An amperometric glucose biosensor based on a MnO2/graphene composite modified electrode

被引:43
|
作者
Liu, Yuge [1 ]
Zhang, Xiumei [1 ]
He, Dongning [2 ]
Ma, Feiyue [1 ]
Fua, Qiong [1 ]
Hu, Yun [3 ]
机构
[1] Chinese Acad Trop Agr Sci, South Subtrop Crop Res Inst, Key Lab Trop Fruit Biol, Minist Agr, Zhanjiang 524091, Peoples R China
[2] Univ So Queensland, Ctr Excellence Engineered Fibre Composites, Toowoomba, Qld 4350, Australia
[3] China Tobacco Guizhou Ind Co Ltd, Ctr Technol, Guiyang 550002, Peoples R China
基金
中国国家自然科学基金;
关键词
DIRECT ELECTROCHEMISTRY; OXIDASE; GRAPHENE; NANOPARTICLES; IMMOBILIZATION; HEMOGLOBIN; NAFION; FILM; PERFORMANCE;
D O I
10.1039/c6ra02680j
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
In this paper, a novel composite of graphene/MnO2 (GR/MnO2) was successfully synthesized by a simple one-step hydrothermal method. The as-synthesized MnO2 and the composite were characterized by scanning electron microscopy (SEM), X-ray diffraction (XRD), X-ray photoelectron spectroscopy (XPS) and Fourier transform infrared spectroscopy (FTIR). The results showed that MnO2 was nanorods and the two materials were perfectly composited. The composite was decorated on a glassy carbon electrode (GCE) and used for the entrapment of glucose oxidase (GOD). Electrochemical results showed that the composite modified electrode showed a pair of well-defined redox peaks, and the direct electron transfer between GOD and the electrode surface was accelerated. The sensor fabricated by the composite modified electrode showed an excellent response to the oxidation of glucose with a wide linear range (0.04 to 2 mM), low detection limit (10 mM), and high sensitivity (3.3 mu A mM(-1) cm(-2)). The sensor also exhibited excellent reproducibility, stability and selectivity, and it can be used in the determination of glucose in real samples.
引用
收藏
页码:18654 / 18661
页数:8
相关论文
共 50 条
  • [1] Fabrication and Applications of Amperometric Glucose Biosensor Based on CuO Nanowires/Graphene Modified Electrode
    Tian, Liangliang
    Su, Yongyao
    Liao, Wenli
    ASIAN JOURNAL OF CHEMISTRY, 2013, 25 (18) : 10071 - 10075
  • [2] MnO2/graphene composite electrode for capacitive deionization
    Rashed, Ahmed E.
    Ibrahim, Mohamed I. A.
    Barakat, Assem O.
    Abd El-Moneim, Ahmed
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2014, 248
  • [3] Mesoporous MnO2 as enzyme immobilization host for amperometric glucose biosensor construction
    Yu, Jingjing
    Zhao, Tian
    Zeng, Baizhao
    ELECTROCHEMISTRY COMMUNICATIONS, 2008, 10 (09) : 1318 - 1321
  • [4] Amperometric determination of glucose with an MnO2 and glucose oxidase bulk-modified screen-printed carbon ink biosensor
    Turkusic, E
    Kalcher, K
    Schachl, K
    Komersova, A
    Bartos, M
    Moderegger, H
    Svancara, I
    Vytras, K
    ANALYTICAL LETTERS, 2001, 34 (15) : 2633 - 2647
  • [5] Preparation and performance of MnO2/graphene composite electrode materials
    Yuan, Lei
    Fu, Zhibing
    Chang, Lijuan
    Yang, Xi
    Zhang, Houqiong
    Wang, Chaoyang
    Tang, Yongjian
    Wang, Chaoyang, 1600, Editorial Office of High Power Laser and Particle Beams (26):
  • [6] Preparation of MnO2/graphene composite as electrode material for supercapacitors
    Qian, Yong
    Lu, Shunbao
    Gao, Fenglei
    JOURNAL OF MATERIALS SCIENCE, 2011, 46 (10) : 3517 - 3522
  • [7] Preparation of MnO2/graphene composite as electrode material for supercapacitors
    Yong Qian
    Shunbao Lu
    Fenglei Gao
    Journal of Materials Science, 2011, 46 : 3517 - 3522
  • [8] AMPEROMETRIC UREA BIOSENSOR BASED PLATINUM ELECTRODE MODIFIED WITH A COMPOSITE FILM
    Branzoi, Viorel
    Musina, Ancuta
    Branzoi, Florina
    REVUE ROUMAINE DE CHIMIE, 2011, 56 (09) : 883 - 893
  • [9] Polyaniline/MnO2 nanocomposites based stainless steel electrode modified enzymatic urease biosensor
    Mahajan, Ashish P.
    Kondawar, Subhash B.
    Mahore, Ritu P.
    Meshram, Bhavana H.
    Virutkar, Priyanka D.
    2ND INTERNATIONAL CONFERENCE ON NANOMATERIALS AND TECHNOLOGIES (CNT 2014), 2015, 10 : 699 - 705
  • [10] Amperometric Acetylcholine Biosensor Based on Graphene-PEDOT:PSS Modified Electrode
    Pitikultham, Piyawat
    Sriprachuabwong, Chakrit
    Sritongkham, Pornpimol
    2014 7TH BIOMEDICAL ENGINEERING INTERNATIONAL CONFERENCE (BMEICON), 2014,