOSENSOR; UREA;
D O I
10.1039/d0ra06470j
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Creatinine is a reliable indicator of renal function and degradation of muscular metabolism. Current analytical techniques for its measurement are limited by their cost and requirement of sophisticated instruments. In this work, we report a highly sensitive amperometric biosensor for creatinine by utilizing the one-step selective conversion of creatinine by creatinine deiminase. The novelty of the proposed sensor relies on the measurement of N-methylhydantoin produced in the reaction. The sensing chemistry comprises of creatinine deiminase as the receptor for creatinine, cobalt chloride as the electrochemically active recognition element, and methylene blue as the redox mediator. The sensing chemistry is immobilized on the glassy carbon electrode surface by physisorption. We have been able to provide a standalone device that reliably quantifies creatinine in serum and even whole blood, without any sample pre-processing. It is possible to measure creatinine in the clinically relevant range from 0.8 to 4 mg dL(-1) with this approach.
引用
收藏
页码:39092 / 39101
页数:10
相关论文
共 50 条
  • [1] Serum Creatinine Electrochemical Biosensor on Printed Electrodes Using Monoenzymatic Pathway to 1-Methylhydantoin Detection
    Dasgupta, Pallavi
    Kumar, Vinay
    Krishnaswamy, Patnam R.
    Bhat, Navakanta
    ACS OMEGA, 2020, 5 (35): : 22459 - 22464
  • [2] Metallophilic interactions in polynuclear Ag(I) complex with 1-methylhydantoin studied by X-ray absorption, electronic and vibrational spectroscopies
    Puszynska-Tuszkanow, Mariola
    Staszak, Zbigniew
    Misiaszek, Tomasz
    Klepka, Marcin T.
    Wolska, Anna
    Drzewiecka-Antonik, Aleksandra
    Faltynowicz, Hanna
    Cieslak-Golonka, Maria
    CHEMICAL PHYSICS LETTERS, 2014, 597 : 94 - 98
  • [3] Impact of the biochemical assay for serum creatinine measurement on the individual carboplatin dosing:: A prospective study
    Léger, F
    Séronie-Vivien, S
    Makdessi, J
    Lochon, I
    Delord, JP
    Sarda, C
    Canal, P
    Chatelut, E
    EUROPEAN JOURNAL OF CANCER, 2002, 38 (01) : 52 - 56
  • [4] DETECTION OF GILVOCARCIN ANTI-TUMOR COMPLEX BY A BIOCHEMICAL INDUCTION ASSAY (BIA)
    WEI, TT
    CHAN, JA
    ROLLER, PP
    WEISS, U
    STROSHANE, RM
    WHITE, RJ
    BYRNE, KM
    JOURNAL OF ANTIBIOTICS, 1982, 35 (04): : 529 - 532
  • [5] An artificial enzyme-based assay: DNA detection using a peroxidase-like copper-creatinine complex
    Singh, Amardeep
    Patra, Srikanta
    Lee, Jeong-Ah
    Park, Kang Hyun
    Yang, Haesik
    BIOSENSORS & BIOELECTRONICS, 2011, 26 (12): : 4798 - 4803
  • [6] Biophotonic sensor for the detection of creatinine concentration in blood serum based on 1D photonic crystal
    Aly, Arafa H.
    Mohamed, Doaa
    A. Mohaseb, Mona
    El-Gawaad, N. S. Abd
    Trabelsi, Y.
    RSC ADVANCES, 2020, 10 (53) : 31765 - 31772
  • [7] Detection of aberrant methylation of HOXA9 and HIC1 through multiplex MethyLight assay in serum DNA for the early detection of epithelial ovarian cancer
    Singh, Alka
    Gupta, Sameer
    Badarukhiya, Jaydeep A.
    Sachan, Manisha
    INTERNATIONAL JOURNAL OF CANCER, 2020, 147 (06) : 1740 - 1752
  • [8] Serum SP1, HPL levels and urinary estrogen/creatinine ratio in the detection of low birth weight
    Zhang Wei-yuan
    Teng Hong
    Wang Li-yan
    Yan Guo-lai
    CHINESEMEDICALJOURNAL, 1993, (06)
  • [9] Efficient Synthesis of Quinazoline-2,4(1H,3H)-dione via Simultaneous Activated CO2 and 2-Aminobenzonitrile by 1-Methylhydantoin Anion-Functionalized Ionic Liquid through the Multiple-Site Cooperative Interactions
    Chen, Tingting
    Zhang, Yunfei
    Xu, Yingjie
    ACS SUSTAINABLE CHEMISTRY & ENGINEERING, 2022, 10 (32): : 10699 - 10711
  • [10] CEREBROSPINAL-FLUID AND SERUM IMMUNE-COMPLEX IN ACUTE INFLAMMATORY POLYNEURITIS - DETECTION BY CLQ BINDING ASSAY
    ANTOINE, JC
    MICHEL, D
    LAMELIN, JP
    LAURENT, B
    SCHOTT, B
    ACTA NEUROLOGICA SCANDINAVICA, 1986, 73 (05): : 477 - 480