Retrieval study on hydroxyapatite-coated acetabular cups

被引:3
|
作者
Hu, WP [1 ]
Lai, KA [1 ]
Lin, CH [1 ]
Yang, CY [1 ]
Chang, GL [1 ]
机构
[1] Natl Cheng Kung Univ, Inst Biomed Engn, Tainan 701, Taiwan
关键词
acetabular cup; total hip replacement (THR); hydroxyapatite; hip implant;
D O I
10.1109/NEBC.2002.999437
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. A large scaled retrieved study of failed HA-coated acetabular cup was done by identifying and quantifying residual HA area on the surface of the retrieved cup for understanding the relevance between HA absorption and failure modes. Design. A total of 49 retrieved acetabular cups were collected through a long period of time for the study. Two main groups among the 49 cups were Landos and Osteonics cups. Methods. Image processing techniques and mathematical operation methods were used to quantify the residual HA area on the retrieved cups. Data incorporated with patients' histories and residual HA area for statistical analysis. Results. The HA coated surface of Landos cup disappears at a rate of 15.14% and 8.8% of the total surface per year for he unstable and the stable cups, respectively. The average loss rate of the HA coating of Osteonics cup in unstable cases is also faster than that in stable cases. Conclusions. The HA coating disappeared significant slower on the stable cups than did the unstable cups. The strong effect of mechanical fixation is necessary to achieve well biological fixation result.
引用
收藏
页码:5 / 6
页数:2
相关论文
共 50 条
  • [21] Hydroxyapatite-coated external fixation pins
    Moroni, A
    Pegreffi, F
    Cadossi, M
    Hoang-Kim, A
    Lio, V
    Giannini, S
    EXPERT REVIEW OF MEDICAL DEVICES, 2005, 2 (04) : 465 - 471
  • [22] HYDROXYAPATITE-COATED TOTAL HIP PROSTHESES
    WICKI, O
    MIKIC, S
    GERBER, P
    HELVETICA CHIRURGICA ACTA, 1990, 57 (01) : 107 - 115
  • [23] BIOCOMPATIBILITY OF HYDROXYAPATITE-COATED HIP PROSTHESES
    SANTAVIRTA, S
    NORDSTROM, D
    YLINEN, P
    KONTTINEN, YT
    SILVENNOINEN, T
    ROKKANEN, P
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1991, 110 (06) : 288 - 292
  • [24] The survival of hydroxyapatite-coated implants is questioned
    Nico Creugers
    Evidence-Based Dentistry, 2002, 3 (3) : 77 - 78
  • [25] Hydroxyapatite-coated antibacterial air filters
    Lee, Seung-Hyun
    Ryu, Su-Chak
    Park, Hong Chae
    JOURNAL OF CERAMIC PROCESSING RESEARCH, 2014, 15 (01): : 21 - 25
  • [26] Hydroxyapatite-coated polyurethane for auricular cartilage replacement:: An in vitro study
    Chetty, Avashnee
    Steynberg, Tenille
    Moolman, Sean
    Nilen, Roger
    Joubert, Annie
    Richter, Wim
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A, 2008, 84A (02) : 475 - 482
  • [27] Hydroxyapatite-coated implants: A case for their use
    Morris, HF
    Ochi, S
    JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1998, 56 (11) : 1303 - 1311
  • [28] Osteoinductive hydroxyapatite-coated titanium implants
    Ripamonti, Ugo
    Roden, Laura C.
    Renton, Louise F.
    BIOMATERIALS, 2012, 33 (15) : 3813 - 3823
  • [29] MIGRATION OF HYDROXYAPATITE-COATED FEMORAL PROSTHESES - A ROENTGEN STEREOPHOTOGRAMMETRIC STUDY
    SOBALLE, K
    TOKSVIGLARSEN, S
    GELINECK, J
    FRUENSGAARD, S
    HANSEN, ES
    RYD, L
    LUCHT, U
    BUNGER, C
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1993, 75 (05): : 681 - 687
  • [30] Ten-year results of ALIZE acetabular cup with hydroxyapatite coating and AURA hydroxyapatite-coated stem in total hip arthroplasty
    Fayard, Jean-Philippe
    Chalencon, Francois
    Passot, Jean-Paul
    Latour, Laurent Dupre
    Edorh, Gregoire
    JOURNAL OF ARTHROPLASTY, 2006, 21 (07): : 1021 - 1025