Comparison of blood pressure measurements of anesthetized dogs obtained noninvasively with a cylindrical blood pressure cuff and an anatomically modified conical blood pressure cuff

被引:0
|
作者
Ramos, Sara J. [1 ]
da Cunha, Anderson F. [1 ]
Domingues, Michelle [1 ]
Shelby, Amanda M. [1 ]
Stout, Rhett W. [2 ]
Acierno, Mark J. [1 ]
机构
[1] Louisiana State Univ, Sch Vet Med, Dept Vet Clin Sci, Baton Rouge, LA 70803 USA
[2] Louisiana State Univ, Sch Vet Med, Div Lab Anim Med, Baton Rouge, LA 70803 USA
关键词
SYSTEMIC HYPERTENSION; AGREEMENT; MONITOR; GUIDELINES; DIAGNOSIS; CATS;
D O I
暂无
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
OBJECTIVE To compare blood pressure measured noninvasively with an oscillometric device that involved use of a novel conical cuff and a traditional cylindrical blood pressure cuff. ANIMALS 17 adult hound-type dogs. PROCEDURES Dogs were anesthetized, and a 20-gauge, 1.5-inch catheter was inserted in the median sacral artery. The catheter was attached to a pressure transducer via fluid-filled noncompliant tubing, and direct blood pressure was recorded with a multifunction monitor. A specially fabricated conical cuff was placed on the antebrachium. Four sets of direct and indirect blood pressure measurements were simultaneously collected every 2 minutes. Four sets of measurements were then obtained by use of a cylindrical cuff. RESULTS The cylindrical cuff met American College of Veterinary Internal Medicine consensus guidelines for validation of indirect blood pressure measurements for mean arterial blood pressure (MAP), systolic arterial blood pressure (SAP), and diastolic arterial blood pressure (DAP). The conical cuff met the consensus guidelines for difference of paired measurements, SD, and percentages of measurements within 10 and 20 mm Hg of the value for the reference method, but it failed a correlation analysis. In addition, although bias for the conical cuff was less than that for the cylindrical cuff for SAP, MAP, and DAP measurements, the limits of agreement for the conical cuff were wider than those for the cylindrical cuff for SAP and MAP measurements. CONCLUSIONS AND CLINICAL RELEVANCE On the basis of results of this study, use of a conical cuff for oscillometric blood pressure measurement cannot be recommended.
引用
收藏
页码:59 / 64
页数:6
相关论文
共 50 条
  • [41] EFFECT OF CUFF WIDTH ON ACCURACY OF INDIRECT MEASUREMENT OF BLOOD PRESSURE IN DOGS
    VALTONEN, MH
    ERIKSSON, LM
    RESEARCH IN VETERINARY SCIENCE, 1970, 11 (04) : 358 - &
  • [42] Blood pressure measurement using finger cuff
    Lee, J. Y.
    Choi, E. Y.
    Jeong, H. J.
    Kim, K. H.
    Park, J. C.
    2005 27TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2005, : 3575 - 3577
  • [43] Pitfalls of Cuff Size and Blood Pressure Measurement
    Gallagher, Benjamin D.
    Bryant, Kelsey B.
    Cluett, Jennifer L.
    JAMA INTERNAL MEDICINE, 2024, 184 (04) : 448 - 449
  • [44] SIZE OF BLOOD PRESSURE CUFF FOR JAPANESE ADULTS
    NUKADA, A
    ISHIZAKI, T
    NAGATA, H
    TOHOKU JOURNAL OF EXPERIMENTAL MEDICINE, 1963, 79 (03): : 293 - 305
  • [45] THE SIGNIFICANCE OF BLOOD PRESSURE RECORDED BY THE CUFF METHOD
    DEXTER, L
    JOURNALS OF GERONTOLOGY, 1951, 6 (04): : 79 - 79
  • [46] Off the cuff : rebooting blood pressure treatment
    Steinhubl, Steven R.
    Muse, Evan D.
    Barrett, Paddy M.
    Topol, Eric J.
    LANCET, 2016, 388 (10046): : 749 - 749
  • [47] Cuff width and accuracy of measurement of blood pressure
    Alpert, BS
    BLOOD PRESSURE MONITORING, 2000, 5 (03) : 151 - 152
  • [48] The influence of cuff size on blood pressure measurement
    C Bakx
    G Oerlemans
    H van den Hoogen
    C van Weel
    T Thien
    Journal of Human Hypertension, 1997, 11 : 439 - 445
  • [49] The PAPERCUFF, a new disposable blood pressure cuff
    Alpert, BS
    Cohen, ML
    AMERICAN JOURNAL OF CARDIOLOGY, 1996, 77 (07): : 531 - &
  • [50] SELECTION OF BLOOD-PRESSURE CUFF DESIGN
    KING, GE
    LANCET, 1982, 2 (8296): : 492 - 493