SCREEN-FILM VERSUS FULL-FIELD DIGITAL MAMMOGRAPHY: RADIATION DOSE AND IMAGE QUALITY IN A LARGE TEACHING HOSPITAL

被引:2
|
作者
Stantic, Tomislav J. [1 ]
Ciraj-Bjelac, Olivera E. [2 ]
Stojanovic, Saida S. [3 ]
Basta-Nikolic, Marijana D. [3 ]
Arandjic, Danijela D. [2 ]
Stoiljkovic, Dragana S. [4 ]
机构
[1] Ctr Diagnost, Subotica, Serbia
[2] Univ Belgrade, Vinca Inst Nucl Sci, Radiat Protect Dept, Belgrade, Serbia
[3] Clin Ctr Vojvodina, Novi Sad, Serbia
[4] St Elizabeth Krankenhaus, Bad Kissingen, Germany
来源
关键词
mammography; image quality; radiation dose; mean glandular dose; REFERENCE LEVEL; BREAST; PERFORMANCE; RADIOGRAPHY; ACCURACY; PROGRAM;
D O I
10.2298/NTRP1304398S
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
The objective of this paper is to measure the radiation dose and image quality in conventional screen-film mammography and full-field digital mammography in women referred to mammography examination. Participants underwent bilateral, two-view screen-film mammography or full-field digital mammography. The visibility of anatomical regions and overall clinical image quality was rated by experienced radiologists. Total of 387 women and 1548 mammograms were enrolled in the study. Image quality was assessed in terms of image quality score, whereas patient dose assessment was performed in terms of mean glandular dose. Average mean glandular dose for cranio-caudal projection was 1.5 mGy and 2.1 mGy in full-field digital mammography and screen-film mammography, respectively. For medio-lateral oblique projection, corresponding values were 2.3 and 2.1 mGy. Overall image quality criteria scoring was 0.82 and 0.99 for screen-film and digital systems, respectively. The scores were in the range from 0.11 to 1.0 for different anatomical structures. Overall, full-field digital mammography was superior both in terms of image quality and dose over the screen-film mammography. The results have indicated that phantom dose values can assist in setting the optimisation activities in mammography and for comparison between mammography units. To obtain accurate diagnostic information with an acceptable radiation dose to breast, it is necessary to periodically perform patient dose and image quality surveys in all mammography units.
引用
收藏
页码:398 / 405
页数:8
相关论文
共 50 条
  • [31] Studies Comparing Screen-Film Mammography and Full-Field Digital Mammography in Breast Cancer Screening: Updated Review
    Skaane, P.
    ACTA RADIOLOGICA, 2009, 50 (01) : 3 - 14
  • [32] Comparison of reading time between screen-film mammography and soft-copied, full-field digital mammography
    Mitsutomi Ishiyama
    Hiroko Tsunoda-Shimizu
    Mari Kikuchi
    Yukihisa Saida
    Sonoe Hiramatsu
    Breast Cancer, 2009, 16 : 58 - 61
  • [33] Erratum: Screen film vs full-field digital mammography: image quality, detectability and characterization of lesions
    S. Obenauer
    S. Luftner-Nagel
    D. von Heyden
    U. Munzel
    F. Baum
    E. Grabbe
    European Radiology, 2002, 12 : 2388 - 2388
  • [34] Two-modality mammography may confer an advantage over either full-field digital mammography or screen-film mammography
    Glueck, Deborah H.
    Lamb, Molly M.
    Lewin, John M.
    Pisano, Etta D.
    ACADEMIC RADIOLOGY, 2007, 14 (06) : 670 - 676
  • [35] Two-modality mammography may confer an advantage over either full-field digital mammography or screen-film mammography
    Glueck, D. H.
    Lamb, M. M.
    Lewin, J. M.
    Pisano, E. D.
    CANCER EPIDEMIOLOGY BIOMARKERS & PREVENTION, 2007, 16 (02) : 359 - 359
  • [36] Characteristics and prognosis of interval cancers after biennial screen-film or full-field digital screening mammography
    Roy J. P. Weber
    Rob M. G. van Bommel
    Marieke W. Louwman
    Joost Nederend
    Adri C. Voogd
    Frits H. Jansen
    Vivianne C. G. Tjan-Heijnen
    Lucien E. M. Duijm
    Breast Cancer Research and Treatment, 2016, 158 : 471 - 483
  • [37] Characteristics and prognosis of interval cancers after biennial screen-film or full-field digital screening mammography
    Weber, Roy J. P.
    van Bommel, Rob M. G.
    Louwman, Marieke W.
    Nederend, Joost
    Voogd, Adri C.
    Jansen, Frits H.
    Tjan-Heijnen, Vivianne C. G.
    Duijm, Lucien E. M.
    BREAST CANCER RESEARCH AND TREATMENT, 2016, 158 (03) : 471 - 483
  • [38] Comparison of the microcalcifications between full-field digital mammography and screen film mammography
    Tohno, E
    Kujiraoka, Y
    Watanabe, Y
    Itai, Y
    RADIOLOGY, 2001, 221 : 189 - 189
  • [39] Image quality and breast dose of 24 screen-film combinations for mammography
    Dimakopoulou, AD
    Tsalafoutas, IA
    Georgiou, EK
    Yakoumakis, EN
    BRITISH JOURNAL OF RADIOLOGY, 2006, 79 (938): : 123 - 129
  • [40] Optimization of technique factors for a silicon diode array full-field digital mammography system and comparison to screen-film mammography with matched mean glandular dose
    Hendrick, RE
    Berns, EA
    RADIOLOGY, 2001, 221 : 338 - 339