Effect of tissue composition on dose distribution in brachytherapy with various photon emitting sources

被引:16
|
作者
Ghorbani, Mahdi [1 ]
Salahshour, Fateme [2 ]
Haghparast, Abbas [2 ]
Moghaddas, Toktam Ahmadi [2 ]
Knaup, Courtney [3 ]
机构
[1] Mashhad Univ Med Sci, Fac Med, Med Phys Res Ctr, Mashhad, Iran
[2] Kermanshah Univ Med Sci, Fac Med, Med Phys & Med Engn Dept, Kermanshah, Iran
[3] Comprehens Canc Ctr Nevada, Las Vegas, NV USA
关键词
brachytherapy; ICRU; photon emitting sources; relative dose; soft tissue; MONTE-CARLO CHARACTERIZATION; IR-192; BRACHYTHERAPY; DOSIMETRY PARAMETERS; PD-103; YB-169; CO-60; WATER; UPDATE; I-125; LUNG;
D O I
10.5114/jcb.2014.42024
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: The aim of this study is to compare the dose in various soft tissues in brachytherapy with photon emitting sources. Material and methods: Pd-103, I-125, Yb-169, Ir-192 brachytherapy sources were simulated with MCNPX Monte Carlo code, and their dose rate constant and radial dose function were compared with the published data. A spherical phantom with 50 cm radius was simulated and the dose at various radial distances in adipose tissue, breast tissue, 4-component soft tissue, brain (grey/white matter), muscle (skeletal), lung tissue, blood (whole), 9-component soft tissue, and water were calculated. The absolute dose and relative dose difference with respect to 9-component soft tissue was obtained for various materials, sources, and distances. Results: There was good agreement between the dosimetric parameters of the sources and the published data. Adipose tissue, breast tissue, 4-component soft tissue, and water showed the greatest difference in dose relative to the dose to the 9-component soft tissue. The other soft tissues showed lower dose differences. The dose difference was also higher for Pd-103 source than for I-125, Yb-169, and Ir-192 sources. Furthermore, greater distances from the source had higher relative dose differences and the effect can be justified due to the change in photon spectrum (softening or hardening) as photons traverse the phantom material. Conclusions: The ignorance of soft tissue characteristics (density, composition, etc.) by treatment planning systems incorporates a significant error in dose delivery to the patient in brachytherapy with photon sources. The error depends on the type of soft tissue, brachytherapy source, as well as the distance from the source.
引用
收藏
页码:54 / 67
页数:14
相关论文
共 50 条
  • [1] Presentation of a matrix-based method to calculate dose distribution in brachytherapy with photon-emitting sources
    Ghorbani, Mahdi
    Mozaffari, Azam
    Akbari-Lalimi, Hossein
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2019, 15 (06) : 1245 - 1253
  • [2] Evaluation of brachytherapy lung implant dose distributions from photon-emitting sources due to tissue heterogeneities
    Yang, Yun
    Rivard, Mark J.
    MEDICAL PHYSICS, 2011, 38 (11) : 5857 - 5862
  • [3] Tissue composition effect on dose distribution in neutron brachytherapy/neutron capture therapy
    Khosroabadi, Mohsen
    Farhood, Bagher
    Ghorbani, Mahdi
    Hamzian, Nima
    Moghaddam, Homa Rezaei
    Davenport, David
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2016, 21 (01) : 8 - 16
  • [4] Photon spectrometry for the determination of the dose-rate constant of low-energy photon-emitting brachytherapy sources
    Chen, Zhe Jay
    Nath, Ravinder
    MEDICAL PHYSICS, 2007, 34 (04) : 1412 - 1430
  • [5] A Generalized Dose-Rate Constant Formalism for Brachytherapy Sources Emitting Anisotropic Photon Energy Spectrum
    Chen, Z.
    Nath, R.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [6] Tissue composition effect on dose distribution in radiotherapy with a 6 MV photon beam of a medical linac
    Ghorbani, Mahdi
    Noghreiyan, Atefeh Vejdani
    Tabatabaei, Zahra Sadat
    Pakravan, Delaram
    Davenport, David
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2019, 15 (01) : 237 - 244
  • [7] A Monte Carlo evaluation of dose enhancement by cisplatin and titanocene dichloride chemotherapy drugs in brachytherapy with photon emitting sources
    Abadi, Akram Yahya
    Ghorbani, Mahdi
    Mowlavi, Ali Asghar
    Knaup, Courtney
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2014, 37 (02) : 327 - 336
  • [8] A Monte Carlo evaluation of dose enhancement by cisplatin and titanocene dichloride chemotherapy drugs in brachytherapy with photon emitting sources
    Akram Yahya Abadi
    Mahdi Ghorbani
    Ali Asghar Mowlavi
    Courtney Knaup
    Australasian Physical & Engineering Sciences in Medicine, 2014, 37 : 327 - 336
  • [9] Evaluation of Tissue Composition Effect On Dose Distribution in Radiotherapy with 6 MV Photon Beam of a Medical Linac
    Ghorbani, M.
    Tabatabaei, Z.
    Noghreiyan, A. Vejdani
    Meigooni, A. Soleimani
    MEDICAL PHYSICS, 2015, 42 (06) : 3428 - 3428
  • [10] RADIATION THERAPY PHOTON BEAMS DOSE CONFORMATION ACCORDING TO DOSE DISTRIBUTION AROUND INTRACAVITARY-APPLIED BRACHYTHERAPY SOURCES
    Jurkovic, Slaven
    Zauhar, Gordana
    Faj, Dario
    Radojcic, Deni Smilovic
    Svabic, Manda
    MEDICAL DOSIMETRY, 2010, 35 (01) : 49 - 52