A comparison of manual counting of rabbit reticulocytes with ADVIA 2120i analyzer counting

被引:2
|
作者
Kaufhold, Annemarie E. [1 ,3 ]
Hirschberger, Johannes [1 ]
Reese, Sven [2 ]
Foerster, Gesine [3 ]
Hein, Jutta [3 ]
机构
[1] Ludwig Maximilians Univ Munchen, Ctr Clin Vet Med, Clin Small Anim Med, Munich, Germany
[2] Ludwig Maximilians Univ Munchen, Ctr Clin Vet Med, Dept Anat Histol & Embryol, Munich, Germany
[3] Synlab Vet Lab, Augsburg, Germany
关键词
EDTA; hematology; Li-heparin; method comparison; rabbits; reticulocyte maturation; HEMATOLOGY ANALYZER; VALUES; CATS; DOGS;
D O I
10.1177/1040638717750428
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
We compared manual counting of reticulocytes in rabbits with automatic counting using an ADVIA 2120i analyzer. Reproducibility and the influence of different anticoagulants (EDTA and Li-heparin) were also examined. Blood samples of 331 rabbits (method comparison, n = 289; reproducibility, n = 33; comparison of anticoagulants, n = 9) were tested. The reticulocyte numbers of each specimen were manually determined twice for method comparison. Passing-Bablok regressions, Bland-Altman plots, and the coefficient of variation (CV) were used to evaluate statistical significance. Good correlation (rs = 0.81) was observed between manual reticulocyte counting (groups 1-4) and the ADVIA 2120i. Quantification with the ADVIA 2120i was reproducible for relative reticulocyte numbers (EDTA, CV = 4.24%; Li-heparin, CV = 3.63%) and absolute reticulocyte numbers (EDTA, CV = 5.64%; Li-heparin, CV = 3.81%). The absolute and relative reticulocyte numbers were significantly higher in Li-heparin samples than in EDTA samples (absolute, p = 0.009; relative, p = 0.016). The ADVIA 2120i is suitable for counting reticulocytes in rabbit blood samples, but reticulocyte numbers are higher by manual counting than by ADVIA 2120i counting. Therefore, microscopic confirmation of quantifications is recommended when high numbers of reticulocytes are observed. The anticoagulant of choice is EDTA.
引用
收藏
页码:337 / 341
页数:5
相关论文
共 50 条
  • [41] Diagnostic Accuracy of Siemens ADVIA 2120i's Five Hematological Parameters in the Identification of Pediatric Acute Lymphoblastic Leukemia
    Canovi, Simone
    Campioli, Daniele
    Gilioli, Lisa
    PEDIATRIC AND DEVELOPMENTAL PATHOLOGY, 2015, 18 (01) : 85 - 86
  • [42] ADVIA 2120I WHITE BLOOD CELL COUNT OF SYNOVIAL FLUIDS: IMPROVEMENT BY HYALURONIDASE PRE-TREATMENT METHOD
    Ciriello, Maria Matilde
    Bensi, Thea
    Calcagno, Lara
    Cattana, Elena
    Doglio, Raffaella
    Guaschino, Roberto
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2016, 38 : 27 - 28
  • [43] 西门子ADVIA 2120i全血细胞分析仪故障4例
    梁镜婵
    医疗装备, 2021, 34 (07) : 133 - 134
  • [44] The diagnostic use of ADVIA 2120i Siemens and an "APL criteria" can help to reduce the rate of early death in the APL
    Rocco, Vincenzo
    Castelli, Christian
    Fumi, Maurizio
    Mancini, Francesca
    Pancione, Ylenia
    Prisciandaro, Michele
    Sale, Silvia
    Tanca, Donatella
    Vagnoni, Davide
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2019, 41 (01) : 124 - 132
  • [45] Evaluation of automated nucleated red blood cells counting on Sysmex XE5000 and Siemens ADVIA 2120
    Pipitone, Silvia
    Pavesi, Fernanda
    Testa, Beatrice
    Bardi, Mirco
    Perri, Giuseppina B.
    Gennari, Daniela
    Lippi, Giuseppe
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2012, 50 (10) : 1857 - 1859
  • [46] AUTOMATED CELL COUNTING OF BODY FLUIDS: EVALUATION OF THE ABX PENTRA DX 120 COMPARED TO THE MANUAL MICROSCOPE METHOD, THE SIEMENS ADVIA 2120 AND THE SYSMEX XE 2100
    Danise, P.
    Rovetti, A.
    Ricciardi, C.
    Esposito, C.
    Di Palma, A.
    Avino, D.
    Pirofalo, M. G.
    Talento, B.
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2011, 49 : S718 - S718
  • [47] Canine differential leukocyte counting with the CellaVision DM96Vision, Sysmex XT-2000iV, and Advia 2120 hematology analyzers and a manual method
    Tvedten, Harold W.
    Lilliehook, Inger E.
    VETERINARY CLINICAL PATHOLOGY, 2011, 40 (03) : 324 - 339
  • [48] ADVIA®2120I AND PANDA CLASSIFICATION: ADVANTAGES FOR LEUKEMIA DIAGNOSIS. TRANSLOCATION T(8;21): A CASE REPORT.
    Simoni, Loredana
    Morelli, Daniele
    Cabras, Antonello Domenico
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2012, 34 : 135 - 135
  • [49] Comparison Study of the Rates of Manual Peripheral Blood Smear Review From 3 Automated Hematology Analyzers, Unicel DxH 800, ADVIA 2120i, and XE 2100, Using International Consensus Group Guidelines
    Kim, Sue Jung
    Kim, Yoonjung
    Shin, Saeam
    Song, Jaewoo
    Choi, Jong Rak
    ARCHIVES OF PATHOLOGY & LABORATORY MEDICINE, 2012, 136 (11) : 1408 - 1413
  • [50] COMPARATIVE ANALYSIS OF RESULTS OF 100 HEMOGRAMS IN TWO DIFFERENT TECHNOLOGY ANALYZERS: SYSMEX XN-2000 VS ADVIA 2120I
    Bardan Rebollar, D.
    Gallardo Escribano, C. C.
    Munoz Merida, M. L.
    Prieto Bonilla, M. D. R.
    Rodriguez Jimenez, A. I.
    Pascual Martinez, A. I.
    Butron Vila, R.
    Rodriguez Fernandez, M. J.
    Palomo Bravo, A.
    Ruiz Sanchez, P. L.
    Heiniger Mazo, A. I.
    HAEMATOLOGICA, 2017, 102 : 155 - 155