SCREEN-FILM VERSUS FULL-FIELD DIGITAL MAMMOGRAPHY: RADIATION DOSE AND IMAGE QUALITY IN A LARGE TEACHING HOSPITAL

被引:2
|
作者
Stantic, Tomislav J. [1 ]
Ciraj-Bjelac, Olivera E. [2 ]
Stojanovic, Saida S. [3 ]
Basta-Nikolic, Marijana D. [3 ]
Arandjic, Danijela D. [2 ]
Stoiljkovic, Dragana S. [4 ]
机构
[1] Ctr Diagnost, Subotica, Serbia
[2] Univ Belgrade, Vinca Inst Nucl Sci, Radiat Protect Dept, Belgrade, Serbia
[3] Clin Ctr Vojvodina, Novi Sad, Serbia
[4] St Elizabeth Krankenhaus, Bad Kissingen, Germany
来源
关键词
mammography; image quality; radiation dose; mean glandular dose; REFERENCE LEVEL; BREAST; PERFORMANCE; RADIOGRAPHY; ACCURACY; PROGRAM;
D O I
10.2298/NTRP1304398S
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
The objective of this paper is to measure the radiation dose and image quality in conventional screen-film mammography and full-field digital mammography in women referred to mammography examination. Participants underwent bilateral, two-view screen-film mammography or full-field digital mammography. The visibility of anatomical regions and overall clinical image quality was rated by experienced radiologists. Total of 387 women and 1548 mammograms were enrolled in the study. Image quality was assessed in terms of image quality score, whereas patient dose assessment was performed in terms of mean glandular dose. Average mean glandular dose for cranio-caudal projection was 1.5 mGy and 2.1 mGy in full-field digital mammography and screen-film mammography, respectively. For medio-lateral oblique projection, corresponding values were 2.3 and 2.1 mGy. Overall image quality criteria scoring was 0.82 and 0.99 for screen-film and digital systems, respectively. The scores were in the range from 0.11 to 1.0 for different anatomical structures. Overall, full-field digital mammography was superior both in terms of image quality and dose over the screen-film mammography. The results have indicated that phantom dose values can assist in setting the optimisation activities in mammography and for comparison between mammography units. To obtain accurate diagnostic information with an acceptable radiation dose to breast, it is necessary to periodically perform patient dose and image quality surveys in all mammography units.
引用
下载
收藏
页码:398 / 405
页数:8
相关论文
共 50 条
  • [1] Comparison of full-field digital and screen-film mammography dose
    Schueler, B.
    MEDICAL PHYSICS, 2006, 33 (06) : 2005 - 2005
  • [2] Screen-film and digital mammography - Image quality and radiation dose considerations
    Haus, AG
    Yaffe, MJ
    RADIOLOGIC CLINICS OF NORTH AMERICA, 2000, 38 (04) : 871 - +
  • [3] Breast mass classification on full-field digital mammography and screen-film mammography
    Shi, Jiazheng
    Sahiner, Berkman
    Chan, Heang-Ping
    Hadjiiski, Lubomir M.
    Ge, Jun
    Wei, Jun
    DIGITAL MAMMOGRAPHY, PROCEEDINGS, 2008, 5116 : 371 - 377
  • [4] Full-field digital mammography in comparison to screen-film mammography - A phantom study
    Obenauer, S
    Schorn, C
    Funke, M
    Fischer, U
    Grabbe, E
    Hermann, KP
    IWDM 2000: 5TH INTERNATIONAL WORKSHOP ON DIGITAL MAMMOGRAPHY, 2001, : 499 - 503
  • [5] Clinical performance comparison of full-field digital mammography to screen-film mammography
    Berns, EA
    Hendrick, RE
    RADIOLOGY, 2001, 221 : 283 - 283
  • [6] Screen-film versus full-field digital mammography:: Comparison in breast cancer screening
    Sigfússon, BF
    ACTA RADIOLOGICA, 2005, 46 (07) : 663 - 663
  • [7] Full-field digital versus screen-film mammography: Comparative accuracy in concurrent screening cohorts
    Del Turco, Marco Rosselli
    Mantellini, Paola
    Ciatto, Stefano
    Bonardi, Rita
    Martinelli, Francesca
    Lazzari, Barbara
    Houssami, Nehmat
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2007, 189 (04) : 860 - 866
  • [8] Performance comparison of full-field digital mammography to screen-film mammography in clinical practice
    Berns, EA
    Hendrick, RE
    Cutter, GR
    MEDICAL PHYSICS, 2002, 29 (05) : 830 - 834
  • [9] Computer-Aided Detection with Full-Field Digital Mammography and Screen-Film Mammography
    Melton, A. R.
    Worrell, S. W.
    Knapp, J.
    Hoffmeister, J. W.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2007, 188 (05)
  • [10] Clinical comparison of full-field digital mammography and screen-film mammography of breast cancer
    Lewin, JM
    D'Orsi, CJ
    Hendrick, RE
    Moss, LJ
    Isaacs, PK
    Karellas, A
    Cutter, GR
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2002, 179 (03) : 671 - 677