Evaluation of A-SPECT: A desktop pinhole SPECT system for small animal Imaging

被引:0
|
作者
McElroy, DP [1 ]
MacDonald, LR [1 ]
Beekman, FJ [1 ]
Wang, YC [1 ]
Patt, BE [1 ]
Iwanczyk, JS [1 ]
Tsui, BMW [1 ]
Hoffman, EJ [1 ]
机构
[1] Univ Calif Los Angeles, Sch Med, Div Nucl Med & Biophys, Los Angeles, CA 90095 USA
关键词
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Pinhole collimation of gamma rays to image distributions of radiolabeled tracers is considered promising for use in small animal imaging. The recent availability of transgenic mice, coupled with the development of I-125 and Tc-99m labeled tracers, has allowed the study of a range of human disease models while creating demand for ultra-high resolution imaging devices. We have developed a compact gamma camera that, in combination with pinhole collimation, allows for accessible, ultra-high resolution in vivo SPECT imaging of small animals. The system is based on a pixilated array of NaI(Tl) crystals coupled to an array of position sensitive photomultiplier tubes (PSPMTs). Interchangeable tungsten pinholes with diameters ranging from 0.5 mm to 3 mm are available, allowing the camera to be optimized for a variety of imaging situations. We use a 3D MLEM algorithm to reconstruct projection images, allowing for 3D positioning of isotope distributions. Our evaluation indicates that high quality, sub-millimeter spatial resolution images can be achieved in living mice. Reconstructed axial spatial resolution was measured to be 0.53 mm and 0.74 mm FWHM for rotation radii of 1 cm and 2 cm, respectively, using the 0.5 mm pinhole. In this configuration, sensitivity is comparable to that of a standard high resolution parallel hole collimator. SPECT images of hot and cold rod phantoms and a highly structured monkey brain phantom illustrate that high quality images can be obtained with the system. Images of living mice demonstrate the ability of the system to obtain high resolution images in vivo.
引用
下载
收藏
页码:1835 / 1839
页数:5
相关论文
共 50 条
  • [21] Evaluation of high-resolution pinhole SPECT using a small rotating animal
    Habraken, JBA
    de Bruin, K
    Shehata, M
    Booij, J
    Bennink, R
    Smit, BLFV
    Sokole, EB
    JOURNAL OF NUCLEAR MEDICINE, 2001, 42 (12) : 1863 - 1869
  • [22] Design of rotating multi-pinhole SPECT system for high resolution small animal imaging.
    Ivanovic, M
    Weber, DA
    Loncaric, S
    JOURNAL OF NUCLEAR MEDICINE, 1998, 39 (05) : 174P - 174P
  • [23] A prototype low-energy, multi-pinhole SPECT system for small-animal imaging
    Peterson, TE
    Wilson, DW
    Barrett, HH
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 2999 - 3002
  • [24] Multi-pinhole spect: Recent results of an animal imaging system.
    Schramm, NU
    Behe, M
    Schurrat, T
    Engeland, U
    Ebel, G
    Behr, TM
    JOURNAL OF NUCLEAR MEDICINE, 2003, 44 (05) : 9P - 10P
  • [25] SPECT-CT system for small animal imaging
    Weisenberger, AG
    Wojcik, R
    Bradley, EL
    Brewer, P
    Majewski, S
    Qian, J
    Ranck, A
    Saha, MS
    Smith, K
    Smith, MF
    Welsh, RE
    2001 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORDS, VOLS 1-4, 2002, : 1840 - 1844
  • [26] SPECT-CT system for small animal imaging
    Weisenberger, AG
    Wojcik, R
    Bradley, EL
    Brewer, P
    Majewski, S
    Qian, J
    Ranck, A
    Saha, MS
    Smith, K
    Smith, MF
    Welsh, RE
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2003, 50 (01) : 74 - 79
  • [27] A compact SPECT/CT system for small animal imaging
    Hong, Key Jo
    Choi, Yong
    Lee, Sang Chul
    Lee, Soo Yeol
    Song, Tae Yong
    Min, Byung Jun
    Jung, Jin Ho
    Choe, Yearn Seong
    Lee, Kyung-Han
    Kim, Byung-Tae
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (05) : 2601 - 2604
  • [28] Performance evaluation of a pinhole SPECT system for myocardial perfusion imaging of mice
    Wu, MC
    Hasegawa, BH
    Dae, MW
    MEDICAL PHYSICS, 2002, 29 (12) : 2830 - 2839
  • [29] Helical pinhole SPECT for small-animal imaging: A method for addressing sampling completeness
    Metzler, SD
    Greer, KL
    Jaszczak, RJ
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2003, 50 (05) : 1575 - 1583
  • [30] CT-SPECT for small animal imaging with A-SPECT.
    Iwata, K
    MacDonald, LR
    Hwang, AB
    Wu, MC
    Patt, BE
    Iwanczyk, JS
    Hasegawa, BH
    JOURNAL OF NUCLEAR MEDICINE, 2002, 43 (05) : 10P - 11P