Minimizing Pedicle Screw Pullout Risks A Detailed Biomechanical Analysis of Screw Design and Placement

被引:80
|
作者
Bianco, Rohan-Jean [1 ,2 ,3 ]
Arnoux, Pierre-Jean [3 ]
Wagnac, Eric [1 ,3 ]
Mac-Thiong, Jean-Marc [2 ,4 ,5 ]
Aubin, Carl-Eric [1 ,2 ,4 ]
机构
[1] Polytech Montreal, Dept Mech Engn, Montreal, PQ, Canada
[2] Sainte Justine Univ Hosp Ctr, Montreal, PQ, Canada
[3] Univ Aix Marseille 1, Lab Biomcca Appl, UMRT24 IFSTTAR, Marseille, France
[4] Univ Montreal, Dept Surg, Montreal, PQ, Canada
[5] Hop Sacre Coeur, Dept Surg, Montreal, PQ, Canada
来源
CLINICAL SPINE SURGERY | 2017年 / 30卷 / 03期
基金
加拿大自然科学与工程研究理事会;
关键词
finite element analysis; pedicle screw; pullout test; spinal instrumentation; FINITE-ELEMENT MODEL; INSERTIONAL TORQUE; FIXATION; BONE; INSTRUMENTATION; CALIBRATION; SCOLIOSIS; STRENGTH; FAILURE; TESTS;
D O I
10.1097/BSD.0000000000000151
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design: Detailed biomechanical analysis of the anchorage performance provided by different pedicle screw designs and placement strategies under pullout loading. Objective: To biomechanically characterize the specific effects of surgeon-specific pedicle screw design parameters on anchorage performance using a finite element model. Summary of Background Data: Pedicle screw fixation is commonly used in the treatment of spinal pathologies. However, there is little consensus on the selection of an optimal screw type, size, and insertion trajectory depending on vertebra dimension and shape. Methods: Different screw diameters and lengths, threads, and insertion trajectories were computationally tested using a design of experiment approach. A detailed finite element model of an L3 vertebra was created including elastoplastic bone properties and contact interactions with the screws. Loads and boundary conditions were applied to the screws to simulate axial pullout tests. Force-displacement responses and internal stresses were analyzed to determine the specific effects of each parameter. Results: The design of experiment analysis revealed significant effects (P<0.01) for all tested principal parameters along with the interactions between diameter and trajectory. Screw diameter had the greatest impact on anchorage performance. The best insertion trajectory to resist pullout involved placing the screw threads closer to the pedicle walls using the straightforward insertion technique, which showed the importance of the cortical layer grip. The simulated cylindrical single-lead thread screws presented better biomechanical anchorage than the conical dual-lead thread screws in axial loading conditions. Conclusions: The model made it possible to quantitatively measure the effects of both screw design characteristics and surgical choices, enabling to recommend strategies to improve single pedicle screw performance under axial loading.
引用
收藏
页码:E226 / E232
页数:7
相关论文
共 50 条
  • [21] Implications of navigation in thoracolumbar pedicle screw placement on screw accuracy and screw diameter/pedicle width ratio
    Mandelka, Eric
    Gierse, Jula
    Zimmermann, Felix
    Gruetzner, Paul A.
    Franke, Jochen
    Vetter, Sven Y.
    BRAIN AND SPINE, 2023, 3
  • [22] Assessment of pedicle screw pullout strength based on various screw designs and bone densities-an ex vivo biomechanical study
    Kim, Young-Yul
    Choi, Woo-Sung
    Rhyu, Kee-Won
    SPINE JOURNAL, 2012, 12 (02): : 164 - 168
  • [23] Biomechanical Investigation of a Novel Revision Device in an Osteoporotic Model Pullout Strength of Pedicle Screw Anchor Versus Larger Screw Diameter
    Manon, Jacinto
    Hussain, Mir M.
    Harris, Jonathan
    Moldavsky, Mark
    La Marca, Frank
    Bucklen, Brandon S.
    CLINICAL SPINE SURGERY, 2017, 30 (06): : 265 - 271
  • [24] Pedicle screw placement accuracy - A meta-analysis
    Kosmopoulos, Victor
    Schizas, Constantin
    SPINE, 2007, 32 (03) : E111 - E120
  • [25] Improving the pullout strength of pedicle screws by screw coupling
    Suzuki, T
    Abe, E
    Okuyama, K
    Sato, K
    JOURNAL OF SPINAL DISORDERS, 2001, 14 (05): : 399 - 403
  • [26] PEDICLE SCREW PULLOUT STRENGTH - CORRELATION WITH INSERTIONAL TORQUE
    ZDEBLICK, TA
    KUNZ, DN
    COOKE, ME
    MCCABE, R
    SPINE, 1993, 18 (12) : 1673 - 1676
  • [27] PEDICLE SCREW PLACEMENT WITH INTRAOSSEOUS ENDOSCOPY
    STAUBER, MH
    BASSETT, GS
    SPINE, 1994, 19 (01) : 57 - 61
  • [28] Impact of lumbar pedicle screw positioning on screw stability - A biomechanical investigation
    Grevenstein, David
    Scheyerer, Max J.
    Meyer, Carolin
    Borggrefe, Jan
    Hackl, Michael
    Oikonomidis, Stavros
    Eysel, Peer
    Prescher, Andreas
    Wegmann, Kilian
    CLINICAL BIOMECHANICS, 2020, 74 (74) : 66 - 72
  • [29] Axial pullout strength comparison of different screw designs: fenestrated screw, dual outer diameter screw and standard pedicle screw
    Christodoulou, Evangelos
    Chinthakunta, Suresh
    Reddy, Divya
    Khalil, Saif
    Apostolou, Thomas
    Drees, Philipp
    Kafchitsas, Konstantinos
    SCOLIOSIS AND SPINAL DISORDERS, 2015, 10
  • [30] Pullout Strength after Expandable Polymethylmethacrylate Transpedicular Screw Augmentation for Pedicle Screw Loosening
    Kang, Suk-Hyung
    Cho, Yong Jun
    Kim, Young-Baeg
    Park, Seung Won
    JOURNAL OF KOREAN NEUROSURGICAL SOCIETY, 2015, 57 (04) : 229 - 234