Radiation protection for an intraoperative X-ray source compared to C-arm fluoroscopy

被引:14
|
作者
Schneider, Frank [1 ]
Clausen, Sven [1 ]
Jahnke, Anika [1 ]
Steil, Volker [1 ]
Bludau, Frederic [2 ]
Suetterlin, Marc [3 ]
Obertacke, Udo [2 ]
Wenz, Frederik [1 ]
机构
[1] Heidelberg Univ, Univ Med Ctr Mannheim, Dept Radiat Oncol, D-68167 Mannheim, Germany
[2] Heidelberg Univ, Univ Med Ctr Mannheim, Dept Trauma Surg, D-68167 Mannheim, Germany
[3] Heidelberg Univ, Univ Med Ctr Mannheim, Dept Obstet & Gynaecol, D-68167 Mannheim, Germany
来源
ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK | 2014年 / 24卷 / 03期
关键词
intraoperative radiotherapy; Intrabeam; radiation protection; SOCIETY CONSENSUS GUIDELINES; AMERICAN BRACHYTHERAPY; BREAST-CANCER; INTERSTITIAL RADIOSURGERY; RADIOTHERAPY-TARGIT; KYPHO-IORT; KYPHOPLASTY; DEVICE; DOSIMETRY;
D O I
10.1016/j.zemedi.2013.10.006
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Intraoperative radiotherapy (IORT) using the INTRABEAM (R) system promises a flexible use regarding radiation protection compared to other approaches such as electron treatment or HDR brachytherapy with Ir-192 or Co-60. In this study we compared dose rate measurements of breast- and Kypho-IORT with C-arm fluoroscopy which is needed to estimate radiation protection areas. Materials and Methods: C-arm fluoroscopy, breast- and Kypho-IORTs were performed using phantoms (silicon breast or bucket of water). Dose rates were measured at the phantom's surface, at 30 cm, 100 cm and 200 cm distance. Those measurements were confirmed during 10 Kypho-IORT and 10 breast-IORT patient treatments. Results: The measured dose rates were in the same magnitude for all three paradigms and ranges from 20 mu Sv/h during a simulated breast-IORT at two meter distance up to 64 mSv/h directly at the surface of a simulated Kypho-IORT. Those measurements result in a circle of controlled area (yearly doses >6 mSv) for each paradigm of about 4 m +/- 2 m. Discussion/Conclusions: All three paradigms show comparable dose rates which implies that the radiation protection is straight forward and confirms the flexible use of the INTRABEAM (R) system.
引用
收藏
页码:243 / 251
页数:9
相关论文
共 50 条
  • [31] EQUIPMENT PERFORMANCE AND RADIATION PROTECTION STATUS IN X-RAY FLUOROSCOPY UNITS IN SUDAN
    Ahmed, Nada A.
    Nayl, A. I.
    Suliman, I. I.
    RADIATION PROTECTION DOSIMETRY, 2012, 148 (02) : 174 - 180
  • [32] Types and Arrangement of Thyroid Shields to Reduce Exposure of Surgeons to Ionizing Radiation During Intraoperative Use of C-arm Fluoroscopy
    Lee, Seung Yeol
    Min, Eungi
    Bae, Jaekeon
    Chung, Chin Youb
    Lee, Kyoung Min
    Kwon, Soon-Sun
    Park, Moon Seok
    Lee, Kisung
    SPINE, 2013, 38 (24) : 2108 - 2112
  • [33] Geometric correction and error analysis of C-arm X-ray image distortion
    Robotics Institute, Beijing University of Aeronautics and Astronautics, Beijing 100083, China
    不详
    Gaojishu Tongxin, 2006, 6 (600-605):
  • [34] Perfusion Estimation in the Peripheral Vasculature Using C-arm X-ray Systems
    Giordano, Marco
    Neukirchen, Christoph
    Bertram, Matthias
    Mali, Willem
    Viergever, Max A.
    Vonken, Evert-Jan
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 3017 - +
  • [35] Hysterosalpingography with use of mobile C-arm fluoroscopy
    Phillips, Jordana
    Cochavi, Saadia
    Silberzweig, James E.
    FERTILITY AND STERILITY, 2010, 93 (06) : 2065 - 2068
  • [36] Assessment of Radiation Dose and Shielding Efficacy in DDR Versus Mobile C-Arm and Portable X-Ray Systems
    Siddique, A.
    Ge, G. Y.
    Lemieux, B.
    Zhang, J.
    MEDICAL PHYSICS, 2024, 51 (10) : 7775 - 7775
  • [37] Targeting of X-ray visible microencapsulated mesenchymal cells with C-arm CT
    Kraitchman, D.
    Kedziorek, D.
    Walczak, P.
    Brost, A.
    Fu, Y.
    Azene, N.
    Wacker, F.
    EUROPEAN HEART JOURNAL SUPPLEMENTS, 2009, 11 (0B) : S73 - S74
  • [38] Mechanical thrombectomy using a combined CT/C-arm X-ray system
    Pfaff, Johannes
    Herweh, Christian
    Pham, Mirko
    Schoenenberger, Silvia
    Boesel, Julian
    Ringleb, Peter A.
    Heiland, Sabine
    Bendszus, Martin
    Moehlenbruch, Markus
    JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2016, 8 (06) : 621 - 625
  • [39] MEASUREMENT OF RADIATION DOSES TO THE EYE LENS DURING ORTHOPEDIC SURGERY USING AN C-ARM X-RAY SYSTEM
    Suzuki, Akira
    Matsubara, Kosuke
    Sasa, Yuko
    RADIATION PROTECTION DOSIMETRY, 2018, 179 (02) : 189 - 195
  • [40] Experimental Evaluation of the Accuracy at the C-Arm Pose Estimation with X-Ray Images
    Thuerauf, Sabine
    Vogt, Florian
    Hornung, Oliver
    Koerner, Mario
    Nasseri, M. Ali
    Knoll, Alois
    2016 38TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2016, : 3859 - 3862