Performance evaluation of a whole-body prototype PET scanner with four-layer DOI detectors

被引:37
|
作者
Akamatsu, Go [1 ]
Tashima, Hideaki [1 ]
Iwao, Yuma [1 ]
Wakizaka, Hidekatsu [1 ]
Maeda, Takamasa [1 ]
Mohammadi, Akram [1 ]
Takyu, Sodai [1 ]
Nitta, Munetaka [1 ]
Nishikido, Fumihiko [1 ]
Rutherford, Harley [2 ,3 ]
Chacon, Andrew [2 ,3 ]
Safavi-Naeini, Mitra [3 ]
Yoshida, Eiji [1 ]
Yamaya, Taiga [1 ]
机构
[1] Natl Inst Quantum & Radiol Sci & Technol QST, NIRS, Chiba, Japan
[2] Univ Wollongong, Ctr Med Radiat Phys, Wollongong, NSW, Australia
[3] ANSTO, Sydney, NSW, Australia
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2019年 / 64卷 / 09期
关键词
PET; DOI; whole-body; NEMA NU 2; OF-FLIGHT PET/CT; AXIAL-FIELD; RESOLUTION; CRYSTALS; SYSTEM; DESIGN; IMPACT; VIEW; PMT;
D O I
10.1088/1361-6560/ab18b2
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Parallax error caused by the detector crystal thickness degrades spatial resolution at the peripheral regions of the field-of-view (FOV) of a scanner. To resolve this issue, depth-of-interaction (DOI) measurement is a promising solution to improve the spatial resolution and its uniformity over the entire FOV. Even though DOI detectors have been used in dedicated systems with a small ring diameter such as for the human brain, breast and small animals, the use of DOI detectors for a large bore whole-body PET system has not been demonstrated yet. We have developed a four-layered DOI detector, and its potential for a brain dedicated system has been proven in our previous development. In the present work, we investigated the use of the four-layer DOI detector for a large bore PET system by developing the world's first whole-body prototype. We evaluated its performance characteristics in accordance with the NEMA NU 2 standard. Furthermore, the impact of incorporating DOI information was evaluated with the NEMA NU 4 image quality phantom. Point source images were reconstructed with a filtered back projection (FBP), and an average spatial resolution of 5.2 +/- 0.7 mm was obtained. For the FBP image, the four-layer DOI information improved the radial spatial resolution by 48% at the 20 cm offset position. The peak noise-equivalent count rate (NECR) was 22.9 kcps at 7.4 kBq ml(-1) and the scatter fraction was 44%. The system sensitivity was 5.9 kcps MBq(-1). For the NEMA NU 2 image quality phantom, the 10 mm sphere was clearly visualized without any artifacts. For the NEMA NU 4 image quality phantom, we measured the phantom at 0, 10 and 20 cm offset positions. As a result, we found the image with four-layer DOI could visualize the 2 mm-diameter hot cylinder although it could not be recognized on the image without DOI. The average improvements in the recovery coefficients for the five hot rods (1-5 mm) were 0.3%, 4.4% and 26.3% at the 0, 10 and 20 cm offset positions, respectively (except for the 1 mm-diameter rod at the 20 cm offset position). Although several practical issues (such as adding end-shields) remain to be addressed before the scanner is ready for clinical use, we showed that the four-layer DOI technology provided higher and more uniform spatial resolution over the FOV and improved contrast for small uptake regions located at the peripheral FOV, which could improve detectability of small and distal lesions such as nodal metastases, especially in obese patients.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] A validated Geant4 model of a whole-body PET scanner with four-layer DOI detectors
    Ahmed, Abdella M.
    Chacon, Andrew
    Rutherford, Harley
    Akamatsu, Go
    Mohammadi, Akram
    Nishikido, Fumihiko
    Tashima, Hideaki
    Yoshida, Eiji
    Yamaya, Taiga
    Franklin, Daniel R.
    Rosenfeld, Anatoly
    Guatelli, Susanna
    Safavi-Naeini, Mitra
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (23):
  • [2] Performance evaluation of a high-resolution brain PET scanner using four-layer MPPC DOI detectors
    Watanabe, Mitsuo
    Saito, Akinori
    Isobe, Takashi
    Ote, Kibo
    Yamada, Ryoko
    Moriya, Takahiro
    Omura, Tomohide
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (17): : 7148 - 7166
  • [3] Development of a C-Shaped Breast PET Scanner Equipped with Four-Layer DOI Detectors
    Kitamura, Keishi
    Ohi, Junichi
    Tonami, Hiromichi
    Yamada, Yoshihiro
    Furumiya, Tetsuo
    Furuta, Masafumi
    Satoh, Masanobu
    Tsuda, Tomoaki
    Nakazawa, Masayuki
    Hashizume, Nobuya
    Yamakawa, Yoshiyuki
    Kawashima, Ayako
    Kumazawa, Yoshihiko
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 5421 - 5424
  • [4] Performance evaluation of a four-layer LSO detector for a small animal DOI PET scanner: jPET-RD
    Tsuda, T
    Murayama, H
    Kitamura, K
    Inadama, N
    Yamaya, T
    Yoshida, E
    Nishikido, F
    Hamamoto, M
    Kawai, H
    Ono, Y
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 3321 - 3324
  • [5] Spatial resolution evaluation with a pair of two four-layer DOI detectors for small animal PET scanner: jPET-RD
    Nishikido, Fumihiko
    Tsuda, Tomoaki
    Yoshida, Eiji
    Inadama, Naoko
    Shibuya, Kengo
    Yamaya, Taiga
    Kitamura, Keishi
    Takahashi, Kei
    Ohmura, Atsushi
    Murayama, Hideo
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2008, 584 (01): : 212 - 218
  • [6] Performance evaluation of a subset of a four-layer LSO detector for a small animal DOI PET scanner: jPET-RD
    Tsuda, T
    Murayama, H
    Kitamura, K
    Inadama, N
    Yamaya, T
    Yoshida, E
    Nishikido, F
    Hamamoto, M
    Kawai, H
    Ono, Y
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (01) : 35 - 39
  • [7] A prototype PET scanner with DOI-encoding detectors
    Yang, Yongfeng
    Wu, Yibao
    Qi, Jinyi
    James, Sara St.
    Du, Huini
    Dokhale, Purushottam A.
    Shah, Kanai S.
    Farrell, Richard
    Cherry, Simon R.
    JOURNAL OF NUCLEAR MEDICINE, 2008, 49 (07) : 1132 - 1140
  • [8] Axial scalable add-on PET/MRI prototype based on four-layer DOI detectors integrated with a RF coil
    Nishikido, Fumihiko
    Obata, Takayuki
    Suga, Mikio
    Iwao, Yuma
    Tashima, Hideaki
    Yoshida, Eiji
    Akram, Md Shahadat Hossain
    Yamaya, Taiga
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2022, 1040
  • [9] Performance evaluation of a whole-body PET scanner using the NEMA protocol
    Brix, G
    Zaers, J
    Adam, LE
    Bellemann, ME
    Ostertag, H
    Trojan, H
    Haberkorn, U
    Doll, J
    Oberdorfer, F
    Lorenz, WJ
    JOURNAL OF NUCLEAR MEDICINE, 1997, 38 (10) : 1614 - 1623
  • [10] Axial scalable add-on PET/MRI prototype based on four-layer DOI detectors integrated with a RF coil
    Nishikido, Fumihiko
    Obata, Takayuki
    Suga, Mikio
    Iwao, Yuma
    Tashima, Hideaki
    Yoshida, Eiji
    Akram, Md Shahadat Hossain
    Yamaya, Taiga
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2022, 1040