Retrieval study on hydroxyapatite-coated acetabular cups

被引:3
|
作者
Hu, WP [1 ]
Lai, KA [1 ]
Lin, CH [1 ]
Yang, CY [1 ]
Chang, GL [1 ]
机构
[1] Natl Cheng Kung Univ, Inst Biomed Engn, Tainan 701, Taiwan
关键词
acetabular cup; total hip replacement (THR); hydroxyapatite; hip implant;
D O I
10.1109/NEBC.2002.999437
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. A large scaled retrieved study of failed HA-coated acetabular cup was done by identifying and quantifying residual HA area on the surface of the retrieved cup for understanding the relevance between HA absorption and failure modes. Design. A total of 49 retrieved acetabular cups were collected through a long period of time for the study. Two main groups among the 49 cups were Landos and Osteonics cups. Methods. Image processing techniques and mathematical operation methods were used to quantify the residual HA area on the retrieved cups. Data incorporated with patients' histories and residual HA area for statistical analysis. Results. The HA coated surface of Landos cup disappears at a rate of 15.14% and 8.8% of the total surface per year for he unstable and the stable cups, respectively. The average loss rate of the HA coating of Osteonics cup in unstable cases is also faster than that in stable cases. Conclusions. The HA coating disappeared significant slower on the stable cups than did the unstable cups. The strong effect of mechanical fixation is necessary to achieve well biological fixation result.
引用
收藏
页码:5 / 6
页数:2
相关论文
共 50 条
  • [1] Failure of dual radius hydroxyapatite-coated acetabular cups
    D'Angelo F.
    Molina M.
    Riva G.
    Zatti G.
    Cherubino P.
    Journal of Orthopaedic Surgery and Research, 3 (1)
  • [2] Clinical and radiological results of hydroxyapatite-coated acetabular cups
    Blacha, J
    Gagala, J
    INTERNATIONAL ORTHOPAEDICS, 2004, 28 (06) : 362 - 365
  • [3] Clinical and radiological results of hydroxyapatite-coated acetabular cups
    J. Blacha
    J. Gągała
    International Orthopaedics, 2004, 28 : 362 - 365
  • [4] Early failure of hemispheric hydroxyapatite-coated acetabular cups
    Kim, Shin-Yoon
    Kim, Do-Heon
    Kim, Yong-Goo
    Oh, Chang-Wug
    Ihn, Joo-Chul
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2006, (446) : 233 - 238
  • [5] Aseptic loosening of 2 rim-fix, hydroxyapatite-coated acetabular cups
    Benkö, TZ
    Santiago-Martin, A
    Ruddlesdin, C
    Selzer, G
    JOURNAL OF ARTHROPLASTY, 2002, 17 (04): : 519 - 523
  • [6] The performance of two hydroxyapatite-coated acetabular cups compared with Charnley cups -: From the Norwegian Arthroplasty Register
    Havelin, LI
    Espehaug, B
    Engesæter, LB
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2002, 84B (06): : 839 - 845
  • [7] Cementless hemispheric hydroxyapatite-coated sockets for acetabular revision
    Geerdink, Carel H.
    Schaafsma, Jelle
    Meyers, Will G. M.
    Grimm, Bernd
    Tonino, Alphons J.
    JOURNAL OF ARTHROPLASTY, 2007, 22 (03): : 369 - 376
  • [8] Survival of hydroxyapatite-coated cups: acetabular screws involve a lower rate of revision surgery due to aseptic loosening
    Natera, Luis
    Valera, Marius
    Moya Gomez, Esther
    Ibanez, Natalia
    Crusi, Xavier
    Sancho, Rogelio
    HIP INTERNATIONAL, 2017, 27 (02) : 153 - 161
  • [9] Failure of hydroxyapatite-coated acetabular cups - Ten-year follow-up of 85 Landos Atoll arthroplasties
    Lai, KA
    Shen, WJ
    Chen, CH
    Yang, CY
    Hu, WP
    Chang, GL
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2002, 84B (05): : 641 - 646
  • [10] RETRIEVAL ANALYSIS OF A HYDROXYAPATITE-COATED HIP-PROSTHESIS
    BLOEBAUM, RD
    MERRELL, M
    GUSTKE, K
    SIMMONS, M
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1991, (267) : 97 - 102