Stereotactic radiotherapy for lung cancer using a flattening filter free Clinac

被引:77
|
作者
Vassiliev, Oleg N. [1 ]
Kry, Stephen F.
Chang, Joe Y. [2 ]
Balter, Peter A.
Titt, Uwe
Mohan, Radhe
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Unit 94, Houston, TX 77030 USA
[2] Univ Texas MD Anderson Canc Ctr, Dept Radiat Oncol, Houston, TX 77030 USA
来源
关键词
lung cancer; flattening filter; clinical linear accelerator; stereotactic radiotherapy; BODY RADIATION-THERAPY; FREE CLINICAL ACCELERATOR; INTENSITY-MODULATED RADIOTHERAPY; MEDICALLY INOPERABLE PATIENTS; SINGLE-DOSE RADIOTHERAPY; HYPOFRACTIONATED RADIOTHERAPY; DOSIMETRIC PROPERTIES; PHOTON BEAMS; MONTE-CARLO; OUTCOMES;
D O I
10.1120/jacmp.v10i1.2880
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The objective of this study was to assess the feasibility of stereotactic radiotherapy for early stage lung cancer using photon beams from a Varian Clinac accelerator operated without a flattening filter. Treatment plans were generated for 10 lung cancer patients with isolated lesions less than 3 cm in diameter. For each patient, two plans were generated, one with and one without the flattening filter. Plans were generated with Eclipse 8.0 (Varian Medical Systems) commissioned with beam data measured on a Clinac 21EX (Varian Medical Systems) operated with and without the flattening filter. Removal of the flattening filter increased the dose rate. The median beam-on time per field was reduced from 25 sec (with the filter) to 11 sec (without the filter), increasing the feasibility of breath-hold treatments and the efficiency of gated treatments. Differences in a dose heterogeneity index for the planning target volume between plans with flattened and unflattened beams were statistically insignificant. Differences in mean doses to organs at risk were small, typically about 10 cGy over the entire treatment. The study concludes that radiotherapy with unflattened beams is feasible and requires substantially less beam-on time, facilitating breath-hold and gating techniques.
引用
下载
收藏
页码:14 / 21
页数:8
相关论文
共 50 条
  • [31] Dosimetric comparison of deep inspiration breath hold and free breathing technique in stereotactic body radiotherapy for localized lung tumor using Flattening Filter Free beam
    Mani, Karthick Raj
    Bhuiyan, Md. Anisuzzaman
    Alam, Md. Mahbub
    Ahmed, Sharif
    Sumon, Mostafa Aziz
    Sengupta, Ashim Kumar
    Rahman, Md. Shakilur
    Islam, Md. S. M. Azharul
    POLISH JOURNAL OF MEDICAL PHYSICS AND ENGINEERING, 2018, 24 (01): : 15 - 24
  • [32] Stereotactic body radiation therapy for prostate cancer: a dosimetric comparison of IMRT and VMAT using flattening filter and flattening filter-free beams
    El-Sayed, Sherif M.
    El-Gebaly, Reem H.
    Fathy, Mohamed M.
    Abdelaziz, Dina M.
    RADIATION AND ENVIRONMENTAL BIOPHYSICS, 2024, 63 (03) : 423 - 431
  • [33] The comparison of properties for radiotherapy with flattening filter-free and flattening filter beam
    Gu, J. H.
    Won, H. S.
    Hong, J. W.
    Chang, N. J.
    Park, J. H.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S978 - S978
  • [34] FLATTENING FILTER FREE BEAMS FOR HYPOFRACTIONATED RADIOTHERAPY OF LOCALIZED PROSTATE CANCER
    Lang, S.
    Hrbacek, J.
    Najafi, Y.
    Streller, T.
    Klock, S.
    Studer, G.
    Zaugg, K.
    Luetolf, U. M.
    Zwahlen, D.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S188 - S189
  • [35] An assessment of the ExacTrac intrafraction imaging capabilities for flattening filter free prostate stereotactic body radiotherapy
    James Rijken
    Mark Sidhom
    Physical and Engineering Sciences in Medicine, 2020, 43 : 849 - 855
  • [36] An assessment of the ExacTrac intrafraction imaging capabilities for flattening filter free prostate stereotactic body radiotherapy
    Rijken, James
    Sidhom, Mark
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2020, 43 (03) : 849 - 855
  • [37] Investigating flattening filter free mode for peripheral lung stereotactic body radiation therapy
    Jensen, Michael
    Garcia, Lourdes M.
    Scora, Daryl
    Vandermeer, Aaron
    MEDICAL PHYSICS, 2017, 44 (08) : 4378 - 4378
  • [38] Stereotactic body radiotherapy with flattening filter-free beams for prostate cancer: assessment of patient-reported quality of life
    Marta Scorsetti
    Filippo Alongi
    Elena Clerici
    Tiziana Comito
    Antonella Fogliata
    Cristina Iftode
    Pietro Mancosu
    Piera Navarria
    Giacomo Reggiori
    Stefano Tomatis
    Elisa Villa
    Luca Cozzi
    Journal of Cancer Research and Clinical Oncology, 2014, 140 : 1795 - 1800
  • [39] Stereotactic body radiotherapy with flattening filter-free beams for prostate cancer: assessment of patient-reported quality of life
    Scorsetti, Marta
    Alongi, Filippo
    Clerici, Elena
    Comito, Tiziana
    Fogliata, Antonella
    Iftode, Cristina
    Mancosu, Pietro
    Navarria, Piera
    Reggiori, Giacomo
    Tomatis, Stefano
    Villa, Elisa
    Cozzi, Luca
    JOURNAL OF CANCER RESEARCH AND CLINICAL ONCOLOGY, 2014, 140 (10) : 1795 - 1800
  • [40] Dosimetric evaluation of different planning techniques based on flattening filter-free beams for central and peripheral lung stereotactic body radiotherapy
    Dwivedi, Shekhar
    Kansal, Sandeep
    Shukla, Jooli
    Bharati, Avinav
    Dangwal, Vinod Kumar
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2021, 7 (06)