Retrieval study on hydroxyapatite-coated acetabular cups

被引:3
|
作者
Hu, WP [1 ]
Lai, KA [1 ]
Lin, CH [1 ]
Yang, CY [1 ]
Chang, GL [1 ]
机构
[1] Natl Cheng Kung Univ, Inst Biomed Engn, Tainan 701, Taiwan
关键词
acetabular cup; total hip replacement (THR); hydroxyapatite; hip implant;
D O I
10.1109/NEBC.2002.999437
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. A large scaled retrieved study of failed HA-coated acetabular cup was done by identifying and quantifying residual HA area on the surface of the retrieved cup for understanding the relevance between HA absorption and failure modes. Design. A total of 49 retrieved acetabular cups were collected through a long period of time for the study. Two main groups among the 49 cups were Landos and Osteonics cups. Methods. Image processing techniques and mathematical operation methods were used to quantify the residual HA area on the retrieved cups. Data incorporated with patients' histories and residual HA area for statistical analysis. Results. The HA coated surface of Landos cup disappears at a rate of 15.14% and 8.8% of the total surface per year for he unstable and the stable cups, respectively. The average loss rate of the HA coating of Osteonics cup in unstable cases is also faster than that in stable cases. Conclusions. The HA coating disappeared significant slower on the stable cups than did the unstable cups. The strong effect of mechanical fixation is necessary to achieve well biological fixation result.
引用
收藏
页码:5 / 6
页数:2
相关论文
共 50 条
  • [31] Plate fixation with hydroxyapatite-coated screws: A comparative loaded study
    Moroni, A
    Faldini, C
    Giannini, S
    Wippermann, B
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (408) : 262 - 267
  • [32] Superior 11-year survival but higher polyethylene wear of hydroxyapatite-coated Mallory-Head cups
    Gottliebsen, Martin
    Rahbek, Ole
    Ottosen, Per F.
    Soballe, Kjeld
    Stilling, Maiken
    HIP INTERNATIONAL, 2012, 22 (01) : 35 - 40
  • [33] Microstructural disintegration in dense hydroxyapatite and hydroxyapatite-coated metal implants
    Seo, Dong Seok
    Lee, Jong Kook
    METALS AND MATERIALS INTERNATIONAL, 2007, 13 (04) : 311 - 316
  • [34] Microstructural disintegration in dense hydroxyapatite and hydroxyapatite-coated metal implants
    Dong Seok Seo
    Jong Kook Lee
    Metals and Materials International, 2007, 13
  • [35] A tribological study of UHMWPE acetabular cups and polyurethane compliant layer acetabular cups
    Smith, SL
    Ash, HE
    Unsworth, A
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 2000, 53 (06): : 710 - 716
  • [36] HYDROXYAPATITE-COATED HIP PROSTHESES - EARLY RESULTS FROM AN INTERNATIONAL STUDY
    TONINO, AJ
    ROMANINI, L
    ROSSI, P
    BORRONI, M
    GRECO, F
    GARCIAARAUJO, C
    GARCIADIHINX, L
    MURCIAMAZON, A
    HEIN, W
    ANDERSON, J
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1995, (312) : 211 - 225
  • [37] Osseointegration of a hydroxyapatite-coated multilayered mesh stem
    Kusakabe, H
    Sakamaki, T
    Nihei, K
    Oyama, Y
    Yanagimoto, S
    Ichimiya, M
    Kimura, J
    Toyama, Y
    BIOMATERIALS, 2004, 25 (15) : 2957 - 2969
  • [38] Hydroxyapatite-coated implants: A case against their use
    Albrektsson, T
    JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1998, 56 (11) : 1312 - 1326
  • [39] EFFECT OF DICLOFENAC ON FIXATION OF HYDROXYAPATITE-COATED IMPLANTS - AN EXPERIMENTAL-STUDY
    JACOBSSON, SA
    DJERF, K
    IVARSSON, I
    WAHLSTROM, O
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (05): : 831 - 833
  • [40] Analysis of retrieved hydroxyapatite-coated hip prostheses
    Gross, KA
    Walsh, W
    Swarts, E
    JOURNAL OF THERMAL SPRAY TECHNOLOGY, 2004, 13 (02) : 190 - 199