A Decade of Outcomes and Predictors of Sac Enlargement after Endovascular Abdominal Aortic Aneurysm Repair Using Zenith Endografts in a Japanese Population

被引:21
|
作者
Iwakoshi, Shinichi [1 ]
Ichihashi, Shigeo [1 ]
Higashiura, Wataru [1 ]
Itoh, Hirofumi [1 ]
Sakaguchi, Shoji [1 ]
Tabayashi, Nobuoki [2 ]
Uchida, Hideo [1 ]
Kichikawa, Kimihiko [1 ]
机构
[1] Nara Med Univ, Dept Radiol, Kashihara, Nara 6348521, Japan
[2] Nara Med Univ, Dept Thorac & Cardiovasc Surg, Kashihara, Nara 6348521, Japan
关键词
NECK ANATOMY; STENT-GRAFT; EXPERIENCE; EUROSTAR; DIAMETER; COMPLICATIONS; HOSTILE;
D O I
10.1016/j.jvir.2014.01.017
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To present 10-year outcomes and risk factors for sac enlargement after endovascular aneurysm repair (EVAR) using the Zenith, AAA Endovascular Graft (Cook, Inc, Bloomington, Indiana) in a Japanese population. Material and Methods: During the period 1999-2011, 127 patients underwent elective EVAR using Zenith endografts at a single institution. A retrospective investigation looked at initial rates of technical success and complications, 10-year rate of freedom from all-cause and aneurysm-related mortality, freedom from secondary intervention and sac,enlargement, and risk factors for second intervention and sac enlargement. Results: The median age of the patients was 78 years, and the median follow-tip time was 43 months. The initial technical success rate was 98.4% (123 of 127 patients). Major adverse events occurred in 7 of 127 (5.5%) patients. Rates of freedom from all-cause and aneurysm-related mortality at 1, 3, 5, mid 10 years were 95%, 87%, 77%, and 39% (all-cause mortality) and 100%, 100%; 99%, and 93% (aneurysm-related mortality). Rates of freedom from secondary intervention at 1, 3 5, and 10 years were 97%, 91%, 88%, and 70%. Rates of primary freedom from sac enlargement at 1, 3, 5, and 10 years were 99%, 87%, 75%, and 67%. Multivariate analysis revealed aneurysm sac diameter as an independent risk factor for a secondary intervention. Preoperative sac diameter combined with an angulated short (AS) proximal neck was a risk factor for sac enlargement. Conclusions: The 10-year results of EVAR using Zenith endografts in a Japanese population were comparable to results from Western countries. Larger aneurysms and AS neck were predictors of sac enlargement after EVAR.
引用
收藏
页码:694 / 701
页数:8
相关论文
共 50 条
  • [41] Postdischarge outcomes after endovascular abdominal aortic aneurysm repair DISCUSSION
    Dosluoglu, Hasan
    Gupta, Prateek K.
    Mell, Matthew
    Sidawy, Anton
    Riles, Thomas
    JOURNAL OF VASCULAR SURGERY, 2014, 59 (04) : 907 - 908
  • [42] Aneurysm sac pressure measurements after endovascular repair of abdominal aortic aneurysms - Discussion
    Hans, SS
    Baum, RA
    Edwards, JD
    Berguer
    Green
    Ohki, T
    Mattos, MA
    White, GH
    Parodi, JC
    JOURNAL OF VASCULAR SURGERY, 2001, 33 (01) : 40 - 41
  • [43] Long-term results after endovascular abdominal aortic aneurysm repair using the Cook Zenith endograft
    Mertens, Johan
    Houthoofd, Sabrina
    Daenens, Kim
    Fourneau, Inge
    Maleux, Geert
    Lerut, Philip
    Nevelsteen, Andre
    JOURNAL OF VASCULAR SURGERY, 2011, 54 (01) : 48 - 57
  • [44] Regarding "Aneurysm sac pressure measurements after endovascular repair of abdominal aortic aneurysms"
    Hinchliffe, R
    Hopkinson, BR
    JOURNAL OF VASCULAR SURGERY, 2002, 35 (03) : 625 - 626
  • [45] Regarding "Aneurysm sac pressure measurements after endovascular repair of abdominal aortic aneurysms"
    Golzarian, J
    JOURNAL OF VASCULAR SURGERY, 2001, 33 (06) : 1317 - 1318
  • [46] A morphovolumetric analysis of aneurysm sac evolution after elective endovascular abdominal aortic repair
    Franchin, Marco
    Serafini, Maddalena
    Tadiello, Marco
    Fontana, Federico
    Rivolta, Nicola
    Venturini, Massimo
    Curti, Marco
    Bush, Ruth L.
    Dorigo, Walter
    Piacentino, Filippo
    Tozzi, Matteo
    Piffaretti, Gabriele
    JOURNAL OF VASCULAR SURGERY, 2021, 74 (04) : 1222 - +
  • [47] Abdominal aortic aneurysm sac shrinkage after endovascular aneurysm repair: Correlation with chronic sac pressure measurement - Discussion
    Lipsitz, EC
    Ellozy, SH
    Dardik, A
    Hingorani, A
    JOURNAL OF VASCULAR SURGERY, 2006, 43 (01) : 6 - 7
  • [48] Aortic neck enlargement after endovascular aneurysm repair using balloon-expandable versus self-expanding endografts
    Savlovskis, Janis
    Krievins, Dainis
    de Vries, Jean-Paul P. M.
    Holden, Andrew
    Kisis, Kaspars
    Gedins, Marcis
    Ezite, Natalija
    Zarins, Christopher K.
    JOURNAL OF VASCULAR SURGERY, 2015, 62 (03) : 541 - 549
  • [49] Symptomatic sac enlargement and rupture due to seroma after open abdominal aortic aneurysm repair with polytetrafluoroethylene graft: Implications for endovascular repair and endotension
    Thoo, CHC
    Bourke, BM
    May, J
    JOURNAL OF VASCULAR SURGERY, 2004, 40 (06) : 1089 - 1094
  • [50] Utility of Noninvasive Aneurysm Sac Pressure Measurement During and After Endovascular Abdominal Aortic Aneurysm Repair
    Hall, Michael J.