Experimental investigation of TRS-483 reference dosimetry correction factors for Leksell Gamma Knife® Icon™ beams

被引:5
|
作者
Hager, Wille [1 ]
Kaveckyte, Vaiva [2 ,3 ]
Benmakhlouf, Hamza [3 ]
机构
[1] Stockholm Univ, Dept Phys, SE-10691 Stockholm, Sweden
[2] Linkoping Univ, Dept Med & Hlth Sci, Radiat Phys, SE-58185 Linkoping, Sweden
[3] Karolinska Univ Hosp, Dept Med Radiat Phys & Nucl Med, SE-17176 Stockholm, Sweden
关键词
code of practice; Leksell Gamma Knife; reference dosimetry; phantom; k(Qmsr; Q0)fmsr; fref; MONTE-CARLO; CALIBRATION-PHANTOM; SCATTER;
D O I
10.1002/mp.14561
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Radiosurgery using the Leksell Gamma Knife (R) (LGK) Icon (TM) is an established technique used for treating intracranial lesions. The largest beam field size the LGK Icon can produce is a 16 mm diameter sphere. Despite this, reference dosimetry on the LGK Icon is typically performed using ionization chambers calibrated in 10 x 10 cm(2) fields. Furthermore, plastic phantoms are widely used instead of liquid water phantoms. In an effort to resolve these issues, the International Atomic Energy Agency (IAEA) in collaboration with American Association of Physicists in medicine (AAPM) recently published Technical Report Series No. 483 (TRS-483) as a Code of Practice for small-field dosimetry. TRS-483 includes small-field correction factors, k(Qmsr,Q0)(fmsr,fref), intended to account for the differences between setups when using small-field modalities such as the LGK Icon, and conventional setups. Since the publication of TRS-483, at least three new sets of values of k(Qmsr,Q0)(fmsr,fref) for the LGK Icon have been published. The purpose of this study was to experimentally investigate the published values of k(Qmsr,Q0)(fmsr,fref) for commonly used phantom and ionization chamber (IC) models for the LGK Icon. Methods: Dose-rates from two LGK units were determined using acrylonitrile butadiene styrene (ABS) and Certified Medical Grade Solid Water (R) (SW) phantoms, and PTW 31010 and PTW 31016 ICs. Correction factors were applied, and the resulting dose-rates compared. Relative validity of the correction factors was investigated by taking the ratios of dose-rate correction factor products. Additionally, dose-rates from the individual sectors were determined in order to calculate the beam attenuation caused by the ABS phantom adapter. Results and Conclusions: It was seen that the dose-rate is underestimated by at least 1% when using the ABS phantom, which was attributed to fluence perturbation caused by the IC and phantom adapter. Published correction factors k(Qmsr,Q0)(fmsr,fref) account for these effects to varying degree and should be used. The SW phantom is unlikely to underestimate the dose-rate by more than 1%, and applying k(Qmsr,Q0)(fmsr,fref) could not be shown to be necessary. Out of the two phantom models, the ABS phantom is not recommended for use in LGK reference dosimetry. The use of newly published values of k(Qmsr,Q0)(fmsr,fref) should be considered. (c) 2020 The Authors. Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine
引用
收藏
页码:434 / 444
页数:11
相关论文
共 42 条
  • [21] Dosimetry of small static fields used in external photon beam radiotherapy: Summary of TRS-483, the IAEA-AAPM international Code of Practice for reference and relative dose determination
    Palmans, Hugo
    Andreo, Pedro
    Huq, M. Saiful
    Seuntjens, Jan
    Christaki, Karen E.
    Meghzifene, Ahmed
    MEDICAL PHYSICS, 2018, 45 (11) : E1123 - E1145
  • [22] Technical note: Consistency of IAEA's TRS-483 and AAPM's extended TG-51 protocols for clinical reference dosimetry of the CyberKnife M6 machine
    Duchaine, Jasmine
    Markel, Daniel
    Ley, Jean-Luc
    Beliveau-Nadeau, Dominique
    Zerouali, Karim
    Doucet, Robert
    Bouchard, Hugo
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2023, 24 (05):
  • [23] Determination of correction factors in small MLC-defined fields for the Razor and microSilicon diode detectors and evaluation of the suitability of the IAEA TRS-483 protocol for multiple detectors
    McGrath, Andrew N.
    Golmakani, Samane
    Williams, Timothy J.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (07):
  • [24] Precision dosimetry for narrow photon beams used in radiosurgery -: Determination of Gamma Knife® output factors
    Mack, A
    Scheib, SG
    Major, J
    Gianolini, S
    Pazmandi, G
    Feist, H
    Czempiel, H
    Kreiner, HJ
    MEDICAL PHYSICS, 2002, 29 (09) : 2080 - 2089
  • [25] GEANT4-based Monte Carlo Simulation of Beam Quality Correction Factors for the Leksell Gamma Knife® Perfexion™
    Thomas Schaarschmidt
    Tae Hoon Kim
    Yong Kyun Kim
    Hye Jeong Yang
    Hyun-Tai Chung
    Journal of the Korean Physical Society, 2018, 73 : 1814 - 1820
  • [26] GEANT4-based Monte Carlo Simulation of Beam Quality Correction Factors for the Leksell Gamma Knife® Perfexion
    Schaarschmidt, Thomas
    Kim, Tae Hoon
    Kim, Yong Kyun
    Yang, Hye Jeong
    Chung, Hyun-Tai
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2018, 73 (12) : 1814 - 1820
  • [27] Determination of field output correction factors of radiophotoluminescence glass dosimeter and CC01 ionization chamber and validation against IAEA-AAPM TRS-483 code of practice
    Yabsantia, Sumalee
    Suriyapee, Sivalee
    Phaisangittisakul, Nakorn
    Oonsiri, Sornjarod
    Sanghangthum, Taweap
    Mirzakhanian, Lalageh
    Heng, Veng Jean
    Seuntjens, Jan
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 88 : 167 - 174
  • [28] Evaluation of radiation detectors for the determination of field output factors in Leksell Gamma Knife dosimetry using 3D printed phantom inserts
    Perrett, Ben
    Seshadri, Venkatakrishnan
    Huxley, Cosmo
    Kumar, Sanjay
    Wawrzak, Michal
    Ramachandran, Prabhakar
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2024, 121
  • [29] Monte Carlo simulations to optimize beams used in Gamma experimental dosimetry of narrow Knife radio-surgery
    Lymperopoulou, G.
    Petrokokkinos, L.
    Papagiannis, P.
    Steiner, M.
    Spevacek, V.
    Semnicka, J.
    Dvorak, P.
    Selmenis, I.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 580 (01): : 548 - 551
  • [30] Current best estimates of beam quality correction factors for reference dosimetry of clinical proton beams
    Palmans, Hugo
    Lourenco, Ana
    Medin, Joakim
    Vatnitsky, Stanislav
    Andreo, Pedro
    PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (19):