Robust three-dimensional registration on optical coherence tomography angiography for speckle reduction and visualization

被引:15
|
作者
Cheng, Yuxuan [1 ]
Chu, Zhongdi [1 ]
Wang, Ruikang K. [1 ,2 ]
机构
[1] Univ Washington, Dept Bioengn, Seattle, WA 98195 USA
[2] Univ Washington, Dept Ophthalmol, Seattle, WA 98195 USA
基金
美国国家卫生研究院;
关键词
Optical coherence tomography (OCT); optical coherence tomography angiography (OCTA); motion artifact; speckle noise; averaging; 3D registration; CHORIOCAPILLARIS FLOW DEFICITS; NEOVASCULARIZATION; OPTIMIZATION; CONTRAST; DOMAIN; SPEED;
D O I
10.21037/qims-20-751
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: In the clinical applications of optical coherence tomography angiography (OCTA), the repeated scanning and averaging method can provide better contrast with reduced speckle noises in the final results, which are useful for visualizing and quantifying vascular components with high accuracy, reproducibility, and reliability. However, the inevitable patient motion presents a challenge to this method. The objective of this study is to meet this challenge by introducing a 3D registration method to register optical coherence tomography (OCT)/OCTA scans for precise volume averaging of multiple scans to improve the signal-to-noise ratio (SNR) and increase quantification accuracy. Methods: The proposed method utilized both rigid affine transformation and non-rigid B-spline transformation in which their parameters were optimized and calculated by the average stochastic gradient descent on OCT structural images. In addition, we also introduced a multi-level resolution approach to further improve the robustness and computational speed of our proposed method. The imaging performance was tested on in vivo imaging of human skin and eye and assessed by SNR, peak signal-to-noise ratio (PSNR) and normalized correlation coefficient (NCC). Results: Five subjects were enrolled in this study for obtaining in vivo images of skin and retina. The proposed registration and averaging method provided substantial improvements of the imaging performance in terms of vessel connectivity and signal to noise ratio. The increase of repeated volume numbers in the averaging improves all the metrics assessed, i.e., SNR, PSNR and NCC. An improvement of the SNR from 10 to 40 dB after 10 repeated volumetric averaging was achieved. Conclusions: The proposed 3D registration and averaging method is effective in reducing speckle noises and suppressing motion artifacts, thereby improving SNR, PSNR and NCC metrics for final averaged images. It is expected that the proposed algorithm would be practically useful in better visualization and more reliable quantification of in vivo OCT and OCTA data, which would be beneficial to OCT clinical applications.
引用
收藏
页码:879 / 894
页数:16
相关论文
共 50 条
  • [21] Three-dimensional optical coherence tomography for biomedical diagnosis
    Chan, KP
    2005 PACIFIC RIM CONFERENCE ON LASERS AND ELECTRO-OPTICS, 2005, : 42 - 43
  • [22] Three-dimensional endomicroscopy using optical coherence tomography
    Desmond C. Adler
    Yu Chen
    Robert Huber
    Joseph Schmitt
    James Connolly
    James G. Fujimoto
    Nature Photonics, 2007, 1 : 709 - 716
  • [23] Three-dimensional endomicroscopy using optical coherence tomography
    Adler, Desmond C.
    Chen, Yu
    Huber, Robert
    Schmitt, Joseph
    Connolly, James
    Fujimoto, James G.
    NATURE PHOTONICS, 2007, 1 (12) : 709 - 716
  • [24] Three-dimensional calibration targets for optical coherence tomography
    Sandrian, Michelle Gabriele
    Tomlins, Pete
    Woolliams, Peter
    Rasakanthan, Janarthanan
    Lee, Graham C. B.
    Yang, Anna
    Povazay, Boris
    Alex, Aneesh
    Sugden, Kate
    Drexler, Wolfgang
    OPTICAL DIAGNOSTICS AND SENSING XII: TOWARD POINT-OF-CARE DIAGNOSTICS AND DESIGN AND PERFORMANCE VALIDATION OF PHANTOMS USED IN CONJUNCTION WITH OPTICAL MEASUREMENT OF TISSUE IV, 2012, 8229
  • [25] Speckle reduction in optical coherence tomography imaging by affine-motion image registration
    Alonso-Caneiro, David
    Read, Scott A.
    Collins, Michael J.
    JOURNAL OF BIOMEDICAL OPTICS, 2011, 16 (11)
  • [26] Detection of Cystoid Macular Edema with Three-Dimensional Optical Coherence Tomography versus Fluorescein Angiography
    Ouyang, Yanling
    Keane, Pearse A.
    Sadda, Srinivas R.
    Walsh, Alexander C.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (10) : 5213 - 5218
  • [27] Normalized field autocorrelation function-based optical coherence tomography three-dimensional angiography
    Tang, Jianbo
    Erdener, Sefik Evren
    Sunil, Smrithi
    Boas, David A.
    JOURNAL OF BIOMEDICAL OPTICS, 2019, 24 (03)
  • [28] Three-dimensional detection of hemodynamic changes in skin microcirculation by optical coherence tomography-angiography
    Nakamichi, Yu
    BIORHEOLOGY, 2021, 58 (3-4) : 116 - 116
  • [29] Speckle noise reduction for optical coherence tomography
    Xiang, SH
    Zhou, L
    Schmitt, JM
    OPTICAL AND IMAGING TECHNIQUES FOR BIOMONITORING III, PROCEEDINGS OF, 1998, 3196 : 79 - 88
  • [30] Statistics and reduction of speckle in optical coherence tomography
    Bashkansky, M
    Reintjes, J
    OPTICS LETTERS, 2000, 25 (08) : 545 - 547